Auxiliary device for calibration of roller pumps: development and experimental results

被引:1
|
作者
Vieira, F. U., Jr. [1 ,2 ]
Dantas de Medeiros, J., Jr. [3 ]
Antunes, N. [2 ]
Vieira, R. W. [2 ]
Costa, E. T. [1 ,3 ]
机构
[1] State Univ Campinas UNICAMP, Ctr Biomed Engn, Sao Paulo, Brazil
[2] State Univ Campinas UNICAMP, Sch Med Sci, Sao Paulo, Brazil
[3] State Univ Campinas UNICAMP, Sch Elect & Comp Engn, Dept Biomed Engn, Sao Paulo, Brazil
来源
PERFUSION-UK | 2012年 / 27卷 / 06期
关键词
roller pump; hemolysis; calibration device; CARDIOPULMONARY BYPASS; CENTRIFUGAL PUMPS; RETROGRADE FLOW; DYNAMIC METHOD; HEMOLYSIS; SURGERY; BLOOD;
D O I
10.1177/0267659112453077
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Roller pumps are commonly used in surgeries involving extracorporeal circulation and its regulation may influence the hemolysis rate. Objective: To describe an auxiliary device for adjusting roller pumps using the dynamic calibration method and show preliminary results of its application in surgical procedures. Method:The device was tested in the laboratory environment with three disposable pressure transducers normally used in surgeries. In the operation room, tests were carried out with two groups of patients. Free hemoglobin plasma rates were measured,5 minutes after perfusion started and immediately before the passage of the residual volume of the pump. For Group 1 (n = 22), occlusion measurements were carried out using the drop rate method and for Group 2 (n = 18), measurements were carried out using the auxiliary device and 300 mmHg calibration (mean pressure for the dynamic calibration method). Results:The auxiliary calibration device has shown calibration results statistically equal to those obtained with a reference device in the laboratory environment. We have found less variability of hemolysis rates in the operating room, using the auxiliary device and we did not observe any influence of unbalanced rollers in the hemolysis rates.
引用
收藏
页码:512 / 519
页数:8
相关论文
共 50 条
  • [31] Development of a temperature calibration device for thermoelastic stress analysis
    Dulieu-Barton, JM
    Quinn, S
    Eyre, C
    Cunningham, PR
    Advances in Experimental Mechanics, 2004, 1-2 : 197 - 202
  • [32] Development of Automatic Device for Calibration of Trace Radon Analyzers
    Huang Fan
    Liao Zhi-Liang
    Huang Long-Zhu
    Zhou Shu-Min
    Tang Bin
    Zhang Xing-Lei
    Chen Huan-Wen
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 39 (10) : 1506 - 1512
  • [33] NEW EXPERIENCE AND RESULTS FROM EXPERIMENTAL MEASUREMENT OF ADHESION ON A ROLLER RIG
    Voltr, Petr
    Zirek, Abdulkadir
    Kayaalp, Bekir Tuna
    XXIII KONFERENCE S MEZINARODNI UCASTI: SOUCASNE PROBLEMY V KOLEJOVYCH VOZIDLECH/ 23RD CONFERENCE: CURRENT PROBLEMS IN RAIL VEHICLES 2017, 2017, : 423 - 432
  • [34] THE DEVELOPMENT OF A NEW DEVICE-SYSTEM FOR VISUAL INSPECTIONS OF INFUSION PUMPS
    Lekscha, Thomas
    Durackova, Daniela
    ANNALS OF DAAAM FOR 2012 & PROCEEDINGS OF THE 23RD INTERNATIONAL DAAAM SYMPOSIUM - INTELLIGENT MANUFACTURING AND AUTOMATION - FOCUS ON SUSTAINABILITY, 2012, 23 : 143 - 146
  • [35] CALIBRATION OF A NUMERICAL MODEL FOR MASONRY WITH APPLICATION TO EXPERIMENTAL RESULTS
    Churilov, Sergey
    Dumova-Jovanoska, Elena
    ARCHITECTURE CIVIL ENGINEERING ENVIRONMENT, 2008, 1 (03) : 41 - 48
  • [36] Calibration of a numerical model for masonry with application to experimental results
    Dumova-Jovanoska, E.
    Churilov, S.
    PROTECTION OF HISTORICAL BUILDINGS - PROHITECH 09, VOL 1 AND 2, 2009, : 1139 - 1145
  • [37] Calibration of NAOS and CONICA static aberrations - Experimental results
    Hartung, A
    Blanc, A
    Fusco, T
    Lacombe, F
    Mugnier, LM
    Rousset, G
    Lenzen, R
    ASTRONOMY & ASTROPHYSICS, 2003, 399 (01) : 385 - 394
  • [38] Prototype fretting device and some experimental results
    Cheikh, M.
    Fermy, A.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2014, 228 (03) : 266 - 275
  • [39] NEW DEVICE FOR PERIPHERAL ATHERECTOMY - EXPERIMENTAL RESULTS
    REDHA, F
    UHLSCHMID, GK
    ANTONUCCI, F
    ZOLLIKOFER, CL
    RADIOLOGY, 1992, 185 : 229 - 229
  • [40] Performance of heat pipes as capillary pumps: modelling and comparison with experimental results
    Margaris, Dionissios P.
    Diamantis, Zisis G.
    Photeinos, Dionysios I.
    Tsahalis, Demos T.
    INTERNATIONAL JOURNAL OF LOW-CARBON TECHNOLOGIES, 2007, 2 (02) : 149 - 161