Comparison between cuff-based and radial tonometry exercise-induced central blood pressure

被引:6
|
作者
Morales-Acuna, Francisco [1 ]
Ratcliffe, Brycen [2 ]
Harrison, Caleb [2 ]
Crowe, Sierra [2 ]
Bockover, Evan [2 ]
Pawlak, Robert [2 ]
Gurovich, Alvaro N. [1 ]
机构
[1] Univ Texas El Paso, Coll Hlth Sci, Dept Rehabil Sci, 1851 Wiggins Way, El Paso, TX 79968 USA
[2] Indiana State Univ, Dept Appl Med & Rehabil, Coll Hlth & Human Serv, Terre Haute, IN 47809 USA
关键词
Exercise-induced central blood pressure; Pulse wave analysis; Cuff-based device; Radial tonometer; AUGMENTATION INDEX; SPHYGMOCOR XCEL; VALIDATION; DEVICE; RELIABILITY; DISEASE; AMPLIFICATION; HYPERTENSION; STANDARDS; RESPONSES;
D O I
10.1007/s00421-019-04079-9
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
PurposeNon-invasive central blood pressure assessed during exercise may provide better cardiovascular prognostic than measurements taken at rest. Radial tonometry is the only technique validated to perform this type of assessment; however, it relies on the experience of the tester. Cuff-based devices have been developed to avoid operator dependency, although these systems have yet to be validated during exercise. The purpose of this study was to compare exercise-induced central blood pressure estimations between a cuff-based device and radial tonometry.MethodsTwenty young healthy subjects were recruited to perform a three-workload steady-state exercise test at blood lactate levels of <2, 2-4, and >4mmol/L, respectively. Central systolic and diastolic blood pressure (cSBP and cDBP, respectively), central pulse pressure (cPP), and augmentation index (AIx) were assessed at rest and during each workload with a cuff-based device and radial tonometry. Statistical analysis included Bland-Altman analysis for agreement between techniques. Agreement was considered when 95% of the data set for each central blood pressure parameter was within 1.96 standard deviations from the mean difference. Significance was considered at =0.05.ResultsCentral blood pressure measurements with the cuff device were obtained only at rest and during low-intensity exercise. During low-intensity exercise, all measurements showed agreement between both devices (cSBP 95% CI [- 6.0 to 10.7], cDBP 95% CI [- 4.5 to 6.3], cPP 95% CI [- 4.7 to 8.3], and AIx (95% CI [- 20.1 to 22.2]).ConclusionA cuff-based device can estimate central blood pressure at low-intensity exercise, without operator dependency, and showing agreement to radial tonometry.
引用
收藏
页码:901 / 911
页数:11
相关论文
共 50 条
  • [31] Central blood pressure obtained by cuff-based oscillometry as a determinant of left ventricular hypertrophy in hypertensive patients
    Abdelmegid, Mohamed Aboel-Kassem F.
    Fouad, Doaa A.
    Beshay, Nardeen W. M.
    Mahran, Dalia G.
    Shams-eddin, Hamdy
    BLOOD PRESSURE MONITORING, 2023, 28 (06) : 322 - 329
  • [32] Comparing blood pressure measurements between a photoplethysmography-based and a standard cuff-based manometry device
    Dean Nachman
    Yftach Gepner
    Nir Goldstein
    Eli Kabakov
    Arik Ben Ishay
    Romi Littman
    Yuval Azmon
    Eli Jaffe
    Arik Eisenkraft
    Scientific Reports, 10
  • [33] CUFF-LESS BLOOD PRESSURE MEASUREMENT USING THE PULSE TRANSIT TIME - A COMPARISON TO CUFF-BASED OSCILLOMETRIC 24 HOUR BLOOD PRESSURE MEASUREMENT IN CHILDREN
    Hulpke-Wette, M.
    Goehler, A.
    Hofmann, E.
    Kuechler, G.
    JOURNAL OF HYPERTENSION, 2018, 36 : E73 - E73
  • [34] Effect of tissue mechanical properties on cuff-based blood pressure measurements
    Lan, H.
    Al-Jumaily, A. M.
    Lowe, A.
    Hing, W.
    MEDICAL ENGINEERING & PHYSICS, 2011, 33 (10) : 1287 - 1292
  • [35] ESTIMATION OF CENTRAL BLOOD PRESSURE BY CAROTID TONOMETRY, RADIAL TONOMETRY, AND BRACHIAL PLETHYSMOGRAPHY: A COMPARATIVE STUDY
    Cheng, Hao-min
    Sung, Shih-Hsien
    Hsu, Pai-Feng
    Wang, Kang-Ling
    Lee, Wen-Shin
    Yu, Wen-Chung
    Lin, Yao-Ping
    Chen, Chen-Huan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [36] COMPARISON OF CENTRAL SYSTOLIC BLOOD PRESSURE (CSBP) ESTIMATED BY DIRECT CENTRAL BLOOD PRESSURE (DCBP) FORMULA AND RADIAL TONOMETRY BY SPIGMOCOR: AN INVASIVE VALIDATION STUDY
    Leone, Dario
    Vallelonga, Fabrizio
    Bruno, Giulia
    Fragapani, Salvatore
    Cesareo, Marco
    Airale, Lorenzo
    Mingrone, Giulia
    Milan, Alberto
    JOURNAL OF HYPERTENSION, 2024, 42
  • [37] Invasive validation of a novel brachial cuff-based oscillometric device (SphygmoCor XCEL) for measuring central blood pressure
    Shoji, Toshihiro
    Nakagomi, Atsushi
    Okada, Sho
    Ohno, Yuji
    Kobayashi, Yoshio
    JOURNAL OF HYPERTENSION, 2017, 35 (01) : 69 - 75
  • [38] Pilot study of comparability of a smartphone blood pressure monitoring algorithm to conventional cuff-based blood pressure measurements
    Vischer, A.
    Rosania, J.
    Socrates, T.
    Blaschke, C.
    Proenca, M.
    Bonnier, G.
    Proust, Y. -M.
    Ferrario, D.
    Eckstein, J.
    Lemay, M.
    Burkard, T.
    EUROPEAN HEART JOURNAL, 2021, 42 : 2350 - 2350
  • [39] Discrepancy Between Tonometric Ambulatory and Cuff-Based Office Blood Pressure Measurements in Patients With Type 1 Diabetes
    Theilade, Simone
    Lajer, Maria
    Joergensen, Christel
    Persson, Frederik
    Rossing, Peter
    JOURNAL OF CLINICAL HYPERTENSION, 2012, 14 (10): : 686 - 693
  • [40] Direct estimation of central systolic blood pressure from radial arterial tonometry
    Izzo, J.
    Kafle, P.
    Osmond, P.
    JOURNAL OF HYPERTENSION, 2008, 26 : S369 - S370