Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples
被引:5
|作者:
Yong, Sharon
[1
]
Liu, Hong
[1
]
Wong, Lingkai
[1
]
Teo, Hui Ling
[1
]
Chen, Yizhao
[1
]
Liu, Qinde
[1
]
Teo, Tang Lin
[1
]
Lee, Tong Kooi
[1
]
机构:
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词:
Urea;
Human serum;
Isotope dilution mass spectrometry;
Liquid chromatography-isotope dilution;
tandem mass spectrometry;
Reference measurement procedure;
External quality assessment;
D O I:
10.1016/j.ijms.2017.01.013
中图分类号:
O64 [物理化学(理论化学)、化学物理学];
O56 [分子物理学、原子物理学];
学科分类号:
070203 ;
070304 ;
081704 ;
1406 ;
摘要:
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文