Evaluation of four pediatric cardiopulmonary bypass circuits in terms of perfusion quality and capturing gaseous microemboli

被引:14
|
作者
Mathis, R. K. [1 ]
Lin, J. [1 ]
Dogal, N. M. [1 ]
Qiu, F. [1 ]
Kunselman, A.
Wang, S. [1 ]
Uendar, A. [1 ,2 ,3 ]
机构
[1] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Penn State Hershey Pediat Cardiovasc Res Ctr,Dept, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Dept Surg, Hershey, PA USA
[3] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Dept Bioengn, Hershey, PA USA
来源
PERFUSION-UK | 2012年 / 27卷 / 06期
关键词
cardiopulmonary bypass; hollow-fiber membrane oxygenators; microemboli; purge line flow; pediatrics; INTEGRATED ARTERIAL FILTERS; MEMBRANE OXYGENATORS;
D O I
10.1177/0267659112453078
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study compared four pediatric cardiopulmonary bypass (CPB) circuits with four different hollow-fiber membrane oxygenators and their specific reservoirs, Capiox RX15,Quadrox-i pediatric, Quadrox-i pediatric with integrated arterial filter (IAF) and KIDS D101, in a simulated CPB circuit identical to that used in the clinical setting at our institution to test their ability to maintain hemodynamic properties, remove gaseous microemboli (GME), and to test the amount of blood "stolen" by the arterial filter purge line. The circuit was first primed with Ringer's Lactate solution, then red blood cells were added and the hematocrit was maintained at 30%. A 5-cc bolus of air was injected just proximal to the venous reservoir over a thirty-second interval and GME were monitored using an Emboli Detection and Classification quantifier. Transducers were placed at pre-oxygenator, post-oxygenator and distal arterial line (post-filter) positions. Flow probes were also placed both pre and post filter. The injections were made at three flow rates, hypothermic and normothermic temperatures, and with the purge line in both the opened and closed positions. Six injections were done at each of the 12 experimental conditions. Results demonstrated that GME in the arterial line increased with increasing temperature and flow rate. The Capiox RX15 had the least GME in the arterial line at all experimental conditions. The KIDS D101 had the largest pressure drop and the lowest retention of hemodynamic energy, while the Capiox had the lowest pressure drop. All of the oxygenators had a similar amount of "stolen" blood flow and it was consistently under 10% of the total flow reaching the patient.
引用
收藏
页码:470 / 479
页数:10
相关论文
共 50 条
  • [1] An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Melchior, Richard W.
    Rosenthal, Tami
    Glatz, Andrew C.
    PERFUSION-UK, 2010, 25 (04): : 255 - 263
  • [2] An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Qiu, Feng
    Talor, Jonathan
    Uendar, Akif
    PERFUSION-UK, 2011, 26 (02): : 167 - 168
  • [3] Commentary on: An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Resley, Justin
    Melchior, Richard
    Beavers, James
    Shearer, Ian
    Riley, Jeffrey
    Groom, Robert
    PERFUSION-UK, 2010, 25 (04): : 265 - 266
  • [4] SOURCES OF GASEOUS MICROEMBOLI DURING CARDIOPULMONARY BYPASS
    HILL, AG
    GROOM, RC
    TEWKSBURY, L
    MUNOZ, R
    MARINO, JA
    LEFRAK, EA
    PROCEEDINGS OF THE AMERICAN ACADEMY OF CARDIOVASCULAR PERFUSION, VOL 9, 1988, 9 : 122 - 130
  • [5] Clinical Real-Time Monitoring of Gaseous Microemboli in Pediatric Cardiopulmonary Bypass
    Wang, Shigang
    Woitas, Karl
    Clark, J. Brian
    Myers, John L.
    Undar, Akif
    ARTIFICIAL ORGANS, 2009, 33 (11) : 1026 - 1030
  • [6] Evaluation of Membrane Oxygenators and Reservoirs in Terms of Capturing Gaseous Microemboli and Pressure Drops
    Guan, Yulong
    Palanzo, David
    Kunselman, Allen
    Undar, Akif
    ARTIFICIAL ORGANS, 2009, 33 (11) : 1037 - 1043
  • [7] Measurement of gaseous microemboli in the prime before the initiation of cardiopulmonary bypass
    Husebraten, Inger M.
    Fiane, Arnt E.
    Ringdal, Mari-Anne I. L.
    Thiara, Amrit P. S.
    PERFUSION-UK, 2018, 33 (01): : 30 - 35
  • [8] Microemboli Detection and Classification During Pediatric Cardiopulmonary Bypass
    Clark, Joseph B.
    Qiu, Feng
    Guan, Yulong
    Woitas, Karl R.
    Myers, John L.
    Undar, Akif
    WORLD JOURNAL FOR PEDIATRIC AND CONGENITAL HEART SURGERY, 2011, 2 (01) : 111 - 114
  • [9] Elimination of Gaseous Microemboli From Cardiopulmonary Bypass Using Hypobaric Oxygenation
    Gipson, Keith E.
    Rosinski, David J.
    Schonberger, Robert B.
    Kubera, Cathryn
    Mathew, Eapen S.
    Nichols, Frank
    Dyckman, William
    Courtin, Francois
    Sherburne, Bradford
    Bordey, Angelique F.
    Gross, Jeffrey B.
    ANNALS OF THORACIC SURGERY, 2014, 97 (03): : 879 - 887
  • [10] Generation, detection and prevention of gaseous microemboli during cardiopulmonary bypass procedure
    Lou, Song
    CARDIOLOGY, 2011, 120 : 103 - 103