Comparing different methods for estimating radiation dose to the conceptus

被引:8
|
作者
Lopez-Rendon, X. [1 ]
Walgraeve, M. S. [2 ]
Woussen, S. [2 ]
Dedulle, A. [1 ]
Zhang, G. [2 ]
Bosmans, H. [1 ,2 ]
Zanca, F. [1 ,3 ]
机构
[1] Katholieke Univ Leuven, Div Med Phys & Qual Assessment, Dept Imaging & Pathol, Herestr 49,Box 7003, B-3000 Leuven, Belgium
[2] Univ Hosp Leuven, Dept Radiol, Herestr 49, B-3000 Leuven, Belgium
[3] GE Healthcare, Buc, France
关键词
Pregnancy; Dose to the conceptus; Tube current modulation; Commercial software package; Monte Carlo simulations; MONTE-CARLO SIMULATIONS; TUBE CURRENT MODULATION; PREGNANT PATIENTS; MULTIDETECTOR CT; GESTATION; PATIENT; PHANTOM; FETUS; DEPTH;
D O I
10.1007/s00330-016-4389-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare different methods available in the literature for estimating radiation dose to the conceptus (D-conceptus) against a patient-specific Monte Carlo (MC) simulation and a commercial software package (CSP). Eight voxel models from abdominopelvic CT exams of pregnant patients were generated. D-conceptus was calculated with an MC framework including patient-specific longitudinal tube current modulation (TCM). For the same patients, dose to the uterus, D-uterus, was calculated as an alternative for D-conceptus, with a CSP that uses a standard-size, non-pregnant phantom and a generic TCM curve. The percentage error between D-uterus and D-conceptus was studied. Dose to the conceptus and percent error with respect to D-conceptus was also estimated for three methods in the literature. The percentage error ranged from -15.9% to 40.0% when comparing MC to CSP. When comparing the TCM profiles with the generic TCM profile from the CSP, differences were observed due to patient habitus and conceptus position. For the other methods, the percentage error ranged from -30.1% to 13.5% but applicability was limited. Estimating an accurate D-conceptus requires a patient-specific approach that the CSP investigated cannot provide. Available methods in the literature can provide a better estimation if applicable to patient-specific cases. aEuro cent A patient's internal anatomy affects the dose to the conceptus. aEuro cent Conceptus position has an influence on its dose estimation. aEuro cent Patient anatomy and specific TCM must be considered for accurate conceptus dosimetry. aEuro cent D (uterus) to a standard-size phantom should not be used as D (conceptus) .
引用
收藏
页码:851 / 858
页数:8
相关论文
共 50 条
  • [1] Comparing different methods for estimating radiation dose to the conceptus
    X. Lopez-Rendon
    M. S. Walgraeve
    S. Woussen
    A. Dedulle
    G. Zhang
    H. Bosmans
    F. Zanca
    European Radiology, 2017, 27 : 851 - 858
  • [2] Comparing different methods of estimating cosmic radiation exposure of airline personnel
    Hammer, GP
    Zeeb, H
    Tveten, U
    Blettner, M
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2000, 39 (04) : 227 - 231
  • [3] Comparing different methods of estimating cosmic radiation exposure of airline personnel
    G. P. Hammer
    Hajo Zeeb
    U. Tveten
    Maria Blettner
    Radiation and Environmental Biophysics, 2000, 39 : 227 - 231
  • [4] THE CLINICAL SIGNIFICANCE OF DIFFERENT METHODS OF ESTIMATING TUMOUR DOSE
    BODEN, G
    COHEN, M
    BRITISH JOURNAL OF RADIOLOGY, 1952, 25 (300): : 646 - 648
  • [5] Radiation dose to conceptus resulting from tangential breast irradiation
    Mazonakis, M
    Varveris, H
    Damilakis, J
    Theoharopoulos, N
    Gourtsoyiannis, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (02): : 386 - 391
  • [6] Estimating Radiation Dose in CT
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2011, 38 (06) : 3703 - +
  • [7] Comparing different extraction methods for estimating phosphorus solubility in various soil types
    Koopmans, GF
    Chardon, WJ
    Dekker, PHM
    Römkens, PFAM
    Schoumans, OF
    SOIL SCIENCE, 2006, 171 (02) : 103 - 116
  • [8] Comparing different recalibrated methods for estimating mean radiant temperature in outdoor environment
    Ouyang, Wanlu
    Liu, Zhixin
    Lau, Kevin
    Shi, Yuan
    Ng, Edward
    BUILDING AND ENVIRONMENT, 2022, 216
  • [9] Comparing Methods for Estimating Habitat Suitability
    Muhammed, Khaleel
    Anandhi, Aavudai
    Chen, Gang
    LAND, 2022, 11 (10)
  • [10] Conceptus radiation dose and risk from cardiac catheter ablation procedures
    Damilakis, J
    Theocharopoulos, N
    Perisinakis, K
    Manios, E
    Dimitriou, P
    Vardas, P
    Gourtsoyiannis, N
    CIRCULATION, 2001, 104 (08) : 893 - 897