ATCA data acquisition system for gamma-ray spectrometry

被引:30
|
作者
Pereira, R. C. [1 ]
Sousa, J. [1 ]
Fernandes, A. M. [1 ]
Patricio, F. [1 ]
Carvalho, B. [1 ]
Neto, A. [1 ]
Varandas, C. A. F. [1 ]
Gorini, G. [2 ]
Tardocchi, M. [2 ]
Gin, D. [3 ]
Shevelev, A. [3 ]
机构
[1] Univ Tecn Lisboa, IST Ctr Fusao Nucl, Inst Super Tecn, EURATOM Assoc, Lisbon, Portugal
[2] EURATOM ENEA CNR Assoc, Ist Fis Plasma, Milan, Italy
[3] Russian Acad Sci, AF Ioffe Physicotech Inst, St Petersburg 194021, Russia
关键词
data acquisition; gamma-ray spectroscopy; PHA; pulse processing;
D O I
10.1016/j.fusengdes.2007.10.011
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The gamma-ray spectrometer JET EP2 (Joint European Torus enhancement project 2) project aims to perform high-resolution gamma spectroscopy at very high count rate (up to few MHz). Traditional analogue electronic has count rate and pulse processing limitations (long dead-time, pile-up challenge). Digital pulse processing (DPP) systems have been shown to have better performance than analogue ones for processing neutrons or/and gamma-ray signals. DPP can synthesize almost any pulse response shape without the signal degradation associated to complex analogue paths. High-speed transient recorders (TR) with auto-trigger functionality are used to digitize and store the detailed shape of pulses. The data acquisition (DAQ) system provides sophisticated analysis/data reduction based on real time algorithms, implemented in field programmable gate arrays (FPGA), for Pulse Height Analysis (PHA) while resolving pulse pile-up of digitized pulses. This paper describes a new DAQ system for real-time pulse analysis. The system is based on the Advanced Telecommunications Computing Architecture (TM) (ATCA (TM)) and contains an ix86-based processor blade with up to 40 GFLOPS and a TR module interconnected through PCI Express (PCIe) links. TR module features: (i) 8 channels of 13 bit resolution with accuracy equal or higher than 11 bit to cope with the expected signal-to-noise ratio (SNR) of the input pulses; (ii) up to 500 MSamples/s sampling rate with the possibility to achieve 1 GSamples/s; and (iii) 2 or 4 GB of local memory. The core of the TR module is two FPGAs able to perform real-time processing algorithms such as PHA and pile-up resolution. This will allow data reduction by a factor of at least 6 and eventually spectra output in real-time. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:341 / 345
页数:5
相关论文
共 50 条
  • [21] Design of data acquisition system and algorithm research for omnidirectional gamma-ray positioning equipment
    胡婷婷
    帅磊
    王培林
    丰宝桐
    张译文
    孙芸华
    李晓辉
    魏书军
    单保慈
    魏龙
    Chinese Physics C, 2014, 38 (01) : 95 - 99
  • [22] Design of data acquisition system and algorithm research for omnidirectional gamma-ray positioning equipment
    胡婷婷
    帅磊
    王培林
    丰宝桐
    张译文
    孙芸华
    李晓辉
    魏书军
    单保慈
    魏龙
    Chinese Physics C, 2014, (01) : 95 - 99
  • [23] AN AIRBORNE GAMMA-RAY SPECTROMETRY SYSTEM FOR GEOLOGICAL SURVEYING
    BRISTOW, Q
    DONHOFFER, DK
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1968, 11 (01): : 66 - +
  • [24] A new versatile underground gamma-ray spectrometry system
    Lutter, Guillaume
    Hult, Mikael
    Marissens, Gerd
    Andreotti, Erica
    Rosengard, Ulf
    Misiaszek, Marcin
    Yuksel, Ayhan
    Sahin, Namik
    APPLIED RADIATION AND ISOTOPES, 2013, 81 : 81 - 86
  • [25] AN ELECTRONICS SYSTEM FOR GAMMA-RAY SPECTROMETRY IN SATELLITE APPLICATIONS
    HARRINGTON, TM
    MARSHALL, JH
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1969, NS16 (01) : 314 - +
  • [26] ONLINE GAMMA-RAY DATA ACQUISITION AT THE POWER BURST FACILITY
    HARTWELL, JK
    KILLIAN, EW
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 242 (03): : 487 - 492
  • [27] JIGSAW - A GAMMA-RAY DATA ACQUISITION, DISPLAY AND ANALYSIS PROGRAM
    HAYWOOD, SE
    BAMFORD, GJ
    RESTER, AC
    COLDWELL, RL
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) : 901 - 903
  • [28] Traceability in gamma-ray spectrometry
    Glavic-Cindro, D.
    Korun, M.
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (7-8) : 1196 - 1199
  • [29] Underground gamma-ray spectrometry
    Hult, M
    Preusse, W
    Gasparro, J
    Köhler, M
    ACTA CHIMICA SLOVENICA, 2006, 53 (01) : 1 - 7
  • [30] Uncertainties in gamma-ray spectrometry
    Lepy, M. C.
    Pearce, A.
    Sima, O.