Setting up computed tomography automatic tube current modulation systems

被引:42
|
作者
Martin, C. J. [1 ]
Sookpeng, S. [2 ]
机构
[1] Univ Glasgow, Gartnavel Royal Hosp, Hlth Phys, Dept Clin Phys, Glasgow G12 0XH, Lanark, Scotland
[2] Naresuan Univ, Fac Allied Hlth Sci, Dept Radiol Technol, Phitsanulok 65000, Thailand
关键词
computed tomography; CT; automatic tube current modulation; CT AEC; CT optimization; diagnostic radiology; IMAGE NOISE; MULTIDETECTOR CT; DOSE REDUCTION; RADIATION-EXPOSURE; VOLTAGE SELECTION; PEDIATRIC CT; PATIENT SIZE; ABDOMINAL CT; PHANTOM; MANAGEMENT;
D O I
10.1088/0952-4746/36/3/R74
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Automatic tube current modulation (ATCM) on CT scanners can yield significant reductions in patient doses. Modulation is based on x-ray beam attenuation in body tissues obtained from scan projection radiographs (SPRs) and aims to maintain the same level of image quality throughout a scan. Noise level is important in judging image quality, but tissues in larger patients exhibit higher contrast resulting from the presence of fat. CT scanner manufacturers use different metrics to assess image quality. Some employ a simple measure of image noise, while others adopt a measure related to a reference image that accepts higher noise levels in more attenuating parts with higher contrast. At the present time there is no standard method for testing ATCM. This paper reviews the operation of different ATCM systems, considers options for testing, and sets out a framework that could be used for optimizing clinical protocols. If dose and image quality can be established for a reference phantom, the modulation performed by ATCM systems can be characterised using anatomical phantoms or geometrical elliptical phantoms which may be conical or include sections of varying dimension. For scanners using a reference image or mAs, selection of the image quality reference determines other factors. However, for scanners using a noise reference, a higher noise level should be selected for larger patients to avoid high doses, and the operator should ensure that appropriate limits are set for mA modulation. Other factors that need to be considered include the SPRs used to plan the ATCM and image thickness. Users should be aware of the mode of operation of the ATCM system on their CT scanner, and be familiar with the effects of changing different protocol parameters. The behaviour of ATCM systems should be established through testing of each CT scanner with suitable phantoms during commissioning.
引用
收藏
页码:R74 / R95
页数:22
相关论文
共 50 条
  • [1] Considerations on an automatic computed tomography tube current modulation system
    Moro, Luca
    Panizza, Denis
    D'Ambrosio, Daniela
    Carne, Irene
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (04) : 525 - 530
  • [2] Design and use of a phantom for testing and comparing the performance of computed tomography automatic tube current modulation systems
    Sookpeng, S.
    Martin, C. J.
    Krisanachinda, A.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (03) : 753 - 773
  • [3] Design and use of a phantom for testing and comparing the performance of computed tomography automatic tube current modulation systems
    Sookpeng, S.
    Martin, C.J.
    Krisanachinda, A.
    Sookpeng, S. (supawitoos@nu.ac.th), 1600, Institute of Physics Publishing (40): : 753 - 773
  • [4] Dosimetric changes with computed tomography automatic tube-current modulation techniques
    Spampinato S.
    Gueli A.M.
    Milone P.
    Raffaele L.A.
    Radiological Physics and Technology, 2018, 11 (2) : 184 - 191
  • [5] Effects of tube potential selection together with computed tomography automatic tube current modulation on CT imaging performance
    Sookpeng, S.
    Martin, C. J.
    Krisanachinda, A.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) : 809 - 831
  • [6] Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
    Yurt, Aysegul
    Ozsoykal, Ismail
    Obuz, Funda
    MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2019, 28 (03) : 96 - 103
  • [7] Size-based protocol optimization using automatic tube current modulation and automatic kV selection in computed tomography
    MacDougall, Robert D.
    Kleinman, Patricia L.
    Callahan, Michael J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 328 - 341
  • [8] Risk-minimizing tube current modulation for computed tomography
    Kachelriess, Marc
    RADIOLOGIE, 2023, 63 (07): : 523 - 529
  • [9] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [10] Image quality and clinical usefulness of automatic tube current modulation technology in female chest computed tomography screening
    Li, Cheng
    Qi, Lin
    Zhang, Yusheng
    Gao, Feng
    Jin, Xiu
    Zhang, Lukai
    Tang, Huan
    Li, Ming
    MEDICINE, 2020, 99 (33) : E21719