Setting up computed tomography automatic tube current modulation systems

被引:42
|
作者
Martin, C. J. [1 ]
Sookpeng, S. [2 ]
机构
[1] Univ Glasgow, Gartnavel Royal Hosp, Hlth Phys, Dept Clin Phys, Glasgow G12 0XH, Lanark, Scotland
[2] Naresuan Univ, Fac Allied Hlth Sci, Dept Radiol Technol, Phitsanulok 65000, Thailand
关键词
computed tomography; CT; automatic tube current modulation; CT AEC; CT optimization; diagnostic radiology; IMAGE NOISE; MULTIDETECTOR CT; DOSE REDUCTION; RADIATION-EXPOSURE; VOLTAGE SELECTION; PEDIATRIC CT; PATIENT SIZE; ABDOMINAL CT; PHANTOM; MANAGEMENT;
D O I
10.1088/0952-4746/36/3/R74
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Automatic tube current modulation (ATCM) on CT scanners can yield significant reductions in patient doses. Modulation is based on x-ray beam attenuation in body tissues obtained from scan projection radiographs (SPRs) and aims to maintain the same level of image quality throughout a scan. Noise level is important in judging image quality, but tissues in larger patients exhibit higher contrast resulting from the presence of fat. CT scanner manufacturers use different metrics to assess image quality. Some employ a simple measure of image noise, while others adopt a measure related to a reference image that accepts higher noise levels in more attenuating parts with higher contrast. At the present time there is no standard method for testing ATCM. This paper reviews the operation of different ATCM systems, considers options for testing, and sets out a framework that could be used for optimizing clinical protocols. If dose and image quality can be established for a reference phantom, the modulation performed by ATCM systems can be characterised using anatomical phantoms or geometrical elliptical phantoms which may be conical or include sections of varying dimension. For scanners using a reference image or mAs, selection of the image quality reference determines other factors. However, for scanners using a noise reference, a higher noise level should be selected for larger patients to avoid high doses, and the operator should ensure that appropriate limits are set for mA modulation. Other factors that need to be considered include the SPRs used to plan the ATCM and image thickness. Users should be aware of the mode of operation of the ATCM system on their CT scanner, and be familiar with the effects of changing different protocol parameters. The behaviour of ATCM systems should be established through testing of each CT scanner with suitable phantoms during commissioning.
引用
收藏
页码:R74 / R95
页数:22
相关论文
共 50 条
  • [21] Effects of scout radiographic imaging conditions on tube current modulation in chest computed tomography
    Inoue, Yusuke
    Itoh, Hiroyasu
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (01) : 253 - 269
  • [22] Signal-to-noise ratio and dose to the lens of the eye for computed tomography examination of the brain using an automatic tube current modulation system
    Sookpeng S.
    Butdee C.
    Emergency Radiology, 2017, 24 (3) : 233 - 239
  • [23] Estimating organ dose in computed tomography using tube current modulation: A Monte Carlo simulation
    Hosseinzadeh, V
    Ghaffari, H.
    Rezaeyan, A.
    Deilami, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 575 - 581
  • [24] Risikominimierende Röhrenstrommodulation in der ComputertomographieRisk-minimizing tube current modulation for computed tomography
    Marc Kachelrieß
    Die Radiologie, 2023, 63 (7) : 523 - 529
  • [25] Comparative Study of Volume Computed Tomography Dose Index and Size-Specific Dose Estimate Head in Computed Tomography Examination for Adult Patients Based on the Mode of Automatic Tube Current Modulation
    Xu, Jian
    He, Xiaolong
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 71 - 76
  • [26] Effect of tube current on computed tomography radiomic features
    Dennis Mackin
    Rachel Ger
    Cristina Dodge
    Xenia Fave
    Pai-Chun Chi
    Lifei Zhang
    Jinzhong Yang
    Steve Bache
    Charles Dodge
    A. Kyle Jones
    Laurence Court
    Scientific Reports, 8
  • [27] Effect of tube current on computed tomography radiomic features
    Mackin, Dennis
    Ger, Rachel
    Dodge, Cristina
    Fave, Xenia
    Chi, Pai-Chun
    Zhang, Lifei
    Yang, Jinzhong
    Bache, Steve
    Dodge, Charles
    Jones, A. Kyle
    Court, Laurence
    SCIENTIFIC REPORTS, 2018, 8
  • [28] Impact of tube current modulation on lesion conspicuity index in Hi-Resolution Chest Computed Tomography
    Szczepura, Katy
    Tomkinson, David
    Manning, David
    MEDICAL IMAGING 2017: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2017, 10136
  • [29] Organ-based tube current modulation and bismuth eye shielding in pediatric head computed tomography
    Stefan Markart
    Tim S. Fischer
    Simon Wildermuth
    Tobias J. Dietrich
    Hatem Alkadhi
    Sebastian Leschka
    Sabine Guesewell
    Michael R. Ditchfield
    Ismail Oezden
    Gerd Lutters
    Stephan L. Waelti
    Pediatric Radiology, 2022, 52 : 2584 - 2594
  • [30] Improving linac integrated cone beam computed tomography image quality using tube current modulation
    Stankovic, Uros
    Ploeger, Lennert S.
    Sonke, Jan-Jakob
    MEDICAL PHYSICS, 2021, 48 (04) : 1739 - 1749