Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study

被引:24
|
作者
Herrero-Climent, Mariano [1 ]
Lemos, Bernardo Ferreira [1 ,2 ]
Herrero-Climent, Federico [3 ]
Falcao, Carlos [1 ,2 ]
Oliveira, Helder [2 ]
Herrera, Manuela [4 ]
Javier Gil, Francisco [5 ]
Rios-Carrasco, Blanca [6 ]
Rios-Santos, Jose-Vicente [6 ]
机构
[1] Porto Dent Inst, P-4150518 Porto, Portugal
[2] Fernando Pessoa Univ, Fac Hlth Sci, P-4249004 Porto, Portugal
[3] Univ Complutense Madrid, Dept Periodontol, Madrid 28040, Spain
[4] Univ Seville, Dept Stomatol, Seville 41009, Spain
[5] Univ Int Catalunya, Fac Dent, Barcelona 08017, Spain
[6] Univ Seville, Dept Periodontol, Seville 41009, Spain
关键词
dental implant; tapered; under-preparation; implant preparation; implant stability; insertion torque; ISQ; RFA; RESONANCE FREQUENCY-ANALYSIS; LOW-DENSITY BONE; DENTAL IMPLANTS; SURGICAL TECHNIQUE; SURFACE-ROUGHNESS; STEPPED OSTEOTOMY; INSERTION TORQUE; HUMAN CADAVERS; OSSEOINTEGRATION; RELIABILITY;
D O I
10.3390/ijerph17124436
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this study was to evaluate the effects of different implant sites an under-preparation sequence associated with two different implant designs on implant primary stability measured by two parameters: insertion torque (IT) and implant stability quotient (ISQ). It used two different implants: one cylindrical as a control and another one with a tapered design. The implants were inserted in type III fresh humid bovine bone and four drilling sequences were used: one control, the one proposed by the implant company (P1), and three different undersized (P2, P3 and P4). P2 was the same as P1 without the cortical drill, P3 was without the last pilot drill and P4 was without both of them. The sample size wasn= 40 for each of the eight groups. Final IT was measured with a torquemeter and the ISQ was measured with Penguin resonance frequency analysis. Results showed that both ISQ and IT have a tendency to increase as the preparation technique reduces the implant site diameter when compared with the standard preparation, P1. The preparations without cortical drill, P2 and P4, showed the best results when compared with the ones with a cortical drill. Tapered implants always showed higher or the same ISQ and IT values when compared with the cylindrical implants. Giving the limitations of this study, it can be concluded that reducing implant preparation can increase IT and ISQ values. Removing the cortical drill and the use of a tapered design implant are also effective methods of increasing primary implant stability.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 50 条
  • [31] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Anna Takács
    Gyula Marada
    Kinga Turzó
    Ákos Nagy
    Orsolya Németh
    Eitan Mijiritsky
    Márton Kivovics
    Attila Mühl
    BMC Oral Health, 23
  • [32] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Takacs, Anna
    Marada, Gyula
    Turzo, Kinga
    Nagy, Akos
    Nemeth, Orsolya
    Mijiritsky, Eitan
    Kivovics, Marton
    Muhl, Attila
    BMC ORAL HEALTH, 2023, 23 (01)
  • [33] In vitro evaluation of the type of implant bed preparation with osteotomes in bone type IV and its influence on the stability of two implant systems
    Garcia-Vives, Nuria
    Andres-Garcia, Rodrigo
    Rios-Santos, Vicente
    Fernandez-Palacin, Ana
    Bullon, Pedro
    Herrero-Climent, Mariano
    Herrero-Climent, Federico
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (09): : E455 - E460
  • [34] In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems
    Andres-Garcia, Rodrigo
    Garcia Vives, Nuria
    Herrero Climent, Federico
    Fernandez Palacin, Ana
    Rios Santos, Vicente
    Herrero Climent, Mariano
    Bullon, Pedro
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (02): : E93 - E97
  • [35] Influence of the Implant Drill Design and Sequence on Temperature Changes During Site Preparation
    Sannino, Gianpaolo
    Cappare, Paolo
    Gherlone, Enrico F.
    Barlattani, Alberto
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (02) : 351 - 358
  • [36] Peri-Implant Bone Damage Procured by Piezoelectric and Conventional Implant Site Preparation: An In Vitro Comparison
    Rebaudi, Alberto
    Rebaudi, Federico
    Barberis, Fabrizio
    Sammartino, Gilberto
    Marenzi, Gaetano
    APPLIED SCIENCES-BASEL, 2020, 10 (24): : 1 - 9
  • [37] Primary Stability of Trabecular Metal Implant in Comparison to Fully Threaded Implants: In Vitro Study Simulating Immediate Implant Placement
    Magic, Marko
    Zupanek, Gorana
    Lazic, Zoran
    El Chaar, Edgard
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (06) : 584 - 589
  • [38] The effect of implant length and diameter on primary stability of tilted implant on D4 bone density: An in vitro study
    Baihaqi, Muhammad, I
    Sumarsongko, Taufik
    Bonifacius, Setyawan
    JOURNAL OF INTERNATIONAL ORAL HEALTH, 2022, 14 (05): : 487 - 493
  • [39] Influence of the dental implant macrogeometry and threads design on primary stability: an in vitro simulation on artificial bone blocks
    Comuzzi, Luca
    Tumedei, Margherita
    De Angelis, Francesco
    Lorusso, Felice
    Piattelli, Adriano
    Iezzi, Giovanna
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (11) : 1242 - 1250
  • [40] Accuracy of implant site preparation in robotic navigated dental implant surgery
    Linn, Thu Ya
    Salamanca, Eisner
    Aung, Lwin Moe
    Huang, Ta-Ko
    Wu, Yi-Fan
    Chang, Wei-Jen
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2023, 25 (05) : 881 - 891