System Architecture of the LabPET Small Animal PET Scanner

被引:32
|
作者
Tetrault, Mare-Andre [1 ]
Viscogliosi, Nicolas [1 ]
Riendeau, Joel [1 ]
Belanger, Francois [1 ]
Michaud, Jean-Baptiste [1 ]
Semmaoui, Hicham [1 ]
Berard, Philippe [2 ]
Lemieux, Francois [1 ]
Arpin, Louis [1 ]
Bergeron, Melanie [2 ]
Cadorette, Jules [2 ]
Pepin, Catherine M. [2 ]
Robert, Ghislain [3 ]
Lepage, Martin D. [3 ]
Lecomte, Roger [2 ]
Fontaine, Rejean [1 ]
机构
[1] Univ Sherbrooke, Dept Elect & Comp Engn, Sherbrooke, PQ J1K 2R1, Canada
[2] Univ Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
[3] Gamma Med Ideas Inc, Sherbrooke, PQ J1K 2R1, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
Positron Emission Tomography (PET);
D O I
10.1109/TNS.2008.2002326
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
To address modern molecular imaging requirements, a digital positron emission tomography (PET) scanner for small animals has been developed at Universite de Sherbrooke. Based on individual readout of avalanche photodiodes (APD) coupled to LYSO/LGSO phoswich detectors, the scanner supports up to 4608 channels in a 16.2 cm diameter, 11.25 cm axial field of view with an isotropic similar to 1.2 mm FWHM intrinsic spatial resolution at the center of the field of view. Custom data acquisition boards preprocess and sample APD signals at 45 MHz and compute in real time crystal identification, energy and timing information of detected events at an average sustained rate of up to 1250 raw counts per second per mm(2) (10 000 cps/channel). Real time digital signal analysis also filters out events outside the pre-selected energy window with crystal granularity to eliminate Compton events and electronic noise. Retained events are then merged into a single stream through a real-time sorting tree, at which end prompt and delayed coincidences are extracted. A single Firewire link handles both control and data transfers with a host computer. The LabPET features four data recording modes, giving the user the choice to retain data for research or to minimize file size for high coincidence count rate and imaging purposes. The electronic system also supports time synchronized data insertion for flags such as vital signs used in gated image reconstruction. Aside from data acquisition, hardware can generate live energy and discrimination spectra suitable for fast, automatic channel calibration.
引用
收藏
页码:2546 / 2550
页数:5
相关论文
共 50 条
  • [31] Features of the NIH atlas small animal PET scanner and its use with a coaxial small animal volume CT scanner
    Seidel, J
    Vaquero, JJ
    Pascau, J
    Desco, M
    Johnson, CA
    Green, MV
    2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, 2002, : 545 - 548
  • [32] The ClearPET™ LSO/LuYAP phoswich scanner:: A high performance small animal PET system
    Ziemons, K
    Auffray, E
    Barbier, R
    Brandenburg, G
    Bruyndonckx, P
    Choi, Y
    Christ, D
    Costes, N
    Declais, Y
    Devroede, O
    Dujardin, C
    Fedorov, AA
    Heinrichs, U
    Hollendung, A
    Korjhik, MJ
    Krieguer, M
    Kuntner, C
    Largeron, G
    Lartizien, C
    Larue, H
    Lecoq, P
    Leonard, S
    Marteau, J
    Morel, C
    Mosset, JB
    Parl, C
    Pedrini, C
    Petrosyan, AG
    Pietrzyk, U
    Rey, M
    Saladino, S
    Sappey-Marinier, D
    Simon, L
    Streun, M
    Tavernier, S
    Vieira, JM
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1728 - 1732
  • [33] System Architecture of a Fully Combined PET/CT Scanner using LabPET™ Electronics with an Upgraded Analog Front-end Optimized for PET and CT Counting Mode Operation
    Traore, Mohamed Seydou
    Thibaudeau, Christian
    Koua, Konin
    Tetrault, Marc-Andre
    Pepin, Catherine M.
    Cadorette, Jules
    Pratte, Jean-Francois
    Lecomte, Roger
    Fontaine, Rejean
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [34] Performance evaluation of the Philips MOSAIC small animal PET scanner
    Huisman, Marc C.
    Reder, Sybille
    Weber, Axel W.
    Ziegler, Sibylle I.
    Schwaiger, Markus
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (04) : 532 - 540
  • [35] Attenuation correction for the NIH ATLAS small animal PET scanner
    Yao, RT
    Seidel, J
    Liow, JS
    Green, MV
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) : 664 - 668
  • [36] Analog processing board for the MiCES small animal PET scanner
    Park, B
    Lewellen, TK
    Miyaoka, RS
    Janes, M
    Kinahan, P
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3920 - 3923
  • [37] Development of a small animal PET scanner using DOI detectors
    Yamada, R.
    Watanabe, M.
    Omura, T.
    Sato, N.
    Shimizu, K.
    Takahashi, M.
    Ote, K.
    Katabe, A.
    Moriya, T.
    Sakai, K.
    Yamashita, T.
    Tanaka, E.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) : 906 - 911
  • [38] A small animal PET scanner with millimeter spatial resolution.
    Del Guerra, A
    Di Domenico, G
    Scandola, M
    Zavattini, G
    Giganti, M
    Uccelli, L
    Piffanelli, A
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 133P - 133P
  • [39] Raytest ClearPET™, a new generation small animal PET scanner
    Roldan, P. Sempere
    Chereul, E.
    Dietzel, O.
    Magnier, L.
    Pautrot, C.
    Rbah, L.
    Sappey-Marinier, D.
    Wagner, A.
    Zimmer, L.
    Janier, M.
    Tarazona, V.
    Dietzel, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 498 - 501
  • [40] Performance evaluation of the Philips MOSAIC small animal PET scanner
    Marc C. Huisman
    Sybille Reder
    Axel W. Weber
    Sibylle I. Ziegler
    Markus Schwaiger
    European Journal of Nuclear Medicine and Molecular Imaging, 2007, 34 : 532 - 540