System Architecture of the LabPET Small Animal PET Scanner

被引:32
|
作者
Tetrault, Mare-Andre [1 ]
Viscogliosi, Nicolas [1 ]
Riendeau, Joel [1 ]
Belanger, Francois [1 ]
Michaud, Jean-Baptiste [1 ]
Semmaoui, Hicham [1 ]
Berard, Philippe [2 ]
Lemieux, Francois [1 ]
Arpin, Louis [1 ]
Bergeron, Melanie [2 ]
Cadorette, Jules [2 ]
Pepin, Catherine M. [2 ]
Robert, Ghislain [3 ]
Lepage, Martin D. [3 ]
Lecomte, Roger [2 ]
Fontaine, Rejean [1 ]
机构
[1] Univ Sherbrooke, Dept Elect & Comp Engn, Sherbrooke, PQ J1K 2R1, Canada
[2] Univ Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
[3] Gamma Med Ideas Inc, Sherbrooke, PQ J1K 2R1, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
Positron Emission Tomography (PET);
D O I
10.1109/TNS.2008.2002326
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
To address modern molecular imaging requirements, a digital positron emission tomography (PET) scanner for small animals has been developed at Universite de Sherbrooke. Based on individual readout of avalanche photodiodes (APD) coupled to LYSO/LGSO phoswich detectors, the scanner supports up to 4608 channels in a 16.2 cm diameter, 11.25 cm axial field of view with an isotropic similar to 1.2 mm FWHM intrinsic spatial resolution at the center of the field of view. Custom data acquisition boards preprocess and sample APD signals at 45 MHz and compute in real time crystal identification, energy and timing information of detected events at an average sustained rate of up to 1250 raw counts per second per mm(2) (10 000 cps/channel). Real time digital signal analysis also filters out events outside the pre-selected energy window with crystal granularity to eliminate Compton events and electronic noise. Retained events are then merged into a single stream through a real-time sorting tree, at which end prompt and delayed coincidences are extracted. A single Firewire link handles both control and data transfers with a host computer. The LabPET features four data recording modes, giving the user the choice to retain data for research or to minimize file size for high coincidence count rate and imaging purposes. The electronic system also supports time synchronized data insertion for flags such as vital signs used in gated image reconstruction. Aside from data acquisition, hardware can generate live energy and discrimination spectra suitable for fast, automatic channel calibration.
引用
收藏
页码:2546 / 2550
页数:5
相关论文
共 50 条
  • [21] Efficient system modeling for a small animal PET scanner with tapered DOI detectors
    Zhang, Mengxi
    Zhou, Jian
    Yang, Yongfeng
    Rodriguez-Villafuerte, Mercedes
    Qi, Jinyi
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 461 - 474
  • [22] Initial evaluation of the Indiana small animal PET scanner
    Rouze, Ned C.
    Soon, Victor C.
    Young, John W.
    Siegel, Stefan
    Hutchins, Gary D.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2394 - 2398
  • [23] Reconstructed Image Resolution of the LabPET Small Animal Scanner Measured Using Simulated Partial Volume Effects
    Dumouchel, Tyler
    Selivanov, Vitali V.
    Cadorette, Jules
    Lecomte, Roger
    deKemp, Robert A.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2689 - 2695
  • [24] Design of a small animal MR compatible PET scanner
    Slates, R
    Cherry, S
    Boutefnouchet, A
    Shao, YP
    Dahlbom, M
    Farahani, K
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) : 565 - 570
  • [25] Design of a small animal MR compatible PET scanner
    UCLA Sch of Medicine, Los Angeles, United States
    IEEE Trans Nucl Sci, 3 II (565-570):
  • [26] Design of a small animal MR compatible PET scanner
    Slates, Randal
    Cherry, Simon
    Boutefnouchet, Abdel
    Shao, Yiping
    Dahlbom, Magnus
    Farahani, Keyvan
    IEEE Nuclear Science Symposium and Medical Imaging Conference, 1999, 2 : 1268 - 1273
  • [27] Performance Evaluation of the LabPET APD-Based Digital PET Scanner
    Bergeron, Melanie
    Cadorette, Jules
    Beaudoin, Jean-Francois
    Lepage, Martin D.
    Robert, Ghislain
    Selivanov, Vitali
    Tetrault, Marc-Andre
    Viscogliosi, Nicolas
    Norenberg, Jeffrey P.
    Fontaine, Rejean
    Lecomte, Roger
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) : 10 - 16
  • [28] Imaging Performance of the LabPET™ APD-based Digital PET Scanner
    Bergeron, Melanie
    Cadorette, Jules
    Beaudoin, Jean-Francois
    Tetrault, Marc-Andre
    Viscogliosi, Nicolas
    Selivanov, Vitali
    Lepage, Martin D.
    Robert, Ghislain
    Norenberg, Jeffrey P.
    Fontaine, Rejean
    Lecomte, Roger
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3116 - +
  • [29] Performance evaluation of the LabPET™ APD-based digital PET scanner
    Bergeron, Melanie
    Cadorette, Jules
    Beaudoin, Jean-Francois
    Rousseau, Jacques A.
    Dumoulin, Marc
    Lepage, Martin D.
    Robert, Ghislain
    Selivanov, Vitali
    Tetrault, Marc-Andre
    Viscogliosi, Nicolas
    Dumouchel, Tyler
    Thorn, Stephanie
    DaSilva, Jean
    deKemp, Robert A.
    Norenberg, Jeffrey P.
    Fontaine, Rejean
    Lecomte, Roger
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4185 - +
  • [30] Predicting Small Lesion Detectability for a Small Animal TOF PET Scanner
    Yussefian, Nikta Zarif
    Gaudin, Emilie
    Lecomte, Roger
    Fontaine, Rejean
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (05) : 601 - 608