Characteristics of highly cross-linked polyethylene wear debris in vivo

被引:29
|
作者
Baxter, Ryan M. [1 ]
MacDonald, Daniel W. [1 ]
Kurtz, Steven M. [1 ,3 ]
Steinbeck, Marla J. [1 ,2 ]
机构
[1] Drexel Univ, Sch Biomed Engn Sci & Hlth Syst, Philadelphia, PA 19104 USA
[2] Drexel Univ, Coll Drexel Med, Philadelphia, PA 19104 USA
[3] Exponent Inc, Philadelphia, PA USA
关键词
total hip replacementhighly cross-linked polyethyleneosteolysiswear debrisbiological activity; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL HIP-ARTHROPLASTY; FUNCTIONAL BIOLOGICAL-ACTIVITY; CONVENTIONAL POLYETHYLENE; UHMWPE PARTICLES; PERIPROSTHETIC TISSUE; MECHANICAL-PROPERTIES; 5-YEAR EXPERIENCE; GAMMA-RADIATION; EARLY FAILURE;
D O I
10.1002/jbm.b.32902
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Despite the widespread implementation of highly cross-linked polyethylene (HXLPE) liners to reduce the clinical incidence of osteolysis, it is not known if the improved wear resistance will outweigh the inflammatory potential of HXLPE wear debris generated in vivo. Thus, we asked: What are the differences in size, shape, number, and biological activity of polyethylene wear particles obtained from primary total hip arthroplasty revision surgery of conventional polyethylene (CPE) versus remelted or annealed HXLPE liners? Pseudocapsular tissue samples were collected from revision surgery of CPE and HXLPE (annealed and remelted) liners, and digested using nitric acid. The isolated polyethylene wear particles were evaluated using scanning electron microscopy. Tissues from both HXLPE cohorts contained an increased percentage of submicron particles compared to the CPE cohort. However, the total number of particles was lower for both HXLPE cohorts, as a result there was no significant difference in the volume fraction distribution and specific biological activity (SBA; the relative biological activity per unit volume) between cohorts. In contrast, based on the decreased size and number of HXLPE wear debris there was a significant decrease in total particle volume (mm3/g of tissue). Accordingly, when the SBA was normalized by total particle volume (mm3/gm tissue) or by component wear volume rate (mm3/year), functional biological activity of the HXLPE wear debris was significantly decreased compared to the CPE cohort. Indications for this study are that the osteolytic potential of wear debris generated by HXLPE liners in vivo is significantly reduced by improvements in polyethylene wear resistance. (c) 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 101B: 467475, 2013.
引用
收藏
页码:467 / 475
页数:9
相关论文
共 50 条
  • [41] Vitamin E-infused highly cross-linked polyethylene did not reduce the number of in vivo wear particles in total knee arthroplasty
    Orita, K.
    Minoda, Y.
    Sugama, R.
    Ohta, Y.
    Ueyama, H.
    Takemura, S.
    Nakamura, H.
    BONE & JOINT JOURNAL, 2020, 102B (11): : 1527 - 1534
  • [42] In Vivo Oxidation and Wear in Remelted Highly Cross-Linked Polyethylene Liners Retrieved at a Minimum of 10 Years After Total Hip Arthroplasty
    Nishimura, Manase
    Nakamura, Takuya
    Nishida, Hideji
    Misaki, Tomonori
    Aikawa, Takao
    JOURNAL OF ARTHROPLASTY, 2024, 39 (12): : 3092 - 3096
  • [43] The effect of neutron radiation on conventional and highly cross-linked ultrahigh-molecular-weight polyethylene wear
    Markel, David C.
    Mendelson, Stephen D.
    Yudelev, Mark
    Essner, Aaron
    Yau, Shi-Shen
    Wang, Aiguo
    JOURNAL OF ARTHROPLASTY, 2008, 23 (05): : 732 - 735
  • [44] Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty
    Samujh, Christopher
    Bhimani, Samrath
    Smith, Langan
    Malkani, Arthur L.
    ORTHOPEDICS, 2016, 39 (06) : E1178 - E1182
  • [45] Clinical Safety and Wear Resistance of the Phospholipid Polymer-Grafted Highly Cross-Linked Polyethylene Liner
    Moro, Toru
    Takatori, Yoshio
    Tanaka, Sakae
    Ishihara, Kazuhiko
    Oda, Hiromi
    Kim, Yoon Taek
    Umeyama, Takashige
    Fukatani, Eisei
    Ito, Hideya
    Kyomoto, Masayuki
    Oshima, Hirofumi
    Tanaka, Takeyuki
    Kawaguchi, Hiroshi
    Nakamura, Kozo
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (09) : 2007 - 2016
  • [46] In Vivo Oxidation in Remelted Highly Cross-Linked Retrievals
    Currier, B. H.
    Van Citters, D. W.
    Currier, J. H.
    Collier, J. P.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (14): : 2409 - 2418
  • [47] Generalized size and shape description of UHMWPE wear debris - A comparison of cross-linked, enhanced fused, and standard polyethylene particles
    Sprecher, CM
    Schneider, E
    Wimmer, MA
    CROSSLINKED AND THERMALLLY TREATED ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE FOR JOINT REPLACEMENTS, 2004, 1445 : 3 - 15
  • [48] Highly cross-linked polyethylene - A new option in tribology of THA
    Rabenseifner, L
    BIOCERAMICS IN JOINT ARTHROPLASTY, 2004, : 83 - 87
  • [49] The Use of Highly Cross-linked Polyethylene in Total Knee Arthroplasty
    Lachiewicz, Paul F.
    Geyer, Mark R.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2011, 19 (03) : 143 - 151
  • [50] Aggressive wear testing of a cross-linked polyethylene in total knee arthroplasty
    Muratoglu, OK
    Bragdon, CR
    O'Connor, DO
    Perinchief, RS
    Jasty, M
    Harris, WH
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (404) : 89 - 95