Automatic localization of vertebral levels in x-ray fluoroscopy using 3D-2D registration: a tool to reduce wrong-site surgery

被引:60
|
作者
Otake, Y. [1 ,2 ]
Schafer, S. [1 ]
Stayman, J. W. [1 ]
Zbijewski, W. [1 ]
Kleinszig, G. [3 ]
Graumann, R. [3 ]
Khanna, A. J. [4 ]
Siewerdsen, J. H. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
[3] Siemens Healthcare XP Div, Erlangen, Germany
[4] Johns Hopkins Med Inst, Dept Orthopaed Surg, Baltimore, MD 21287 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 17期
基金
美国国家卫生研究院;
关键词
CMA EVOLUTION STRATEGY; IMAGE REGISTRATION; 2D/3D REGISTRATION; 2D-3D REGISTRATION; INTRAOPERATIVE LOCALIZATION; SCREW PLACEMENT; CT; INTENSITY; ARM; RECONSTRUCTION;
D O I
10.1088/0031-9155/57/17/5485
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Surgical targeting of the incorrect vertebral level (wrong-level surgery) is among the more common wrong-site surgical errors, attributed primarily to the lack of uniquely identifiable radiographic landmarks in the mid-thoracic spine. The conventional localization method involves manual counting of vertebral bodies under fluoroscopy, is prone to human error and carries additional time and dose. We propose an image registration and visualization system (referred to as LevelCheck), for decision support in spine surgery by automatically labeling vertebral levels in fluoroscopy using a GPU-accelerated, intensity-based 3D-2D (namely CT-to-fluoroscopy) registration. Agradient information (GI) similarity metric and a CMA-ES optimizer were chosen due to their robustness and inherent suitability for parallelization. Simulation studies involved ten patient CT datasets from which 50 000 simulated fluoroscopic images were generated from C-arm poses selected to approximate the C-arm operator and positioning variability. Physical experiments used an anthropomorphic chest phantom imaged under real fluoroscopy. The registration accuracy was evaluated as the mean projection distance (mPD) between the estimated and true center of vertebral levels. Trials were defined as successful if the estimated position was within the projection of the vertebral body (namely mPD <5 mm). Simulation studies showed a success rate of 99.998% (1 failure in 50 000 trials) and computation time of 4.7 s on a midrange GPU. Analysis of failure modes identified cases of false local optima in the search space arising from longitudinal periodicity in vertebral structures. Physical experiments demonstrated the robustness of the algorithm against quantum noise and x-ray scatter. The ability to automatically localize target anatomy in fluoroscopy in near-real-time could be valuable in reducing the occurrence of wrong-site surgery while helping to reduce radiation exposure. The method is applicable beyond the specific case of vertebral labeling, since any structure defined in pre-operative (or intra-operative) CT or cone-beam CT can be automatically registered to the fluoroscopic scene.
引用
收藏
页码:5485 / 5508
页数:24
相关论文
共 50 条
  • [41] A novel approach to 2D/3D registration of X-ray images using Grangeat's relation
    Frysch, Robert
    Pfeiffer, Tim
    Rose, Georg
    MEDICAL IMAGE ANALYSIS, 2021, 67
  • [42] A novel approach to 2D/3D registration of X-ray images using Grangeat's relation
    Frysch, Robert
    Pfeiffer, Tim
    Rose, Georg
    Medical Image Analysis, 2021, 67
  • [43] Intraoperative 2D/3D Image Registration via Differentiable X-ray Rendering
    Gopalakrishnan, Vivek
    Dey, Neel
    Golland, Polina
    2024 IEEE/CVF CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR), 2024, : 11662 - 11672
  • [44] Bundling 3D-and 2D-based registration of MRI to X-ray breast tomosynthesis
    Smole, P. Cotic
    Ruiter, N. V.
    Kaiser, C.
    Krammer, J.
    Hopp, T.
    MEDICAL IMAGING 2018: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2018, 10576
  • [45] Effect of segmentation errors on 3D-to-2D registration of implant models in X-ray images
    Mahfouz, MR
    Hoff, WA
    Komistek, RD
    Dennis, DA
    JOURNAL OF BIOMECHANICS, 2005, 38 (02) : 229 - 239
  • [46] An approach to 2D/3D registration of a vertebra in 2D x-ray fluoroscopies with 3D CT images
    Weese, J
    Buzug, TM
    Lorenz, C
    Fassnacht, C
    CVRMED-MRCAS'97: FIRST JOINT CONFERENCE - COMPUTER VISION, VIRTUAL REALITY AND ROBOTICS IN MEDICINE AND MEDICAL ROBOTICS AND COMPUTER-ASSISTED SURGERY, 1997, 1205 : 119 - 128
  • [47] Gradient-based registration of 3D MR and 2D x-ray images
    Tomazevic, D
    Likar, B
    Pernus, F
    CARS 2001: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2001, 1230 : 327 - 332
  • [48] Aortic 3D Deformation Reconstruction using 2D X-ray Fluoroscopy and 3D Pre-operative Data for Endovascular Interventions
    Zhang, Yanhao
    Zhao, Liang
    Huang, Shoudong
    2020 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2020, : 2393 - 2399
  • [49] 3D registration of MR and X-ray spine images using an articulated model
    Harmouche, Rola
    Cheriet, Farida
    Labelle, Hubert
    Dansereau, Jean
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2012, 36 (05) : 410 - 418
  • [50] SGReg: segmentation guided 3D/2D rigid registration for orthogonal X-ray and CT images in spine surgery navigation
    Mi, Jia
    Yin, Wenhao
    Zhao, Lei
    Chen, Yangfan
    Zhou, Yujia
    Feng, Qianjin
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (13):