Performance of an active fixation bipolar left ventricular lead vs passive fixation quadripolar leads in cardiac resynchronization therapy, a randomized trial

被引:4
|
作者
Keilegavlen, Havard [1 ,2 ]
Schuster, Peter [1 ,2 ]
Hovstad, Thomas [1 ]
Faerestrand, Svein [1 ,2 ]
机构
[1] Haukeland Hosp, Dept Heart Dis, Pb 1400, N-5021 Bergen, Norway
[2] Univ Bergen, Dept Clin Sci, Bergen, Norway
关键词
biventricular pacemaker; cardiac resynchronization therapy; heart failure; PLACEMENT; POSITION; DYSSYNCHRONY; IMPLANTATION; MORTALITY; SURVIVAL; SITES;
D O I
10.1002/joa3.12450
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Usage of active fixation bipolar left ventricular (LV) leads represents an alternative approach to the more commonly used passive fixation quadripolar leads in cardiac resynchronization therapy (CRT). We compared a bipolar LV lead with a side screw for active fixation and passive fixation quadripolar LV leads. Methods: Sixty-two patients were before CRT implantations randomly allocated to receive a bipolar (n = 31) or quadripolar (n = 31) LV leads. Speckle-tracking radial strain echocardiography was used to define the LV segment with latest mechanical activation as the target LV segment. The electrophysiological measurements and the capability to obtain a proximal position in a coronary vein placed over the target segment were assessed. Results: Upon implantation, the quadripolar lead demonstrated a lower pacing capture threshold than the bipolar lead, but at follow-up, there was no difference. There were no differences in the LV lead implant times or radiation doses. The success rate in reaching the target location was not significantly different between the two LV leads. Conclusions: The pacing capture thresholds were low, with no significant difference between active fixation bipolar leads and quadripolar leads. Active fixation leads did not promote a more proximal location of the stimulating electrode or a higher grade of concordance to the target segment than passive fixation leads.
引用
收藏
页码:212 / 218
页数:7
相关论文
共 50 条
  • [31] Hospitalization rates and associated cost analysis of cardiac resynchronization therapy with an implantable defibrillator and quadripolar vs. bipolar left ventricular leads: a comparative effectiveness study
    Forleo, Giovanni B.
    Di Biase, Luigi
    Bharmi, Rupinder
    Dalal, Nirav
    Panattoni, Germana
    Pollastrelli, Annalisa
    Tesauro, Manfredi
    Santini, Luca
    Natale, Andrea
    Romeo, Francesco
    EUROPACE, 2015, 17 (01): : 101 - 107
  • [32] Repetitive optimizing left ventricular pacing configurations with quadripolar leads improves response to cardiac resynchronization therapy A single-center randomized clinical trial
    Gu, Min
    Jin, Han
    Hua, Wei
    Fan, Xiao-Han
    Ding, Li-Gang
    Wang, Jing
    Niu, Hong-Xia
    Cai, Chi
    Zhang, Shu
    MEDICINE, 2017, 96 (37)
  • [33] Results of the European and Canadian safety and performance studies for the smallest passive and active fixation quadripolar defibrillator leads
    Kuepper, B.
    Della Bella, P.
    Schmitz, D.
    Workel, M.
    Janssen, J.
    Heuer, H.
    EUROPEAN HEART JOURNAL, 2005, 26 : 272 - 272
  • [34] Extraction of chronically implanted coronary sinus leads active fixation vs passive fixation leads
    Crossley, George H.
    Sorrentino, Robert A.
    Exner, Derek V.
    Merliss, Andrew D.
    Tobias, Serge M.
    Martin, David O.
    Augostini, Ralph
    Piccini, Jonathan P.
    Schaerf, Raymond
    Li, Shelby
    Miller, Clayton T.
    Adler, Stuart W.
    HEART RHYTHM, 2016, 13 (06) : 1253 - 1259
  • [35] Active fixation lead can improve clinical response to cardiac resynchronization therapy
    Mezzetti, M.
    Morelli, M.
    Caccamo, L.
    Casale, M.
    Busacca, P.
    EUROPEAN JOURNAL OF HEART FAILURE, 2018, 20 : 559 - 559
  • [36] Active fixation of a thin transvenous left-ventricular lead by a side helix facilitates targeted and stable placement in cardiac resynchronization therapy
    Keilegavlen, Havard
    Hovstad, Thomas
    Faerestrand, Svein
    EUROPACE, 2016, 18 (08): : 1235 - 1240
  • [37] Vector Selection of a Quadripolar Left Ventricular Pacing Lead Affects Acute Hemodynamic Response to Cardiac Resynchronization Therapy: A Randomized Cross-Over Trial
    Asbach, Stefan
    Hartmann, Maximilian
    Wengenmayer, Tobias
    Graf, Erika
    Bode, Christoph
    Biermann, Juergen
    PLOS ONE, 2013, 8 (06):
  • [38] Benefits of a Quadripolar Left Ventricular Lead in Patients Undergoing Cardiac Resynchronization Therapy with Underlying Myocardial Scar
    Hussain, Mohsin A.
    Bhamra-Ariza, Paul
    Jacques, Adam
    Wilkinson, Peter
    Odemuyiwa, Olusola
    Fluck, David
    Beeton, Ian
    Kaba, Riyaz A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (02): : e45 - e47
  • [39] Electrically vs. imaging-guided left ventricular lead placement in cardiac resynchronization therapy: a randomized controlled trial
    Stephansen, Charlotte
    Sommer, Anders
    Kronborg, Mads Brix
    Jensen, Jesper Moller
    Norgaard, Bjarne Linde
    Gerdes, Christian
    Kristensen, Jens
    Jensen, Henrik Kjaerulf
    Fyenbo, Daniel Benjamin
    Bouchelouche, Kirsten
    Nielsen, Jens Cosedis
    EUROPACE, 2019, 21 (09): : 1369 - 1377
  • [40] Effective non-apical left ventricular pacing with quadripolar leads for cardiac resynchronization therapy: a multicenter study
    Forleo, G. B.
    Panattoni, G.
    Solimene, F.
    Schillaci, V.
    Covino, G.
    Sassara, M.
    Savarese, G.
    Santini, L.
    Donzelli, S.
    Badolati, S.
    Gerosa, C.
    Valsecchi, S.
    Mangone, G.
    Sergi, D.
    EUROPEAN HEART JOURNAL, 2017, 38 : 1170 - 1170