Effect of Calibration and Environmental Condition on the Performance of Direct-Reading Organic Vapor Monitors

被引:10
|
作者
Coffey, Christopher [1 ]
LeBouf, Ryan [2 ]
Lee, Larry [3 ]
Slaven, James [3 ]
Martin, Stephen [2 ]
机构
[1] NIOSH, Natl Personal Protect Technol Lab, Morgantown, WV 26505 USA
[2] NIOSH, Div Resp Dis Studies, Morgantown, WV 26505 USA
[3] NIOSH, Hlth Effects Lab Div, Morgantown, WV 26505 USA
关键词
direct-reading organic vapor monitors; flame ionization detector; photoionization detector; SapphIRe; toxic vapor analyzer; IONIZATION GAS-CHROMATOGRAPH; PHOTOIONIZATION DETECTOR; ATMOSPHERES; EXPOSURE;
D O I
10.1080/15459624.2012.725015
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The performance of three MIRAN SapphIRe Portable Infrared Ambient Air Analyzers and three Century Portable Toxic Vapor Analyzers equipped with photoionization (PID) and flame ionization (FID) detectors was compared with charcoal tube sampling. Relationships were investigated using two different calibration methods at four cyclohexane concentrations, three temperatures, and four relative humidities. For the first method, the TVA monitors were calibrated with a single concentration of methane for the FID, and isobutylene for the PID. The SapphIRe monitors were zeroed and the monitor's manufacturer-supplied library was used. For the second method, a five-point cyclohexane calibration curve was created for each monitor Comparison of the monitor results of each calibration method (pooled data) indicated a significant difference between methods (t-test, p < 0.001), The SapphIRe group had results closer to the charcoal tubes with the second calibration method, while the PID and FID monitor groups performed better using the first calibration method. The PID monitor group's performance was affected only at the 90% relative humidity (RH) condition. Using the first method, the monitor readings were compared with the charcoal tube average using mixed linear model analyses of variance (ANOVAs) and regression. The ANOVA results showed there was a statistically significant difference among readings from all monitor types (p <0.0001). The regression results demonstrated that the SapphIRe (r(2) = 0.97) and FID (r(2) = 0.92) monitor groups correlated well with the charcoal tubes. The PID monitor group had a similar correlation when 90% RH was excluded (r(2) = 0.94) but had a weaker correlation when it was included (r(2) = 0.58). The operator should take care when using these monitors at high concentrations and the PID monitors at high humidities, consider the variability between units of the same monitor, and conduct performance verification of the monitor being used.
引用
收藏
页码:670 / 680
页数:11
相关论文
共 50 条
  • [31] EFFECT OF BODY CONDITION AND ENVIRONMENTAL-TEMPERATURE ON REPRODUCTIVE-PERFORMANCE IN THE SOW
    MEYEN, BA
    MABRY, JW
    RAMPACEK, GB
    JONES, RD
    KRAELING, RR
    GEORGIA AGRICULTURAL EXPERIMENT STATION SPECIAL PUBLICATION, 1985, (37): : 12 - 14
  • [32] Effect of working fluids on the performance of phase change material storage based direct vapor generation solar organic Rankine cycle system
    Alvi, Jahan Zeb
    Feng, Yongqiang
    Wang, Qian
    Imran, Muhammad
    Pei, Gang
    ENERGY REPORTS, 2021, 7 (07) : 348 - 361
  • [33] Effect of Organic Impurities on the Performance of Direct Formic Acid Fuel Cells
    Law, Wai Lung
    Platt, Alison M.
    Wimalaratne, Priyantha D. C.
    Blair, Sharon L.
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2009, 156 (05) : B553 - B557
  • [34] Analytical Performance Issues Direct Reading Particle Counters: Calibration Verification and Multiple Instrument Agreement via Bump Testing
    Jankovic, John
    Zontek, Tracy L.
    Ogle, Burton R.
    Hollenbeck, Scott
    JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2015, 12 (07) : D116 - D122
  • [35] Performance Assessment of Direct Vapor Generation Solar Organic Rankine Cycle System Coupled with Heat Storage
    Alvi, Jahan Zeb
    Yu Jinghu
    Feng, Yongqiang
    Asim, Muhammad
    Qian, Wang
    Pei, Gang
    SUSTAINABILITY, 2022, 14 (22)
  • [36] AN EVALUATION OF THE RESPONSE OF SOME PORTABLE, DIRECT-READING 10.2 EV AND 11.8 EV PHOTOIONIZATION DETECTORS, AND A FLAME IONIZATION GAS-CHROMATOGRAPH FOR ORGANIC VAPORS IN HIGH HUMIDITY ATMOSPHERES
    BARSKY, JB
    HEE, SSQ
    CLARK, CS
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1985, 46 (01): : 9 - 14
  • [37] Effect of organic solvent vapor treatment on transistor performance and contact resistance of copper phthalocyanine based organic field -effect transistors
    Kumar, Pankaj
    Yadav, Sarita
    Kumar, Naresh
    Kumar, Lokendra
    SOLID-STATE ELECTRONICS, 2021, 176
  • [38] Effect of Evaporation Temperature on the Performance of Organic Rankine Cycle in Near-Critical Condition
    Wang, Yuping
    Ding, Xiaoyi
    Tang, Lei
    Weng, Yiwu
    JOURNAL OF ENERGY RESOURCES TECHNOLOGY-TRANSACTIONS OF THE ASME, 2016, 138 (03):
  • [39] EVALUATION OF THE EFFECT OF VARIOUS SURFICIAL POLLUTANTS AND ENVIRONMENTAL CONDITION ON SURFACE FRICTION PERFORMANCE OF ROAD PAVEMENT
    Mehdi Khabiri, Mohammad
    Afkhamy Meybodi, Pooya
    Montazeri, Ali Mohammad
    SCIENTIFIC JOURNAL OF SILESIAN UNIVERSITY OF TECHNOLOGY-SERIES TRANSPORT, 2021, 112 : 99 - 111
  • [40] Effect of differential fertilisation treatments on maize hybrid quality and performance under environmental stress condition in Hungary
    Szeles, Adrienn
    Nagy, Janos
    Ratonyi, Tamas
    Harsanyi, Endre
    MAYDICA, 2019, 64 (02):