Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels

被引:90
|
作者
Alfa, Michelle J. [1 ,2 ,3 ]
Fatima, Iram [3 ]
Olson, Nancy [3 ]
机构
[1] Diagnost Serv Manitoba, Clin Microbiol, Winnipeg, MB R2H 2A6, Canada
[2] St Boniface Res Ctr, Microbiol Lab, Winnipeg, MB, Canada
[3] Univ Manitoba, Dept Med Microbiol, Winnipeg, MB, Canada
关键词
Simulated use; Endoscope channel; Cleaning-compliance; Benchmark; Relative light units; DECONTAMINATION; EFFICACY;
D O I
10.1016/j.ajic.2012.03.018
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Compliance with cleaning of flexible endoscope channels cannot be verified using visual inspection. Adenosine triphosphate (ATP) has been suggested as a possible rapid cleaning monitor for flexible endoscope channels. There have not been published validation studies to specify the level of ATP that indicates inadequate cleaning has been achieved. Objective: The objective of this study was to validate the Clean-Trace (3M Inc, St. Paul, MN) ATP water test method for monitoring manual cleaning of flexible endoscopes. Methods: This was a simulated use study using a duodenoscope as the test device. Artificial test soil containing 10(6) colony-forming units of Pseudomonas aeruginosa and Enterococcus faecalis was used to perfuse all channels. The flush sample method for the suction-biopsy (L1) or air-water channel (L2) using 40 and 20 mLs sterile reverse osmosis water, respectively, was validated. Residuals of ATP, protein, hemoglobin, and bioburden were quantitated from channel samples taken from uncleaned, partially cleaned, and fully cleaned duodenoscopes. The benchmarks for clean were as follows: <6.4 mu g/cm(2) protein, <2.2 mu g/cm(2) hemoglobin, and <4-log(10) colony-forming units/cm(2) bioburden. Results: The average ATP in clean channel samples was 27.7 RLUs and 154 RLUs for L1 and L2, respectively (<200 RLUs for all channels). The average protein, hemoglobin, and bioburden benchmarks were achieved if <200 RLUs were detected. If the channel sample was >200 RLUs, the residual organic and bioburden levels would exceed the acceptable benchmarks. Conclusion: Our data validated that flexible endoscopes that have complete manual cleaning will have <200 RLUs by the Clean-Trace ATP test. Copyright (C) 2013 by the Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 50 条
  • [31] Comparison of manual Steering and Steering via Joystick of a flexible Rhino Endoscope
    Eckl, R.
    Gumprecht, J. D. J.
    Strauss, G.
    Hofer, M.
    Dietz, A.
    Lueth, T. C.
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 1234 - 1237
  • [32] Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning
    Alfa, Michelle J.
    Olson, Nancy
    DeGagne, Pat
    AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (09) : 561 - 570
  • [33] Implementation of cleaning and evaluation process for mobile patient equipment using adenosine triphosphate
    Reese, Sara M.
    Knepper, Bryan C.
    Kurtz, Jennifer
    LeQuire, D. Christy
    Van Winks, Tina
    Bonn, Jennifer
    Young, Heather L.
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2019, 40 (07): : 798 - 800
  • [34] Monitoring the Effectiveness of Hospital Cleaning Practices by Use of an Adenosine Triphosphate Bioluminescence Assay
    Boyce, John M.
    Havill, Nancy L.
    Dumigan, Diane G.
    Golebiewski, Michael
    Balogun, Ola
    Rizvani, Ramo
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2009, 30 (07): : 678 - 684
  • [35] Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination
    van der Ploeg, Koen
    Severin, Juliette A.
    Vos, Margreet C.
    Erler, Nicole S.
    Bulkmans, Adriana J. C.
    Bruno, Marco
    Mason-Slingerland, Bibi C. G. C.
    ENDOSCOPY INTERNATIONAL OPEN, 2025, 13
  • [36] Assessment of endoscope reprocessing at World Gastroenterology Organisation training centers using adenosine triphosphate testing
    Visrodia, Kavel
    Jones, Dianne
    Jennings, Melissa
    Conway, Catherine
    Burgos, Herbert
    Malani, Jioji
    Ogutu, Elly
    Pausawasdi, Nonthalee
    Sabbagh, Luis
    Leddin, Desmond
    Kassmeyer, Blake
    Lennon, Ryan J.
    Topazian, Mark D.
    ENDOSCOPY INTERNATIONAL OPEN, 2024, 12 (01) : E68 - E77
  • [37] Study of the removal effects of flexible endoscope cleaning brushes on endoscopic Pseudomonas Aeruginosa
    Ma, Jiu-Hong
    Huang, Xi
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2013, 28 : 484 - 484
  • [38] ADENOSINE TRIPHOSPHATE-SENSITIVE POTASSIUM CHANNELS IN THE CARDIOVASCULAR-SYSTEM
    NICHOLS, CG
    LEDERER, WJ
    AMERICAN JOURNAL OF PHYSIOLOGY, 1991, 261 (06): : H1675 - H1686
  • [39] Reprocessing of GI endoscopes: Standardization and validation of a manual cleaning protocol
    Foliente, RL
    Chen, YK
    Kovacs, BJ
    Aprecio, RM
    Kettering, JD
    GASTROINTESTINAL ENDOSCOPY, 1999, 49 (04) : AB141 - AB141
  • [40] Beyond What the Eye Can See: Verifying Surgical Instrument and Endoscope Cleanliness With Adenosine Triphosphate Technology
    Mitchell, Christy V.
    Anderson, Richard D.
    Taylor, Laura A.
    Rodriguez, Jose A.
    Bradley, David F., Jr.
    Phillips, Angela K.
    Romito, Kenneth J.
    AORN JOURNAL, 2023, 118 (02) : 70 - 71