Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels

被引:90
|
作者
Alfa, Michelle J. [1 ,2 ,3 ]
Fatima, Iram [3 ]
Olson, Nancy [3 ]
机构
[1] Diagnost Serv Manitoba, Clin Microbiol, Winnipeg, MB R2H 2A6, Canada
[2] St Boniface Res Ctr, Microbiol Lab, Winnipeg, MB, Canada
[3] Univ Manitoba, Dept Med Microbiol, Winnipeg, MB, Canada
关键词
Simulated use; Endoscope channel; Cleaning-compliance; Benchmark; Relative light units; DECONTAMINATION; EFFICACY;
D O I
10.1016/j.ajic.2012.03.018
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Compliance with cleaning of flexible endoscope channels cannot be verified using visual inspection. Adenosine triphosphate (ATP) has been suggested as a possible rapid cleaning monitor for flexible endoscope channels. There have not been published validation studies to specify the level of ATP that indicates inadequate cleaning has been achieved. Objective: The objective of this study was to validate the Clean-Trace (3M Inc, St. Paul, MN) ATP water test method for monitoring manual cleaning of flexible endoscopes. Methods: This was a simulated use study using a duodenoscope as the test device. Artificial test soil containing 10(6) colony-forming units of Pseudomonas aeruginosa and Enterococcus faecalis was used to perfuse all channels. The flush sample method for the suction-biopsy (L1) or air-water channel (L2) using 40 and 20 mLs sterile reverse osmosis water, respectively, was validated. Residuals of ATP, protein, hemoglobin, and bioburden were quantitated from channel samples taken from uncleaned, partially cleaned, and fully cleaned duodenoscopes. The benchmarks for clean were as follows: <6.4 mu g/cm(2) protein, <2.2 mu g/cm(2) hemoglobin, and <4-log(10) colony-forming units/cm(2) bioburden. Results: The average ATP in clean channel samples was 27.7 RLUs and 154 RLUs for L1 and L2, respectively (<200 RLUs for all channels). The average protein, hemoglobin, and bioburden benchmarks were achieved if <200 RLUs were detected. If the channel sample was >200 RLUs, the residual organic and bioburden levels would exceed the acceptable benchmarks. Conclusion: Our data validated that flexible endoscopes that have complete manual cleaning will have <200 RLUs by the Clean-Trace ATP test. Copyright (C) 2013 by the Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 50 条
  • [21] ADENOSINE-TRIPHOSPHATE SENSITIVE POTASSIUM CHANNELS IN ANOXIA
    KRNJEVIC, K
    STROKE, 1990, 21 (11) : 190 - 193
  • [22] ADENOSINE 5'-TRIPHOSPHATE-SENSITIVE POTASSIUM CHANNELS
    ASHCROFT, FM
    ANNUAL REVIEW OF NEUROSCIENCE, 1988, 11 : 97 - 118
  • [23] Is automated cleaning as good as manual cleaning in the reprocessing of flexible GI endoscopes?
    Gabriel, M
    Aprecio, RM
    Foliente, RL
    Kettering, JD
    Chen, YK
    GASTROINTESTINAL ENDOSCOPY, 2000, 51 (04) : AB251 - AB251
  • [24] Simulated-use evaluation of rapid ChannelCheckTM cleaning test for optimal detection of organic residues in flexible endoscope channels
    Kulkarni, K.
    Gavette, M.
    Alfa, M. J.
    JOURNAL OF HOSPITAL INFECTION, 2024, 152 : 66 - 72
  • [25] Adenosine triphosphate bioluminescence to evaluate the effectiveness of surface cleaning: an integrative review
    de Oliveira, Adriana Cristina
    Viana, Roberta El Hariri
    REVISTA BRASILEIRA DE ENFERMAGEM, 2014, 67 (06) : 987 - 993
  • [26] The Use of ATPase Monitoring to Determine Endoscope Bioburden and Assure Satisfactory Manual Cleaning
    Boroda, Konstantin
    Lugay, Melinda
    Sabuda, Marla D.
    Lewis, Blair S.
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2016, 111 : S153 - S153
  • [27] Development and Validation of a New Generation of Flexible Endoscope for NOTES
    Swanstrom, Lee
    Swain, Paul
    Denk, Peter
    SURGICAL INNOVATION, 2009, 16 (02) : 104 - 110
  • [28] Molecular biology of adenosine triphosphate-sensitive potassium channels
    Aguilar-Bryan, L
    Bryan, J
    ENDOCRINE REVIEWS, 1999, 20 (02) : 101 - 135
  • [29] Adenosine triphosphate-sensitive potassium channels and cardiomyopathies (Review)
    Liu, Zhongwei
    Cai, Hui
    Dang, Yonghui
    Qiu, Chuan
    Wang, Junkui
    MOLECULAR MEDICINE REPORTS, 2016, 13 (02) : 1447 - 1454
  • [30] Validation and comparison of three adenosine triphosphate luminometers for monitoring hospital surface sanitization: A Rosetta Stone for adenosine triphosphate testing
    Sciortino, Carmen V.
    Giles, R. Allen
    AMERICAN JOURNAL OF INFECTION CONTROL, 2012, 40 (08) : E233 - E239