The novel magnesium-titanium hybrid cannulated screws for the treatment of vertical femoral neck fractures: Biomechanical evaluation

被引:6
|
作者
Zhang, Qida [1 ,6 ]
Chen, Zhenxian [2 ]
Peng, Yinghu [3 ]
Jin, Zhongmin [4 ,5 ]
Qin, Ling [1 ,6 ]
机构
[1] Chinese Univ Hong Kong, Dept Orthopaed & Traumatol, Musculoskeletal Res Lab, Hong Kong, Peoples R China
[2] Changan Univ, Sch Mech Engn, Key Lab Rd Construction Technol & Equipment, Minist Educ, Xian, Peoples R China
[3] Chinese Acad Sci, Shenzhen Inst Adv Technol, CAS Key Lab Human Machine Intelligence Synergy Sys, Shenzhen, Peoples R China
[4] Southwest Jiaotong Univ, Tribol Res Inst, Sch Mech Engn, Chengdu, Peoples R China
[5] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds, England
[6] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Orthopaed & Traumatol, Musculoskeletal Res Lab,Shatin, 30-32 Ngan Shing St, Hong Kong, Peoples R China
基金
中国国家自然科学基金;
关键词
Biomechanics; Cannulated screws; Finite element analysis; Musculoskeletal multibody model; Vertical femoral neck fractures; MECHANICAL-PROPERTIES; INTERNAL-FIXATION; TRABECULAR BONE; PREDICTION; DENSITY; LOAD; TRANSLATION; FORCE;
D O I
10.1016/j.jot.2023.08.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Conventional cannulated screws are commonly used for internal fixation in the treatment of vertical femoral neck fractures. However, the noticeably high rates of undesirable outcomes such as nonunion, malunion, avascular necrosis, and fixation failure still troubled the patients and surgeons. It is urgent to develop new cannulated screws to improve the above clinical problems. The purpose of this study was to design a novel magnesium-titanium hybrid cannulated screw and to further evaluate its biomechanical performance for the treatment of vertical femoral neck fractures. Methods: A novel magnesium-titanium hybrid cannulated screw was designed, and the conventional titanium cannulated screw was also modeled. The finite element models for vertical femoral neck fractures with magnesium-titanium hybrid cannulated screws and conventional cannulated screws were respectively established. The hip joint contact force during walking gait calculated by a subject-specific musculoskeletal multibody dynamics model, was used as loads and boundary conditions for both finite element models. The stress and displacement distributions of the cannulated screws and the femur, the micromotion of the fracture surfaces of the femoral neck, and the overall stiffness were calculated and analyzed using finite element models. The biomechanical performance of the Magnesium-Titanium hybrid cannulated screws was evaluated. Results: The maximum stresses of the magnesium-titanium hybrid cannulated screws and the conventional cannulated screws were 451.5 MPa and 476.8 MPa, respectively. The maximum stresses of the femur with the above different cannulated screws were 140.3 MPa and 164.8 MPa, respectively. The maximum displacement of the femur with the hybrid cannulated screws was 6.260 mm, lower than the femur with the conventional cannulated screws, which was 7.125 mm. The tangential micromotions in the two orthogonal directions at the fracture surface of the femoral neck with the magnesium-titanium hybrid cannulated screws were comparable to those with the conventional cannulated screws. The overall stiffness of the magnesium-titanium hybrid cannulated screw system was 490.17 N/mm, higher than that of the conventional cannulated screw system, which was 433.92 N/mm. Conclusion: The magnesium-titanium hybrid cannulated screw had superior mechanical strength and fixation stability for the treatment of the vertical femoral neck fractures, compared with those of the conventional cannulated screw, indicating that the magnesium-titanium hybrid cannulated screw has great potential as a new fixation strategy in future clinical applications. The translational potential of this article: This study highlights an innovative design of the magnesium-titanium hybrid cannulated screw for the treatment of vertical femoral neck fractures. The novel magnesium-titanium
引用
收藏
页码:127 / 136
页数:10
相关论文
共 50 条
  • [31] Value of Washers in Internal Fixation of Femoral Neck Fractures With Cancellous Screws: A Biomechanical Evaluation
    Zlowodzki, Michael P.
    Wijdicks, Coen A.
    Armitage, Bryan M.
    Cole, Peter A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (02) : E69 - E72
  • [32] Biomechanical Research of Three Parallel Cannulated Compression Screws in Oblique Triangle Configuration for Fixation of Femoral Neck Unstable Fractures
    Zhang, Ru-Yi
    Zhang, Wu-Peng
    Yang, Guang-Min
    Wang, Dao-Feng
    Su, Peng
    Zhang, Yi
    Nie, Shao-Bo
    Li, Jia
    Zhao, Zhe
    Li, Jian-Tao
    Zhang, Li-Cheng
    Tang, Pei-Fu
    ORTHOPAEDIC SURGERY, 2024, 16 (04) : 953 - 964
  • [33] Prediction of Reoperation of Femoral Neck Fractures Treated With Cannulated Screws in Elderly Patients
    Gregersen, Merete
    Krogshede, Anna
    Brink, Ole
    Damsgaard, Else Marie
    GERIATRIC ORTHOPAEDIC SURGERY & REHABILITATION, 2015, 6 (04) : 322 - 327
  • [34] The Effect of Configuration of Rhombic Cannulated Screws on Internal Fixation of Femoral Neck Fractures
    Guo, Jialiang
    Dong, Weichong
    Yin, Yingchao
    Zhang, Ruipeng
    Hou, Zhiyong
    Zhang, Yingze
    ORTHOPEDICS, 2020, 43 (02) : E72 - E78
  • [35] Fixation of femoral neck fractures using divergent versus parallel cannulated screws
    Papanastassiou I.D.
    Mavrogenis A.F.
    Kokkalis Z.T.
    Nikolopoulos K.
    Skourtas K.
    Papagelopoulos P.J.
    Journal of Long-Term Effects of Medical Implants, 2011, 21 (01) : 63 - 69
  • [36] Evaluation of a Bi-Planar Robot Navigation System for Insertion of Cannulated Screws in Femoral Neck Fractures
    He, Meng
    Han, Wei
    Zhao, Chun-peng
    Su, Yong-gang
    Zhou, Li
    Wu, Xin-bao
    Wang, Jun-qiang
    ORTHOPAEDIC SURGERY, 2019, 11 (03) : 373 - 379
  • [37] Use of fully threaded cannulated screws decreases femoral neck shortening after fixation of femoral neck fractures
    Yoram A. Weil
    Feras Qawasmi
    Meir Liebergall
    Rami Mosheiff
    Amal Khoury
    Archives of Orthopaedic and Trauma Surgery, 2018, 138 : 661 - 667
  • [38] Use of fully threaded cannulated screws decreases femoral neck shortening after fixation of femoral neck fractures
    Weil, Yoram A.
    Qawasmi, Feras
    Liebergall, Meir
    Mosheiff, Rami
    Khoury, Amal
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2018, 138 (05) : 661 - 667
  • [39] Treatment of Delbet II/III Pediatric Femoral Neck Fractures With Proximal Femoral Locking Plate Versus Cannulated Screws
    Haider, Shamrez
    Harris, Timothy J.
    Turner, Alexander C.
    Podeszwa, David A.
    Hartman, Courtney A.
    Morris, William Z.
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2024, 44 (04) : 213 - 220
  • [40] Biomechanical optimization of the magnesium alloy bionic cannulated screw for stabilizing femoral neck fractures: a finite element analysis
    Cui, Yunwei
    Ding, Kai
    Lv, Hongzhi
    Cheng, Xiaodong
    Fan, Zixi
    Sun, Dacheng
    Zhang, Yifan
    Chen, Wei
    Zhang, Yingze
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12