Paediatric effective radiation doses during brain computed tomography angiography procedure

被引:2
|
作者
Salah, H. [1 ,2 ]
Rabbaa, Mohammad [3 ]
Abuljoud, Mohammad [3 ]
Babikir, E. [4 ]
Alkhorayef, M. [5 ]
Tamam, N. [6 ]
Tahir, Dahlang [7 ]
Sulieman, A. [8 ]
Bradley, D. A. [9 ,10 ]
机构
[1] INAYA Med Collage, Nucl Med Dept, Riyadh, Saudi Arabia
[2] Sudan Univ Sci & Technol, Coll Med Radiol Sci, Khartoum, Sudan
[3] Riyadh Care Hosp, Radiol Dept, Riyadh, Saudi Arabia
[4] Univ Bahrain, Coll Hlth & Sport Sci, Radiol Technol Program, Sakhir Campus,POB 32038, Zallaq, Bahrain
[5] King Saud Univ, Coll Appl Med Sci, Dept Radiol Sci, POB 10219, Riyadh 11433, Saudi Arabia
[6] Princess Nourah bint Abdulrahman Univ, Coll Sci, Dept Phys, POB 84428, Riyadh 11671, Saudi Arabia
[7] Hasanuddin Univ, Dept Phys, Makassar 90245, Indonesia
[8] Prince Sattam bin Abdulaziz Univ, Coll Appl Med Sci, Radiol & Med Imaging Dept, POB 422, Alkharj 11942, Saudi Arabia
[9] Univ Surrey, Ctr Nucl & Radiat Phys, Dept Phys, Guildford, England
[10] Sunway Univ, Ctr Appl Phys & Radiat Technol, Sch Engn & Technol, Bandar Sunway 47500, Selangor, Malaysia
关键词
CT brain; Diagnostic radiology; Paediatirc imaging; Radiation dosimetry; DLP; Effective dose; DIAGNOSTIC REFERENCE LEVELS; CT; REDUCTION; CHILDREN; QUALITY;
D O I
10.1016/j.apradiso.2022.110610
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In comparison to adults and paediatric are more sensitive to ionizing radiation exposure. Computed tomography (CT) is now the dominant source of medical radiologic tests for patients, accounting for more than 70% of total doses to the general public. Paediatric CT brain scans (with and without contrast) are routinely performed for a variety of clinical reasons. As a result, this parameter must be calculated in order to determine relative radiation risk. The goal of this study is to assess the radiation risk to children during CT brain diagnostic procedures. Three hundred fifty three child patients' radiation risk doses were assessed over the course of a year. The mean and ranged of the children's radiation doses were 40.6 +/- 8.8 (27.8-45.8) CTDIvol (mGy) and 850 +/- 230 (568.1-1126.4) DLP (mGy.cm) for the brain with contrast medium. For CT brain without contrast, the patients' doses were 40.9 +/- 9.4 (14.27-64.07) CTDIvol (mGy), and 866.1 +/- 289.3 (203.6-2484.9) DLP (mGy.cm). The characteristics related to the radiation dose were retrieved from the scan protocol generated by the CT system by the participating physicians after each procedure. Furthermore, optimizing the CT acquisition parameter is critical for increasing the benefit while lowering the procedure's radiogenic risk. The patients' radiation dose is comparable with the most previously published studies and international diagnostic reference levels (DRLs). Radiation dose optimization is recommended due to high sensitivity of the paediatric patients to ionizing radiation.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Spinal intraarterial computed tomography angiography as an effective adjunct for spinal angiography
    Yamamoto, Shinji
    Kanaya, Hideaki
    Kim, Phyo
    JOURNAL OF NEUROSURGERY-SPINE, 2015, 23 (03) : 360 - 367
  • [32] Radiation dose reduction in paediatric coronary computed tomography: assessment of effective dose and image quality
    Geryes, Bouchra Habib
    Calmon, Raphael
    Khraiche, Diala
    Boddaert, Nathalie
    Bonnet, Damien
    Raimondi, Francesca
    EUROPEAN RADIOLOGY, 2016, 26 (07) : 2030 - 2038
  • [33] Radiation dose reduction in paediatric coronary computed tomography: assessment of effective dose and image quality
    Bouchra Habib Geryes
    Raphael Calmon
    Diala Khraiche
    Nathalie Boddaert
    Damien Bonnet
    Francesca Raimondi
    European Radiology, 2016, 26 : 2030 - 2038
  • [34] Radiation exposure during paediatric CT in Sudan: CT dose, organ and effective doses
    Suliman, I. I.
    Khamis, H. M.
    Ombada, T. H.
    Alzimami, K.
    Alkhorayef, M.
    Sulieman, A.
    RADIATION PROTECTION DOSIMETRY, 2015, 167 (04) : 513 - 518
  • [35] Organ and effective doses in newborn patients during helical multislice computed tomography examination
    Staton, Robert J.
    Lee, Choonik
    Lee, Choonsik
    Williams, Matt D.
    Hintenlang, David E.
    Arreola, Manuel M.
    Williams, Jonathon L.
    Bolch, Wesley E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (20): : 5151 - 5166
  • [36] Radiation Dose to Radiosensitive Organs during Head/Sinus Computed Tomography Procedure
    Taha, Taha Mohamed
    ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2023, 56 (01): : 112 - 116
  • [37] Effective radiation dose in angiography of coronary artery bypass grafts using multislice computed tomography
    Smith, D. K.
    Negus, I. S.
    Murphy, G.
    Coles, D. R.
    Wilde, P.
    Angelini, G. D.
    Karsch, K. R.
    Baumbach, A.
    EUROPEAN HEART JOURNAL, 2006, 27 : 290 - 290
  • [38] Computed tomography scan radiation doses: Call for optimisation
    Parthasarathy, K. S.
    JOURNAL OF MEDICAL PHYSICS, 2007, 32 (02) : 49 - 50
  • [39] Effective radiation dose in angiography of coronary artery bypass grafts using multislice computed tomography
    Smith, D.
    Negus, I.
    Hamilton, M.
    Baumbach, A.
    HEART, 2007, 93 : A63 - A64
  • [40] Computed Tomography Perfusion is aUseful Adjunct to Computed Tomography Angiography in the Diagnosis of Brain Death
    Sawicki, M.
    Solek-Pastuszka, J.
    Chamier-Cieminska, K.
    Walecka, A.
    Walecki, J.
    Bohatyrewicz, R.
    CLINICAL NEURORADIOLOGY, 2019, 29 (01) : 101 - 108