Optimal Protocol for Contrast-enhanced Free-running 5D Whole-heart Coronary MR Angiography at 3T

被引:7
|
作者
Ishida, Masaki [1 ]
Yerly, Jerome [2 ,3 ,4 ]
Ito, Haruno [1 ]
Takafuji, Masafumi [1 ]
Nakamori, Shiro [5 ]
Takase, Shinichi [1 ]
Ichiba, Yoshito [6 ]
Komori, Yoshiaki [6 ]
Dohi, Kaoru [5 ]
Piccini, Davide [2 ,3 ,4 ,7 ]
Bastiaansen, Jessica A. M. [2 ,3 ,4 ,8 ]
Stuber, Matthias [2 ,3 ,4 ]
Sakuma, Hajime [1 ]
机构
[1] Mie Univ Hosp, Dept Radiol, 2-174 Edobashi, Tsu, Mie 5148507, Japan
[2] Lausanne Univ Hosp CHUV, Dept Diagnost & Intervent Radiol, Lausanne, Switzerland
[3] Univ Lausanne UNIL, Lausanne, Switzerland
[4] Ctr Biomed Imaging CIBM, Lausanne, Switzerland
[5] Mie Univ Hosp, Dept Cardiol, Tsu, Mie, Japan
[6] Siemens Healthcare KK, Tokyo, Japan
[7] Siemens Healthcare AG, Adv Clin Imaging Technol, Lausanne, Switzerland
[8] Univ Bern, Dept Diagnost Intervent & Pediat Radiol DIPR, Inselspital Bern Univ Hosp, Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
contrast medium; coronary Magnetic Resonance angiography; fat suppression; free-running framework; gradient echo; VISUALIZATION;
D O I
10.2463/mrms.tn.2022-0086
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Free-running 5D whole-heart coronary MR angiography (MRA) is gaining in popularity because it reduces scanning complexity by removing the need for specific slice orientations, respiratory gating, or cardiac triggering. At 3T, a gradient echo (GRE) sequence is preferred in combination with contrast injection. However, neither the injection scheme of the gadolinium (Gd) contrast medium, the choice of the RF excitation angle, nor the dedicated image reconstruction parameters have been established for 3T GRE free-running 5D whole-heart coronary MRA. In this study, a Gd injection scheme, RF excitation angles of lipid-insensitive binominal off-resonance RF excitation (LIBRE) pulse for valid fat suppression and continuous data acquisition, and compressed-sensing reconstruction regularization parameters were optimized for contrast-enhanced free-running 5D whole-heart coronary MRA using a GRE sequence at 3T. Using this optimized protocol, contrast-enhanced free-running 5D whole-heart coronary MRA using a GRE sequence is feasible with good image quality at 3T.
引用
收藏
页码:225 / 237
页数:13
相关论文
共 50 条
  • [31] Contrast-Enhanced Whole-Heart Coronary MRI with Bolus Infusion of Gadobenate Dimeglumine at 1.5 T
    Hu, Peng
    Chan, Jonathan
    Ngo, Long H.
    Smink, Jouke
    Goddu, Beth
    Kissinger, Kraig V.
    Goepfert, Lois
    Hauser, Thomas H.
    Rofsky, Neil M.
    Manning, Warren J.
    Nezafat, Reza
    MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (02) : 392 - 398
  • [32] Contrast-enhanced whole-heart coronary magnetic resonance angiography (MRA) in less than 5 minutes using radial EPI
    Himanshu Bhat
    Qi Yang
    Sven Zuehlsdorff
    Debiao Li
    Journal of Cardiovascular Magnetic Resonance, 12 (Suppl 1)
  • [33] Non-contrast compressed sensing whole-heart coronary magnetic resonance angiography at 3T: A comparison with conventional imaging
    Nakamura, Masashi
    Kido, Tomoyuki
    Kido, Teruhito
    Watanabe, Kouki
    Schmidt, Michaela
    Forman, Christoph
    Mochizuki, Teruhito
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 104 : 43 - 48
  • [34] Whole-heart coronary magnetic resonance angiography - Contrast-enhanced high-resolution, time-resolved 3D imaging
    Warmuth, Carsten
    Schnorr, Joerg
    Kaufels, Nicola
    Wagner, Susanne
    Pilgrimm, Herbert
    Hamm, Bernd
    Taupitz, Matthias
    INVESTIGATIVE RADIOLOGY, 2007, 42 (08) : 550 - 557
  • [35] An automated approach to fully self-gated free-running cardiac and respiratory motion-resolved 5D whole-heart MRI
    Di Sopra, Lorenzo
    Piccini, Davide
    Coppo, Simone
    Stuber, Matthias
    Yerly, Jerome
    MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (06) : 2118 - 2132
  • [36] Self-gated free-running 5D whole-heart MRI using blind source separation for automated cardiac motion extraction
    Monton Quesada, Isabel
    Ogier, Augustin C.
    Ishida, Masaki
    Takafuji, Masafumi
    Ito, Haruno
    Sakuma, Hajime
    Romanin, Ludovica
    Roy, Christopher W.
    Prsa, Milan
    Richiardi, Jonas
    Yerly, Jerome
    Stuber, Matthias
    van Heeswijk, Ruud B.
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (03) : 961 - 974
  • [37] Contrast-enhanced MR angiography at 3T in the evaluation of intracranial aneurysms: A comparison with time-of-flight MR angiography
    Nael, K.
    Villablanca, J. P.
    Saleh, R.
    Pope, W.
    Nael, A.
    Laub, G.
    Finn, J. P.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2006, 27 (10) : 2118 - 2121
  • [38] Evaluation of Brain and Head and Neck Tumors with 4D Contrast-Enhanced MR Angiography at 3T
    Nishimura, S.
    Hirai, T.
    Shigematsu, Y.
    Kitajima, M.
    Morioka, M.
    Kai, Y.
    Minoda, R.
    Uetani, H.
    Murakami, R.
    Yamashita, Y.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (03) : 445 - 448
  • [39] Contrast-Enhanced Whole-Heart Coronary Magnetic Resonance Angiography at 3.0-T A Comparative Study With X-Ray Angiography in a Single Center
    Yang, Qi
    Li, Kuncheng
    Liu, Xin
    Bi, Xiaoming
    Liu, Zhi
    An, Jing
    Zhang, Al
    Jerecic, Renate
    Li, Debiao
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (01) : 69 - 76
  • [40] Whole-heart coronary MR angiography using a 3D cones phyllotaxis trajectory
    Malave, Mario O.
    Baron, Corey A.
    Addy, Nii Okai
    Cheng, Joseph Y.
    Yang, Phillip C.
    Hu, Bob S.
    Nishimura, Dwight G.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (02) : 1092 - 1103