FLASH Radiotherapy and the Use of Radiation Dosimeters

被引:23
|
作者
Siddique, Sarkar [1 ]
Ruda, Harry E. [2 ,3 ]
Chow, James [4 ,5 ]
机构
[1] Toronto Metropolitan Univ, Dept Phys, Toronto, ON M5B 2K3, Canada
[2] Univ Toronto, Fac Appl Sci & Engn, Ctr Adv Nanotechnol, Toronto, ON M5S 3E4, Canada
[3] Univ Toronto, Dept Mat Sci & Engn, Toronto, ON M5S 3E4, Canada
[4] Univ Hlth Network, Princess Margaret Canc Ctr, Radiat Med Program, Toronto, ON M5G 1X6, Canada
[5] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5T 1P5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
FLASH radiotherapy; ultra-high dose rate radiotherapy; detectors; dosimetry; CHAMBER;
D O I
10.3390/cancers15153883
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Simple Summary: FLASH radiotherapy (RT) delivering ultra-high dose rate radiation can reduce normal tissue toxicity while effectively treating tumors. However, implementing FLASH RT in clinical settings faces challenges like limited depth penetration and complex treatment planning. Monte Carlo simulation is a valuable tool to optimize FLASH RT. Radiation detectors, including diamond detectors like microDiamond and ionization chambers, play a crucial role in accurately measuring dose delivery. Moreover, optically stimulated luminescence dosimeters and radiochromic films are used for validation. Advancements are being made to improve detector accuracy in FLASH RT. Further research is needed to refine treatment planning and detector performance for widespread FLASH RT implementation, which can potentially revolutionize cancer treatment. Radiotherapy (RT) using ultra-high dose rate (UHDR) radiation, known as FLASH RT, has shown promising results in reducing normal tissue toxicity while maintaining tumor control. However, implementing FLASH RT in clinical settings presents technical challenges, including limited depth penetration and complex treatment planning. Monte Carlo (MC) simulation is a valuable tool for dose calculation in RT and has been investigated for optimizing FLASH RT. Various MC codes, such as EGSnrc, DOSXYZnrc, and Geant4, have been used to simulate dose distributions and optimize treatment plans. Accurate dosimetry is essential for FLASH RT, and radiation detectors play a crucial role in measuring dose delivery. Solid-state detectors, including diamond detectors such as microDiamond, have demonstrated linear responses and good agreement with reference detectors in UHDR and ultra-high dose per pulse (UHDPP) ranges. Ionization chambers are commonly used for dose measurement, and advancements have been made to address their response nonlinearities at UHDPP. Studies have proposed new calculation methods and empirical models for ion recombination in ionization chambers to improve their accuracy in FLASH RT. Additionally, strip-segmented ionization chamber arrays have shown potential for the experimental measurement of dose rate distribution in proton pencil beam scanning. Radiochromic films, such as Gafchromic (TM) EBT3, have been used for absolute dose measurement and to validateMCsimulation results in high-energy X-rays, triggering the FLASH effect. These films have been utilized to characterize ionization chambers and measure off-axis and depth dose distributions in FLASH RT. In conclusion, MC simulation provides accurate dose calculation and optimization for FLASH RT, while radiation detectors, including diamond detectors, ionization chambers, and radiochromic films, offer valuable tools for dosimetry in UHDR environments. Further research is needed to refine treatment planning techniques and improve detector performance to facilitate the widespread implementation of FLASH RT, potentially revolutionizing cancer treatment.
引用
收藏
页数:16
相关论文
共 50 条
  • [21] CALIBRATION OF RADIOTHERAPY DOSIMETERS AGAINST SECONDARY STANDARD DOSIMETERS - AN ANOMALOUS RESULT
    LAW, J
    FOSTER, CJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (08): : 1039 - 1043
  • [22] USE OF THERMOLUMINESCENT DOSIMETERS FOR MEASUREMENT OF EXTERNAL RADIATION EXPOSURE IN GUARAPARIS BRAZIL
    CULLEN, TL
    HEALTH PHYSICS, 1965, 11 (08): : 806 - +
  • [23] Development of MOS-FET dosimeters for use in high radiation fields
    Kramberger, G.
    Ambrozic, K.
    Gurer, U.
    Hiti, B.
    Karacali, H.
    Mandic, I
    Yilmaz, E.
    Yilmaz, O.
    Zavrtanik, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 978
  • [24] Feasibility of Plastic Scintillator Dosimeters for FLASH Therapy
    Cecchi, D.
    Therriault-Proulx, F.
    Bazalova-Carter, M.
    MEDICAL PHYSICS, 2020, 47 (06) : E783 - E783
  • [25] THE USE OF RADIATION-DAMAGE TO MATCH THE SENSITIVITY OF LIF TLD DOSIMETERS
    ELLIOTT, JH
    HEALTH PHYSICS, 1983, 45 (01): : 205 - 206
  • [26] Breeding behavior of radiation-induced effects in organic materials and their possible use as radiation dosimeters
    Abdelaal, Saad
    Abdelhady, A. M.
    Tokhy, Hatem Hussny
    Eid, A. M.
    Rammah, Y. S.
    Awad, El-Sayed Mahmoud
    Elmaghraby, Elsayed K.
    JOURNAL OF PHYSICS AND CHEMISTRY OF SOLIDS, 2021, 150
  • [27] CVD diamond detectors as dosimeters for radiotherapy
    Buttar, CM
    Conway, J
    Meyfarth, R
    Scarsbrook, G
    Sellin, PJ
    Whitehead, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3): : 281 - 284
  • [28] FLASH radiotherapy
    Moeckli, R.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S555 - S555
  • [29] FLASH Proton Radiotherapy on a Model of Radiation-Induced Myocardial Fibrosis
    Verginadis, I.
    Kim, K.
    Kim, M. M.
    Skoufos, G.
    Diffenderfer, E. S.
    Motlagh, S. A. O.
    Kokkorakis, M.
    Koliaki, I.
    Morcos, G.
    Shoniyozov, K.
    Griffin, J.
    Hatzigeorgiou, A.
    Metz, J. M.
    Lin, A.
    Feigenberg, S. J.
    Cengel, K. A.
    Ky, B.
    Koumenis, C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : S136 - S136
  • [30] Correction: FLASH-Radiotherapy: A Potential Innovation Driver in Radiation Therapy
    Young-Seok Seo
    Woo-Yoon Park
    Byung Jun Min
    Kyungdon Choi
    Ki-Whwan Kim
    Journal of the Korean Physical Society, 2020, 77 : 537 - 537