Spectral analysis of Monte Carlo calculated fluence correction and cema conversion factors for high- energy photon beams at different depths

被引:0
|
作者
Roers, Julian [1 ,2 ]
Czarnecki, Damian [1 ]
Alissa, Mohamad [1 ]
Zink, Klemens [1 ,3 ,4 ]
机构
[1] Univ Appl Sci Giessen, Inst Med Phys & Radiat Protect IMPS, Giessen, Germany
[2] Univ Hosp Munster, Dept Radiotherapy, Munster, Germany
[3] Univ Med Ctr Giessen Marburg, Dept Radiotherapy & Radiooncol, Marburg, Germany
[4] Marburg Iontherapy Ctr MIT, Marburg, Germany
关键词
radiation dosimetry; CEMA; charged particle fluence; detector response; perturbation factor; build-up region; Monte Carlo simulation; REPLACEMENT CORRECTION FACTORS; DETECTOR DOSE-RESPONSE; REFERENCE DOSIMETRY; MEGAVOLTAGE PHOTON; ION CHAMBERS; PERTURBATION; QUALITY;
D O I
10.3389/fphy.2022.1075514
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Purpose: The aim of this study is to investigate the depth-dependent detector response of detailed thimble air-filled ionization chambers by calculating spectral charged particle fluence correction factors at different depths in water. Those spectral correction factors will help to understand, how the detector response varies at different depths and what kind of influences disparate effects have on the spectral detector response.Methods: The cema-approach can be used to obtain spectral charged particle fluence-based correction factors for various measurement conditions by substituting the commonly well-known dose conversion factor with a conversion factor based on the dosimetric quantity cema ( "converted-energy per unit mass "). The resulting spectral fluence correction factors were calculated with the EGSnrc software toolkit and analyzed for two air-filled cylindrical ionization chambers (PTW type 31021 Semiflex 3D, SNC125c (TM)) at different depths in a water phantom irradiated with a 6 MV linear accelerator x-ray spectrum. The ionization chamber models have been stepwise decomposed to investigate the perturbation caused by internal and external effects on the fluence distribution within the detector.Results: Monte Carlo calculated fluence-based perturbation correction factors revealed that for all investigated detectors, considerable fluence disturbances occur, especially in the build-up region of depth-dose curves. Our results have shown that even slight variations in depth can have major consequences on the differential charged particle fluence within the ionization chamber, mainly due to internal cavity-specific effects. Furthermore, the results showed that in the case of relative dose measurements, the depth-depending detector response can significantly differ from unity in a range of 1.4%-2.8% depending on the ionization chamber design.Conclusion: The complexity of different effects on the fluence disturbance could be broken down with regard to their influence on the spectral fluence distribution in the sensitive volume of the investigated detectors. It could be demonstrated, that the displacement of water is a depth-depending effect, which can not be compensated or corrected ideally for each investigated water depth by the shift of the effective point of measurement. Generally, the spectral analysis of those energy-dependent correction factors serves to a deeper understanding of the detector response under various conditions.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beams
    Kuenzler, Thomas
    Fotina, Irina
    Stock, Markus
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (24): : 7363 - 7377
  • [42] Monte Carlo and Water Calorimetric Determination of Ionization Chamber Correction Factors for Medium Energy X-Ray Beams
    Bancheri, J.
    Ketelhut, S.
    Bueermann, L.
    Seuntjens, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E816 - E816
  • [43] Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams
    Fotina, Irina
    Winkler, Peter
    Kuenzler, Thomas
    Reiterer, Jochen
    Simmat, Isabell
    Georg, Dietmar
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (03) : 645 - 653
  • [44] Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo study
    Benmakhlouf, Hamza
    Sempau, Josep
    Andreo, Pedro
    MEDICAL PHYSICS, 2014, 41 (04)
  • [45] Absorbed-dose beam quality conversion factors for cylindrical chambers in high energy photon beams
    Seuntjens, JP
    Ross, CK
    Shortt, KR
    Rogers, DWO
    MEDICAL PHYSICS, 2000, 27 (12) : 2763 - 2779
  • [46] Experimentally determined and Monte Carlo-calculated energy dependence of NaCl pellets read by optically stimulated luminescence for photon beams in the energy range 30 keV to 1.25 MeV
    Waldner, Lovisa
    Raaf, Christopher
    Hinrichsen, Yvonne
    Herrnsdorf, Lars
    Bernhardsson, Christian
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (04) : 1321 - 1335
  • [47] The dose discrepancies between Monte Carlo calculations and measurements in the build-up region for high-energy photon beams
    Fu, Y
    Li, S
    Ayyangar, K
    Enke, C
    MEDICAL PHYSICS, 2005, 32 (06) : 2014 - 2014
  • [48] Monte Carlo calculation of detector perturbation and quality correction factors in a 1.5 T magnetic resonance guided radiation therapy small photon beams
    Cervantes, Yunuen
    Duchaine, Jasmine
    Billas, Ilias
    Duane, Simon
    Bouchard, Hugo
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (22):
  • [49] Dose-mean lineal energy values for electrons by different Monte Carlo codes: Consequences for estimates of radiation quality in photon beams
    Lindborg, Lennart
    Lillhok, Jan
    Kyriakou, Ioanna
    Emfietzoglou, Dimitris
    MEDICAL PHYSICS, 2022, 49 (02) : 1286 - 1296
  • [50] Correction factors for the NMi free-air ionization chamber for medium-energy x-rays calculated with the Monte Carlo method
    Grimbergen, TWM
    van Dijk, E
    de Vries, W
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (11): : 3207 - 3224