Spectral analysis of Monte Carlo calculated fluence correction and cema conversion factors for high- energy photon beams at different depths

被引:0
|
作者
Roers, Julian [1 ,2 ]
Czarnecki, Damian [1 ]
Alissa, Mohamad [1 ]
Zink, Klemens [1 ,3 ,4 ]
机构
[1] Univ Appl Sci Giessen, Inst Med Phys & Radiat Protect IMPS, Giessen, Germany
[2] Univ Hosp Munster, Dept Radiotherapy, Munster, Germany
[3] Univ Med Ctr Giessen Marburg, Dept Radiotherapy & Radiooncol, Marburg, Germany
[4] Marburg Iontherapy Ctr MIT, Marburg, Germany
关键词
radiation dosimetry; CEMA; charged particle fluence; detector response; perturbation factor; build-up region; Monte Carlo simulation; REPLACEMENT CORRECTION FACTORS; DETECTOR DOSE-RESPONSE; REFERENCE DOSIMETRY; MEGAVOLTAGE PHOTON; ION CHAMBERS; PERTURBATION; QUALITY;
D O I
10.3389/fphy.2022.1075514
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Purpose: The aim of this study is to investigate the depth-dependent detector response of detailed thimble air-filled ionization chambers by calculating spectral charged particle fluence correction factors at different depths in water. Those spectral correction factors will help to understand, how the detector response varies at different depths and what kind of influences disparate effects have on the spectral detector response.Methods: The cema-approach can be used to obtain spectral charged particle fluence-based correction factors for various measurement conditions by substituting the commonly well-known dose conversion factor with a conversion factor based on the dosimetric quantity cema ( "converted-energy per unit mass "). The resulting spectral fluence correction factors were calculated with the EGSnrc software toolkit and analyzed for two air-filled cylindrical ionization chambers (PTW type 31021 Semiflex 3D, SNC125c (TM)) at different depths in a water phantom irradiated with a 6 MV linear accelerator x-ray spectrum. The ionization chamber models have been stepwise decomposed to investigate the perturbation caused by internal and external effects on the fluence distribution within the detector.Results: Monte Carlo calculated fluence-based perturbation correction factors revealed that for all investigated detectors, considerable fluence disturbances occur, especially in the build-up region of depth-dose curves. Our results have shown that even slight variations in depth can have major consequences on the differential charged particle fluence within the ionization chamber, mainly due to internal cavity-specific effects. Furthermore, the results showed that in the case of relative dose measurements, the depth-depending detector response can significantly differ from unity in a range of 1.4%-2.8% depending on the ionization chamber design.Conclusion: The complexity of different effects on the fluence disturbance could be broken down with regard to their influence on the spectral fluence distribution in the sensitive volume of the investigated detectors. It could be demonstrated, that the displacement of water is a depth-depending effect, which can not be compensated or corrected ideally for each investigated water depth by the shift of the effective point of measurement. Generally, the spectral analysis of those energy-dependent correction factors serves to a deeper understanding of the detector response under various conditions.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Monte Carlo calculation of beam quality correction factors for PTW cylindrical ionization chambers in photon beams
    Gimenez-Alventosa, Vicent
    Gimenez, Vicente
    Ballester, Facundo
    Vijande, Javier
    Andreo, Pedro
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (20):
  • [22] Monte Carlo studies on dose conversion factors from graphite to water for high energy X-ray beams
    Kim, Y. H.
    Yi, C. Y.
    Kim, I. J.
    Kim, B. C.
    Kim, J. H.
    Seong, Y. M.
    Kim, H. M.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 171
  • [23] Absorbed dose conversion factors for therapeutic kilovoltage and megavoltage x-ray beams calculated by the Monte Carlo method
    Knight, RT
    MEDICAL PHYSICS, 1997, 24 (02) : 336 - 336
  • [24] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Park, Dong-wook
    Lee, Jai-ki
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (04) : 647 - 652
  • [25] Monte Carlo calculated ionization chamber correction factors in clinical proton beams - deriving uncertainties from published data
    Baumann, Kilian-Simon
    Goma, Carles
    Wulff, Joerg
    Kretschmer, Jana
    Zink, Klemens
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 113
  • [26] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Dong-wook Park
    Jai-ki Lee
    Journal of the Korean Physical Society, 2016, 69 : 647 - 652
  • [27] Monte Carlo-based correction factors for ion chamber dosimetry in heterogeneous phantoms for megavoltage photon beams
    Araki, Fujio
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22): : 7615 - 7627
  • [28] The replacement correction factors for cylindrical chambers in high-energy photon beams
    Wang, L. L. W.
    Rogers, D. W. O.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06): : 1609 - 1620
  • [29] Monte Carlo dosimetric evaluation of high energy vs low energy photon beams in low density tissues
    Tsiakalos, Miltiadis F.
    Stathakis, Sotirios
    Plataniotis, George A.
    Kappas, Constantin
    Theodorou, Kiki
    RADIOTHERAPY AND ONCOLOGY, 2006, 79 (01) : 131 - 138
  • [30] Photon Energy Spectral Analysis in the Mucosal Dose with and Without the Flattening Filter: A Monte Carlo Study
    Chow, J.
    Owrangi, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E292 - E292