In Vivo Dosimetry in the Urethra During Prostate Carbon Ion Radiotherapy

被引:0
|
作者
Matsumoto, S. [1 ,4 ]
Matsufuji, N. [1 ]
Koba, Y. [2 ]
Tsuji, H. [3 ]
机构
[1] Natl Inst Quantum Sci & Technol, Inst Quantum Med Sci, Dept Accelerator & Med Phys, Quantum Med Sci Directorate, Chiba, Japan
[2] Natl Inst Quantum Sci & Technol, Natl Inst Radiol Sci, Dept Radiat Regulatory Sci Res, Quantum Med Sci Directorate, Chiba, Japan
[3] Natl Inst Quantum Sci & Technol, Inst Quantum Med Sci, QST Hosp, Chiba, Japan
[4] Natl Inst Quantum Sci & Technol, 4 9 1 Anagawa,Inage ku, Chiba 2638555, Japan
关键词
Radiation dosimeters; prostate cancer; urinary catheters; in vivo dosimetry; SYSTEM;
D O I
10.21873/anticanres.16389
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background/Aim: In vivo dosimetry can prevent dose delivery errors by directly measuring the dose of radiation administered to a patient. However, a method for in vivo dosimetry during carbon ion radiotherapy (CIRT) has not been established. Therefore, we investigated data from in vivo dosimetry of the urethra during CIRT for prostate cancer using small spherical diode dosimeters (SSDDs). Patients and Methods: This study included five patients enrolled in a clinical trial (jRCT identifier: jRCTs032190180) on which the use of four-fraction CIRT for prostate cancer was examined. The urethral dose during CIRT for prostate cancer was measured using the SSDDs inserted into the ureteral catheter. The relative error between the in vivo and calculated doses obtained using the Xio-N treatment planning system was determined. Additionally, a dose-response stability test for the in vivo dosimeter was performed under clinical conditions. Results: The relative error between the in vivo and calculated urethral doses ranged from 6 to 12%. The dose-response stability under clinical conditions of the measured dose was <= 1%. Therefore, an error >1% would be due to an interfractional patient setup error in the large dose gradient in the urethra. Conclusion: The usefulness of in vivo dosimetry using SSDDs in CIRT and SSDDs' potential for detecting dose delivery errors during CIRT is herein highlighted.
引用
收藏
页码:2259 / 2264
页数:6
相关论文
共 50 条
  • [21] In vivo dosimetry during conformal radiotherapy - Requirements for and findings of a routine procedure
    Lanson, JH
    Essers, M
    Meijer, GJ
    Minken, AWH
    Uiterwaal, GJ
    Mijnheer, BJ
    RADIOTHERAPY AND ONCOLOGY, 1999, 52 (01) : 51 - 59
  • [22] Development of a New Endorectal Balloon with An Unfoldable Radiochromic Film for In-Vivo Rectal Dosimetry During Prostate Cancer Radiotherapy
    Jeang, E.
    Lim, Y.
    Cho, K.
    Hwang, U.
    Kwak, J.
    Choi, S.
    Jeong, J.
    Kim, H.
    Kim, S.
    Lee, S.
    Shin, D.
    Park, J.
    Kim, J.
    MEDICAL PHYSICS, 2015, 42 (06) : 3392 - 3392
  • [23] In vivo neutron dosimetry during high-energy bremsstrahlung radiotherapy
    D'Errico, F
    Nath, R
    Silvano, G
    Tana, L
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (05): : 1185 - 1192
  • [24] In vivo dosimetry with an electronic portal imaging device for prostate cancer radiotherapy with an endorectal balloon
    Sterckx, Bo
    Steinseifer, Isabell
    Wendling, Markus
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2019, 12 : 7 - 9
  • [25] Comparison of pre-treatment and in-vivo dosimetry for advanced radiotherapy of prostate cancer
    Bartnikowska, Agnieszka
    Cieslik, Grzegorz
    Mlodzik, Mateusz
    Garcia-Argibay, Miguel
    Pract, Rep
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (02) : 189 - 197
  • [26] DEVELOPMENT OF PROCEDURES FOR IN VIVO DOSIMETRY IN RADIOTHERAPY
    Mayles, P.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S225 - S225
  • [27] In vivo dosimetry in external beam radiotherapy
    Mijnheer, Ben
    Beddar, Sam
    Izewska, Joanna
    Reft, Chester
    MEDICAL PHYSICS, 2013, 40 (07)
  • [28] The role of in vivo dosimetry in modern radiotherapy
    Bulski, W
    Rostkowska, J
    Kania, M
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S358 - S358
  • [29] THE ROLE OF IN VIVO DOSIMETRY IN MODERN RADIOTHERAPY
    Bulski, W.
    Rostkowska, J.
    Kania, M.
    Walewska, A.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S401 - S401
  • [30] In vivo dosimetry with alanine/electron spin resonance dosimetry to evaluate the urethra dose during high-dose-rate brachytherapy
    Wagner, Daniela
    Herrnann, Markus
    Hille, Andrea
    BRACHYTHERAPY, 2017, 16 (04) : 815 - 821