Bipolar radiofrequency ablation lesion areas and confluence: An ex vivo study and technical report

被引:2
|
作者
Wondra, Andrew [1 ,2 ]
Gulhar, Shelly [1 ,2 ]
Brestle, Mason D. [3 ,4 ]
Chou, Jiling [5 ]
Desai, Mehul J. [3 ,6 ,7 ]
机构
[1] MedStar Georgetown Univ, Dept Rehabil Med, Washington, DC USA
[2] MedStar Natl Rehabil Hosp, Washington, DC USA
[3] Int Spine Pain & Performance Ctr, Washington, DC USA
[4] Philadelphia Coll Osteopath Med, Suwanee, GA USA
[5] MedStar Hlth Res Inst, Hyattsville, MD USA
[6] George Washington Univ, Sch Med & Hlth Sci, Washington, DC USA
[7] Int Spine Pain & Performance Ctr, 2021 K St NW,Suite 615, Washington, DC 20006 USA
关键词
bipolar; interprobe distance; lesion confluence; lesion midpoint area; radiofrequency ablation; FACET JOINT DENERVATION; LOW-BACK-PAIN; DOUBLE-BLIND; NEUROTOMY; VOLUMES; ENERGY;
D O I
10.1111/papr.13323
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
IntroductionRadiofrequency ablation (RFA) has been used for nearly 100 years, treating an array of medical conditions including chronic pain. Radiofrequency (RF) energy depolarizes and repolarizes tissues adjacent to a probe producing heat and causing direct thermal injury. When positioned adjacent to neural structures, it leads to neural tissue injury and cell death interrupting pain signaling with the ultimate goal of providing lasting pain relief. Today, RFA is commonly used to treat cervical, thoracic, and lumbar zygapophyseal joints, sacroiliac joint, and more recently large peripheral joint-mediated pain. There are several applications of RFA systems, including bipolar, conventional thermal, cooled, protruding, and pulsed. As yet, no study has determined the best technical practice for bipolar RFA.ObjectiveThis ex vivo study examines RFA lesion midpoint (LMP) area and lesion confluence comparing three different commonly used gauge (g) probes (18-g, 20-g, and 22-g) with 10-mm active tips at various interprobe distances (IPD) to guide best technical practices for its clinical application.MethodsBipolar RFA lesions were generated in preservative-free chicken breast specimens using three different gauge probes (18-g, 20-g, and 22-g) with 10-mm active tips at various IPD (6, 8, 10, 12, 14, 16, and 18 mm). RF was applied for 105 s (15-s ramp time) at 80 degrees C for each lesion at both room and human physiological temperature. The specimen tissues were dissected through the lesion to obtain a length, width, and depth, which were used to calculate the LMP area (mm2). The LMP areas of each thermal ablation were investigated using visualization and descriptive analysis. The Kruskal-Wallis test was performed to compare LMP areas between the two temperature groups and the three different gauge probe subgroups at the various IPDs.ResultsOf the 36 RF lesions (14: 18-g, 12: 20-g, and 10: 22-g) performed, 24 demonstrated lesion confluence. The average time to reach 80 degrees C was 16-17 s; therefore, the average time of RF-energy delivery (at goal temperature) was 88-89 s despite varying needle size or IPD. Comparing the 25 and 37 degrees C groups, 18-g probes produced mean LMP areas of 73.7 and 79.2 mm2, respectively; 20-g probes produced mean LMP areas of 66 and 66.8 mm2, respectively; 22-g probes produced mean LMP areas of 56.6 and 59.7 mm2, respectively. There was no statistical evidence to state a difference regarding LMP area between temperature groups; however, the 18-g probes produced consistently larger LMP areas in the 37 degrees C compared to 25 degrees C specimen groups at each IPD. Lesion confluence was lost for 18-g, 20-g, and 22-g probes at IPD of 14, 12, and 10 mm, respectively, in both 25 and 37 degrees C groups. LMP area was similar between 6 and 8 mm IPD in all of the three-gauge groups; however, there was a significant drop in LMP area from 8 mm IPD to 10 mm and greater. The 18-g, 20-g, and 22-g probes all demonstrated a sharp decline in LMP area when increasing the IPD from 8 to 10 mm.ConclusionThis ex vivo technical study evaluated bipolar RFA LMP areas and lesion confluence, and determined the recommended IPD of 18-g, 20-g, and 22-g probes to be less than 12, 10, and 8 mm, respectively, for best clinical practice. Placing bipolar probes at an IPD greater than 14, 12, and 10 mm, respectively, risks the loss of lesion confluence and failure to produce a clinically significant treatment response due to lack of nerve capture. In clinical practice, the use of injectate may produce larger lesions than demonstrated in this study. Additionally, in vivo factors may impact ablation zone size and ablation patterns. As there are a paucity of studies comparing various RFA applications and conventional RFA needles are least expensive, it is possible that bipolar conventional RFA is more cost-effective than other techniques.
引用
收藏
页码:489 / 501
页数:13
相关论文
共 50 条
  • [41] Bipolar radiofrequency ablation of tibialchondroblastomas: A report of three cases
    Rajalakshmi, Prathiba
    Srivastava, Deep N.
    Rastogi, Shishir
    Julka, Pramod Kumar
    Bhatnagar, Sushma
    Gamanagatti, Shivanand
    WORLD JOURNAL OF RADIOLOGY, 2012, 4 (07): : 335 - 340
  • [42] Radiofrequency ablation of subcapsular lesions with internally cooled perfusion electrode: ex vivo study
    Kim, Myung Sub
    Hong, Hyun Pyo
    Kang, Kyung A.
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2022, 31 (02) : 216 - 222
  • [43] Comparison of bipolar and unipolar radiofrequency ablation in an in vivo experimental model
    Bugge, E
    Nicholson, IA
    Thomas, SP
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 28 (01) : 76 - 80
  • [44] The effect of radiofrequency ablation on different organs: Ex vivo and in vivo comparative studies
    Kim, Yoo Na
    Rhim, Hyunchul
    Choi, Dongil
    Kim, Young-sun
    Lee, Min Woo
    Chang, Ilsoo
    Lee, Won Jae
    Lim, Hyo K.
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (02) : 526 - 532
  • [45] Multipolar Ablation With Simultaneous versus Bipolar Radiofrequency Delivery Have Different Ablation Lesion Characteristics
    Borne, Ryan
    Sauer, William H.
    Zipse, Matthew
    Zheng, Lijun
    Tzou, Wendy
    Nguyen, Duy T.
    CIRCULATION, 2017, 136
  • [46] Heat Sink Effect on Tumor Ablation Characteristics as Observed in Monopolar Radiofrequency, Bipolar Radiofrequency, and Microwave, Using Ex Vivo Calf Liver Model
    Pillai, Krishna
    Akhter, Javid
    Chua, Terence C.
    Shehata, Mena
    Alzahrani, Nayef
    Al-Alem, Issan
    Morris, David L.
    MEDICINE, 2015, 94 (09) : e580
  • [47] Juxtametallic Bipolar Bone Radiofrequency Ablation: Thermal Monitoring in an Ex-Vivo Model with Specimen MRI and Histopathologic Correlation
    Sag, Alan A.
    Sperduto, William A. L.
    Eward, William
    Ronald, James
    Davis, Haley
    Jiang, Xiaoyin Sara
    Enterline, David S.
    Visgauss, Julia
    Brigman, Brian
    Goodwin, C. Rory
    Qadri, Yawar J.
    Kim, Charles Y.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 33 (12) : 1594 - 1600
  • [48] Radiofrequency Ablation of Spine An Experimental Study in an Ex Vivo Bovine and In Vivo Swine Model for Feasibility in Spine Tumor
    You, Nam Kyu
    Lee, Hye Young
    Shin, Dong Ah
    Choi, Gwi Hyun
    Yi, Seong
    Kim, Keung Nyun
    Yoon, Do Heum
    Park, Jon
    SPINE, 2013, 38 (18) : E1121 - E1127
  • [49] Hepatic lesion ablation with bipolar saline-enhanced radiofrequency in the audible spectrum
    Burdío, F
    Güemes, A
    Burdío, JM
    Castiella, T
    De Gregorio, MA
    Lozano, R
    Livraghi, T
    ACADEMIC RADIOLOGY, 1999, 6 (11) : 680 - 686
  • [50] In vivo Lesion Index (LSI) validation in percutaneous radiofrequency catheter ablation
    Themistoclakis, Sakis
    Calzolari, Vittorio
    De Mattia, Luca
    China, Paolo
    Dello Russo, Antonio
    Fassini, Gaetano
    Casella, Michela
    Caporaso, Igor
    Indiani, Stefano
    Addis, Alessandro
    Basso, Cristina
    Della Barbera, Mila
    Thiene, Gaetano
    Tondo, Claudio
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (05) : 874 - 882