Bipolar radiofrequency ablation lesion areas and confluence: An ex vivo study and technical report

被引:2
|
作者
Wondra, Andrew [1 ,2 ]
Gulhar, Shelly [1 ,2 ]
Brestle, Mason D. [3 ,4 ]
Chou, Jiling [5 ]
Desai, Mehul J. [3 ,6 ,7 ]
机构
[1] MedStar Georgetown Univ, Dept Rehabil Med, Washington, DC USA
[2] MedStar Natl Rehabil Hosp, Washington, DC USA
[3] Int Spine Pain & Performance Ctr, Washington, DC USA
[4] Philadelphia Coll Osteopath Med, Suwanee, GA USA
[5] MedStar Hlth Res Inst, Hyattsville, MD USA
[6] George Washington Univ, Sch Med & Hlth Sci, Washington, DC USA
[7] Int Spine Pain & Performance Ctr, 2021 K St NW,Suite 615, Washington, DC 20006 USA
关键词
bipolar; interprobe distance; lesion confluence; lesion midpoint area; radiofrequency ablation; FACET JOINT DENERVATION; LOW-BACK-PAIN; DOUBLE-BLIND; NEUROTOMY; VOLUMES; ENERGY;
D O I
10.1111/papr.13323
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
IntroductionRadiofrequency ablation (RFA) has been used for nearly 100 years, treating an array of medical conditions including chronic pain. Radiofrequency (RF) energy depolarizes and repolarizes tissues adjacent to a probe producing heat and causing direct thermal injury. When positioned adjacent to neural structures, it leads to neural tissue injury and cell death interrupting pain signaling with the ultimate goal of providing lasting pain relief. Today, RFA is commonly used to treat cervical, thoracic, and lumbar zygapophyseal joints, sacroiliac joint, and more recently large peripheral joint-mediated pain. There are several applications of RFA systems, including bipolar, conventional thermal, cooled, protruding, and pulsed. As yet, no study has determined the best technical practice for bipolar RFA.ObjectiveThis ex vivo study examines RFA lesion midpoint (LMP) area and lesion confluence comparing three different commonly used gauge (g) probes (18-g, 20-g, and 22-g) with 10-mm active tips at various interprobe distances (IPD) to guide best technical practices for its clinical application.MethodsBipolar RFA lesions were generated in preservative-free chicken breast specimens using three different gauge probes (18-g, 20-g, and 22-g) with 10-mm active tips at various IPD (6, 8, 10, 12, 14, 16, and 18 mm). RF was applied for 105 s (15-s ramp time) at 80 degrees C for each lesion at both room and human physiological temperature. The specimen tissues were dissected through the lesion to obtain a length, width, and depth, which were used to calculate the LMP area (mm2). The LMP areas of each thermal ablation were investigated using visualization and descriptive analysis. The Kruskal-Wallis test was performed to compare LMP areas between the two temperature groups and the three different gauge probe subgroups at the various IPDs.ResultsOf the 36 RF lesions (14: 18-g, 12: 20-g, and 10: 22-g) performed, 24 demonstrated lesion confluence. The average time to reach 80 degrees C was 16-17 s; therefore, the average time of RF-energy delivery (at goal temperature) was 88-89 s despite varying needle size or IPD. Comparing the 25 and 37 degrees C groups, 18-g probes produced mean LMP areas of 73.7 and 79.2 mm2, respectively; 20-g probes produced mean LMP areas of 66 and 66.8 mm2, respectively; 22-g probes produced mean LMP areas of 56.6 and 59.7 mm2, respectively. There was no statistical evidence to state a difference regarding LMP area between temperature groups; however, the 18-g probes produced consistently larger LMP areas in the 37 degrees C compared to 25 degrees C specimen groups at each IPD. Lesion confluence was lost for 18-g, 20-g, and 22-g probes at IPD of 14, 12, and 10 mm, respectively, in both 25 and 37 degrees C groups. LMP area was similar between 6 and 8 mm IPD in all of the three-gauge groups; however, there was a significant drop in LMP area from 8 mm IPD to 10 mm and greater. The 18-g, 20-g, and 22-g probes all demonstrated a sharp decline in LMP area when increasing the IPD from 8 to 10 mm.ConclusionThis ex vivo technical study evaluated bipolar RFA LMP areas and lesion confluence, and determined the recommended IPD of 18-g, 20-g, and 22-g probes to be less than 12, 10, and 8 mm, respectively, for best clinical practice. Placing bipolar probes at an IPD greater than 14, 12, and 10 mm, respectively, risks the loss of lesion confluence and failure to produce a clinically significant treatment response due to lack of nerve capture. In clinical practice, the use of injectate may produce larger lesions than demonstrated in this study. Additionally, in vivo factors may impact ablation zone size and ablation patterns. As there are a paucity of studies comparing various RFA applications and conventional RFA needles are least expensive, it is possible that bipolar conventional RFA is more cost-effective than other techniques.
引用
收藏
页码:489 / 501
页数:13
相关论文
共 50 条
  • [1] Bipolar Radiofrequency Ablation of the Adrenal Gland: In Vivo and Ex Vivo Experiments
    Kinoshita, Tomo
    Seiji, Kazumasa
    Oguro, Sota
    Ono, Yoshikiyo
    Morimoto, Ryo
    Satoh, Fumitoshi
    Takase, Kei
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 2023, 259 (02): : 127 - 133
  • [2] Heat sink effects in thyroid bipolar radiofrequency ablation: an ex vivo study
    Klimek, Konrad
    Mader, Nicolai
    Happel, Christian
    Sabet, Amir
    Grunwald, Frank
    Groener, Daniel
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [3] Direct Current Combined With Bipolar Radiofrequency Ablation: An Ex Vivo Feasibility Study
    Tanaka, Toshihiro
    Penzkofer, Tobias
    Isfort, Peter
    Bruners, Philipp
    Disselhorst-Klug, Catherine
    Junker, Elmar
    Kichikawa, Kimihiko
    Schmitz-Rode, Thomas
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 34 (03) : 631 - 636
  • [4] Heat sink effects in thyroid bipolar radiofrequency ablation: an ex vivo study
    Konrad Klimek
    Nicolai Mader
    Christian Happel
    Amir Sabet
    Frank Grünwald
    Daniel Groener
    Scientific Reports, 13
  • [5] Direct Current Combined With Bipolar Radiofrequency Ablation: An Ex Vivo Feasibility Study
    Toshihiro Tanaka
    Tobias Penzkofer
    Peter Isfort
    Philipp Bruners
    Catherine Disselhorst-Klug
    Elmar Junker
    Kimihiko Kichikawa
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2011, 34 : 631 - 636
  • [6] Ex Vivo Evaluation of a Bipolar Application Concept for Radiofrequency Ablation
    Zurbuchen, Urte
    Frericks, Bernd
    Roggan, Andre
    Lehmann, Kai
    Boessenroth, Daniela
    Buhr, Heinz J.
    Ritz, Joerg-Peter
    ANTICANCER RESEARCH, 2009, 29 (04) : 1309 - 1314
  • [7] Fluid-modulated bipolar radiofrequency ablation:: An ex-vivo evaluation study
    Bruners, P.
    Mueller, H.
    Guenther, R. W.
    Schmitz-Rode, T.
    Mahnken, A. H.
    ACTA RADIOLOGICA, 2008, 49 (03) : 258 - 266
  • [8] Ex vivo human bile duct radiofrequency ablation with a bipolar catheter
    Mustafa Atar
    Abdurrahman Kadayifci
    Ebubekir Daglilar
    Catherine Hagen
    Carlos Fernandez-del Castillo
    William R. Brugge
    Surgical Endoscopy, 2018, 32 : 2808 - 2813
  • [9] Ex vivo human bile duct radiofrequency ablation with a bipolar catheter
    Atar, Mustafa
    Kadayifci, Abdurrahman
    Daglilar, Ebubekir
    Hagen, Catherine
    Fernandez-del Castillo, Carlos
    Brugge, William R.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2018, 32 (06): : 2808 - 2813