Accuracy of low-cost alternative facial scanners: a prospective cohort study

被引:10
|
作者
Bartella, Alexander K. [1 ]
Laser, Josefine [1 ]
Kamal, Mohammad [2 ]
Krause, Matthias [1 ]
Neuhaus, Michael [1 ]
Pausch, Niels C. [1 ]
Sander, Anna K. [1 ]
Lethaus, Bernd [1 ]
Zimmerer, Rudiger [1 ]
机构
[1] Univ Leipzig, Dept Oral & Maxillofacial Surg, Liebigstr 12, D-04103 Leipzig, Germany
[2] Kuwait Univ, Fac Dent, Dept Surg Sci, Safat, Kuwait
来源
关键词
Face scan; Three-dimensional face scan; Photogrammetry; Smart device; iPhone; SYSTEM; MOBILE;
D O I
10.1007/s10006-022-01050-5
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction Three-dimensional facial scans have recently begun to play an increasingly important role in the peri-therapeutic management of oral and maxillofacial and head and neck surgery cases. Face scan images can be generated by optical facial scanners utilizing line-laser, stereophotography, or structured light modalities, as well as from volumetric data: for example, from cone beam computed tomography (CBCT). This study aimed to evaluate whether two low-cost procedures for the creation of three-dimensional face scan images were capable of producing sufficiently accurate data sets for clinical analysis. Materials and methods Fifty healthy volunteers were included in the study. Two test objects with defined dimensions (Lego bricks) were attached to the forehead and the left cheek of each volunteer. Facial anthropometric values (i.e., the distances between the medial canthi, the lateral canthi, the nasal alae, and the angles of the mouth) were first measured manually. Subsequently, face scans were performed with a smart device and manual photogrammetry and the values obtained were compared with the manually measured data sets. Results The anthropometric distances deviated, on average, 2.17 mm from the manual measurements (smart device scanning deviation 3.01 mm, photogrammetry deviation 1.34 mm), with seven out of eight deviations being statistically significant. For the Lego brick, from a total of 32 angles, 19 values demonstrated a significant difference from the original 90 degrees angles. The average deviation was 6.5 degrees (smart device scanning deviation 10.1 degrees, photogrammetry deviation 2.8 degrees). Conclusion Manual photogrammetry demonstrated greater accuracy when creating three-dimensional face scan images; however, smart devices are more user-friendly. Dental professionals should monitor camera and smart device technical improvements carefully when choosing and adequate technique for 3D scanning.
引用
收藏
页码:33 / 41
页数:9
相关论文
共 50 条
  • [21] A low-cost alternative to atomizer in talc pleurodesis
    Jindal, Rohit
    Mishra, Atul
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2020, 152 (07) : 251 - 251
  • [22] Immersion tin - A low-cost alternative to HASL
    Langan, JP
    PLATING AND SURFACE FINISHING, 1997, 84 (12): : 46 - 47
  • [23] CWDM: A low-cost alternative for increased capacity
    Richard, JA
    PHOTONICS SPECTRA, 2003, 37 (05) : 62 - 64
  • [24] Seabird offers low-cost helicopter alternative
    Aviation Week and Space Technology (New York), 1994, 140 (10):
  • [25] A Novel, Low-Cost Alternative to Traditional Glaucoma
    Kong, Cheng F.
    Yek, John
    Clegg, Philippa
    Graham, Kathleen
    Gupta, Rohan
    Boughton, Philip
    Billson, Mark
    White, Andrew
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2024, 13 (08): : 1 - 8
  • [26] Depth accuracy of the RealSense F200: Low-cost 4D facial imaging
    ten Harkel, Timen C.
    Speksnijder, Caroline M.
    van der Heijden, Ferdinand
    Beurskens, Carien H. G.
    Ingels, Koen J. A. O.
    Maal, Thomas J. J.
    SCIENTIFIC REPORTS, 2017, 7
  • [27] Depth accuracy of the RealSense F200: Low-cost 4D facial imaging
    Timen C. ten Harkel
    Caroline M. Speksnijder
    Ferdinand van der Heijden
    Carien H. G. Beurskens
    Koen J. A. O. Ingels
    Thomas J. J. Maal
    Scientific Reports, 7
  • [28] Low-Cost Ironmaking: Metallurgical Coal Prospective
    H. P. Naresh Sharma
    Coke and Chemistry, 2023, 66 : 641 - 658
  • [29] Low-Cost Ironmaking: Metallurgical Coal Prospective
    Sharma, Naresh
    Tiwari, H. P.
    COKE AND CHEMISTRY, 2023, 66 (12) : 641 - 658
  • [30] Pregnancy rate after icsi compared to a low-cost simplified ivf system in a selected population: A prospective cohort study
    Vertessen, V.
    Ombelet, W.
    Van der Auwera, I.
    Janssen, M.
    Onofre, J.
    Dhont, N.
    Campo, R.
    EUROPEAN JOURNAL OF OBSTETRICS & GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 2022, 270 : E88 - E88