Clinical application of breathing-adapted 4D CT: image quality comparison to conventional 4D CT

被引:7
|
作者
Werner, Rene [1 ]
Szkitsak, Juliane [2 ,3 ]
Madesta, Frederic [1 ]
Buettgen, Laura [1 ]
Wimmert, Lukas [1 ]
Sentker, Thilo [1 ]
Fietkau, Rainer [2 ,3 ]
Haderlein, Marlen [2 ,3 ]
Bert, Christoph [2 ,3 ]
Gauer, Tobias [1 ]
Hofmann, Christian [4 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, D-20246 Hamburg, Germany
[2] Univ Klinikum Erlangen, Dept Radiat Oncol, D-91054 Erlangen, Germany
[3] Friedrich Alexander Univ Erlangen Nurnberg, D-91054 Erlangen, Germany
[4] Siemens Healthcare GmbH, D-91301 Forchheim, Germany
关键词
4D radiotherapy; 4D CT; Image artifacts; Image quality; LUNG; ARTIFACTS; GUIDANCE; STRATEGY;
D O I
10.1007/s00066-023-02062-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: 4D CT imaging is an integral part of 4D radiotherapy workflows. However, 4D CT data often contain motion artifacts that mitigate treatment planning. Recently, breathing-adapted 4D CT (i4DCT) was introduced into clinical practice, promising artifact reduction in in-silico and phantom studies. Here, we present an image quality comparison study, pooling clinical patient data from two centers: a new i4DCT and a conventional spiral 4D CT patient cohort.Methods: The i4DCT cohort comprises 129 and the conventional spiral 4D CT cohort 417 4D CT data sets of lung and liver tumor patients. All data were acquired for treatment planning. The study consists of three parts: illustration of image quality in selected patients of the two cohorts with similar breathing patterns; an image quality expert rater study; and automated analysis of the artifact frequency.Results: Image data of the patients with similar breathing patterns underline artifact reduction by i4DCT compared to conventional spiral 4D CT. Based on a subgroup of 50 patients with irregular breathing patterns, the rater study reveals a fraction of almost artifact-free scans of 89% for i4DCT and only 25% for conventional 4D CT; the quantitative analysis indicated a reduction of artifact frequency by 31% for i4DCT.Conclusion: The results demonstrate 4D CT image quality improvement for patients with irregular breathing patterns by breathing-adapted 4D CT in this first corresponding clinical data image quality comparison study.
引用
收藏
页码:686 / 691
页数:6
相关论文
共 50 条
  • [11] Optimization of 4D PET/CT
    Pan, Tinsu
    Mawlawi, Osama
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [12] 4D pet/ct in oncology
    Maranzana, D.
    Muni, A.
    Gandini, O.
    Muratori, M.
    Sumatra, M.
    Tornambe, G.
    Zanaga, M.
    Orsi, K.
    Valentini, D.
    Testori, O.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S456 - S456
  • [13] 4D PET and 4D CT image fusion for accurate radiotherapy planning of lung cancer patients
    Wolthaus, JWH
    van Herk, M
    Muller, SH
    Bois, DJA
    Rossi, MMG
    Belderbos, JSA
    Lebesque, JV
    Damen, EMF
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S162 - S163
  • [14] 4D Radiomics: Impact of 4D Image Quality On Radiomic Analysis
    Zhang, Z.
    Huang, M.
    Jiang, Z.
    Chang, Y.
    Torok, J.
    Yin, F.
    Ren, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E312 - E313
  • [15] 4D Flow Meets CT: Can It Compete with 4D Flow MRI?
    Schoepf, U. Joseph
    Varga-Szemes, Akos
    RADIOLOGY, 2018, 289 (01) : 59 - 60
  • [16] ENHANCED 4D PET OPTIMIZATION BASED ON 4D CT MOTION MODELING
    Gianoli, C.
    Fontana, G.
    Riboldi, M.
    Cavedon, C.
    Baroni, G.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S67 - S68
  • [17] ADVANCES IN DEFORMABLE IMAGE REGISTRATION OF THORACIC 4D CT
    Sarrut, D.
    Vandemeulebroucke, J.
    Rit, S.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S188 - S189
  • [18] A multiple points method for 4D CT image sorting
    Gianoli, Chiara
    Riboldi, Marco
    Spadea, Maria Francesca
    Travaini, Laura Lavinia
    Ferrari, Mahila
    Mei, Riccardo
    Orecchia, Roberto
    Baroni, Guido
    MEDICAL PHYSICS, 2011, 38 (02) : 656 - 667
  • [19] 4D CT image reconstruction with diffeomorphic motion model
    Hinkle, Jacob
    Szegedi, Martin
    Wang, Brian
    Salter, Bill
    Joshi, Sarang
    MEDICAL IMAGE ANALYSIS, 2012, 16 (06) : 1307 - 1316
  • [20] How to explore alternative 4D CT image binning
    Fitzpatrick, MJ
    Starkschall, G
    Balter, P
    Pan, T
    MEDICAL PHYSICS, 2005, 32 (06) : 1924 - 1924