Clinical application of breathing-adapted 4D CT: image quality comparison to conventional 4D CT

被引:7
|
作者
Werner, Rene [1 ]
Szkitsak, Juliane [2 ,3 ]
Madesta, Frederic [1 ]
Buettgen, Laura [1 ]
Wimmert, Lukas [1 ]
Sentker, Thilo [1 ]
Fietkau, Rainer [2 ,3 ]
Haderlein, Marlen [2 ,3 ]
Bert, Christoph [2 ,3 ]
Gauer, Tobias [1 ]
Hofmann, Christian [4 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, D-20246 Hamburg, Germany
[2] Univ Klinikum Erlangen, Dept Radiat Oncol, D-91054 Erlangen, Germany
[3] Friedrich Alexander Univ Erlangen Nurnberg, D-91054 Erlangen, Germany
[4] Siemens Healthcare GmbH, D-91301 Forchheim, Germany
关键词
4D radiotherapy; 4D CT; Image artifacts; Image quality; LUNG; ARTIFACTS; GUIDANCE; STRATEGY;
D O I
10.1007/s00066-023-02062-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: 4D CT imaging is an integral part of 4D radiotherapy workflows. However, 4D CT data often contain motion artifacts that mitigate treatment planning. Recently, breathing-adapted 4D CT (i4DCT) was introduced into clinical practice, promising artifact reduction in in-silico and phantom studies. Here, we present an image quality comparison study, pooling clinical patient data from two centers: a new i4DCT and a conventional spiral 4D CT patient cohort.Methods: The i4DCT cohort comprises 129 and the conventional spiral 4D CT cohort 417 4D CT data sets of lung and liver tumor patients. All data were acquired for treatment planning. The study consists of three parts: illustration of image quality in selected patients of the two cohorts with similar breathing patterns; an image quality expert rater study; and automated analysis of the artifact frequency.Results: Image data of the patients with similar breathing patterns underline artifact reduction by i4DCT compared to conventional spiral 4D CT. Based on a subgroup of 50 patients with irregular breathing patterns, the rater study reveals a fraction of almost artifact-free scans of 89% for i4DCT and only 25% for conventional 4D CT; the quantitative analysis indicated a reduction of artifact frequency by 31% for i4DCT.Conclusion: The results demonstrate 4D CT image quality improvement for patients with irregular breathing patterns by breathing-adapted 4D CT in this first corresponding clinical data image quality comparison study.
引用
收藏
页码:686 / 691
页数:6
相关论文
共 50 条
  • [1] Clinical application of breathing-adapted 4D CT: image quality comparison to conventional 4D CT
    René Werner
    Juliane Szkitsak
    Frederic Madesta
    Laura Büttgen
    Lukas Wimmert
    Thilo Sentker
    Rainer Fietkau
    Marlen Haderlein
    Christoph Bert
    Tobias Gauer
    Christian Hofmann
    Strahlentherapie und Onkologie, 2023, 199 : 686 - 691
  • [2] Comparison of intelligent 4D CT sequence scanning and conventional spiral 4D CT: A first comprehensive phantom study
    Werner R.
    Szkitsak J.
    Sentker T.
    Madesta F.
    Schwarz A.
    Fernolendt S.
    Vornehm M.
    Gauer T.
    Bert C.
    Hofmann C.
    Physics in Medicine and Biology, 2021, 66 (01):
  • [3] Evaluation of Liver Deformation During Breathing Using Deformable Image Registration: A Comparison Between 4D CT and 4D MRI
    Liang, X.
    Czito, B.
    Palta, M.
    Bashir, M.
    Yin, F.
    Cai, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 171 - 171
  • [4] Intracardiac Flow at 4D CT: Comparison with 4D Flow MRI
    Lantz, Jonas
    Gupta, Vikas
    Henriksson, Lilian
    Karlsson, Matts
    Persson, Anders
    Carlhall, Carl-Johan
    Ebbers, Tino
    RADIOLOGY, 2018, 289 (01) : 51 - 58
  • [5] Automated breathing motion modeling for 4D CT
    ElNaqa, I
    Low, D
    Deasy, J
    Amini, A
    Nystrom, M
    MEDICAL PHYSICS, 2003, 30 (06) : 1415 - 1415
  • [6] An Approach to Convert 4D Geometry into a 4D CT Scan
    Villard, P. F.
    Beuve, M.
    Shariat, B.
    WSCG 2006: SHORT PAPERS PROCEEDINGS: 14TH INTERNATIONAL CONFERENCE IN CENTRAL EUROPE ON COMPUTER GRAPHICS, VISUALIZATION AND COMPUTER VISION 2006, 2006, : 163 - 169
  • [7] Comparison of 4D cone beam CT and 4D spiral scan CT using a dedicated chest phantom
    Nill, S.
    Jetter, S.
    Tacking, T.
    Moser, T.
    Remmert, G.
    Biederer, J.
    Fabel-Schulte, M.
    Dinkel, J.
    Oelfke, U.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S213 - S214
  • [8] 4D CT lung ventilation images are affected by the 4D CT sorting method
    Yamamoto, Tokihiro
    Kabus, Sven
    Lorenz, Cristian
    Johnston, Eric
    Maxim, Peter G.
    Diehn, Maximilian
    Eclov, Neville
    Barquero, Cristian
    Loo, Billy W., Jr.
    Keall, Paul J.
    MEDICAL PHYSICS, 2013, 40 (10)
  • [9] Reduce Artifacts for Helical 4D CT Image
    Han, D.
    Bayouth, J.
    Bhatia, S.
    Sonka, M.
    Wu, X.
    MEDICAL PHYSICS, 2010, 37 (06) : 3126 - +
  • [10] Registration of 4D CBCT and 4D CT for extracranial stereotactic treatments
    Elder, E.
    Schreibmann, E.
    Li, T.
    Fox, T.
    Xing, L.
    Crocker, J.
    Landry, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S651 - S651