V Flow Measurements of Pulsatile Flow in Femoral-Popliteal Bypass Proximal Anastomosis Compared with CFD Simulation

被引:1
|
作者
Yukhnev, Andrey [1 ]
Tikhomolova, Ludmila [1 ]
Gataulin, Yakov [1 ]
Marinova, Alexandra [1 ]
Smirnov, Evgueni [1 ]
Vrabiy, Andrey [2 ]
Suprunovich, Andrey [2 ]
Khubulava, Gennady [2 ]
机构
[1] Peter Great St Petersburg Polytech Univ, Higher Sch Appl Math & Computat Phys, 29 Polytechnicheskaya Str, St Petersburg 195251, Russia
[2] Pavlov First St Petersburg State Med Univ, Dept Fac Surg, 6-8 Lva Tolstogo Str, St Petersburg 197022, Russia
基金
俄罗斯科学基金会;
关键词
pulsatile blood flow; 3D numerical simulation; 2D ultrasound vector imaging; femoral popliteal bypass surgery; proximal anastomosis; personalized model; WALL SHEAR-STRESS; BLOOD-FLOW; VESSEL;
D O I
10.3390/fluids9030064
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
This paper presents the experience of using the V Flow high-frame-rate ultrasound vector imaging method to study the pulsatile velocity fields in the area of the proximal anastomosis for femoral popliteal bypass surgery in vitro and in vivo. A representative (average) anastomosis model and the experimental setup designed for in vitro studies covering forward and reverse flow phases throughout the cycle are described. The results of the measurements are presented for areas with a relatively uniform velocity distribution and for areas with pronounced spatial inhomogeneities due to the jet or recirculating nature of the flow. The results of ultrasonic studies of the velocity field of the three-dimensional pulsatile flow in vitro and in vivo are compared with the data of numerical simulations carried out for the average and personalized models based on the Navier-Stokes equations. Acceptable consistency between the results of experimental and numerical studies is demonstrated.