Stroke Volume Measurements by Echocardiography and Ultrasonic Cardiac Output Monitor in Children

被引:0
|
作者
Fremuth, Jiri [1 ,2 ]
Huml, Michal [1 ]
Pomahacova, Tereza [1 ]
Kobr, Jiri [1 ]
Kormunda, Stanislav [1 ]
Sykora, Josef [1 ]
机构
[1] Charles Univ Prague, Fac Med Pilsen, Fac Hosp Pilsen, Dept Pediat, Prague, Czech Republic
[2] Charles Univ Prague, Fac Hosp Pilsen, Fac Med Pilsen, Dept Pediat, Alej Svobody 80, Plzen 30460, Czech Republic
关键词
ultrasonic cardiac output monitor; echocardiography; cardiac output; METAANALYSIS; PRECISION; INDEX;
D O I
10.1097/PEC.0000000000003018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
ObjectivesStroke volume (SV) and cardiac output monitoring is a cornerstone of hemodynamic assessment. Noninvasive technologies are increasingly used in children. This study compared SV measurements obtained by transcutaneous Doppler ultrasound techniques (ultrasonic cardiac output monitor [USCOM]), transthoracic echocardiography jugular (TTE-J), and parasternal (TTE-P) views performed by pediatric intensivists (OP-As) with limited training in cardiac sonography (20 previous examinations) and pediatric cardiologists (OP-Bs) with limited training in USCOM (30 previous examinations) in spontaneously ventilating children.MethodsA single-center study was conducted in 37 children. Each operator obtained 3 sets of USCOM SV measurements within a period of 3 to 5 minutes, followed with TTE measurements from both apical and jugular views. The investigators were blinded to each other's results to prevent visual and auditory bias.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ConclusionsOur findings show that the methods are not interchangeable because SV values by USCOM are higher in comparison with the SV values obtained by TTE. Both methods have low level of intraobserver variability. The SV measurements obtained by TTE-P were significantly lower compared with the TTE-J for the operator with limited training in echocardiography. The TTE-P requires longer practice compared with the TTE-J; therefore, we recommend to prefer TTE-J to TTE-P for inexperienced operators.
引用
收藏
页码:680 / 684
页数:5
相关论文
共 50 条
  • [21] STROKE VOLUME AND CARDIAC-OUTPUT MEASUREMENT WITH A NEW CATHETER TIP ULTRASONIC INSTRUMENT
    MARTIN, RW
    POLLACK, GH
    PHILLIPS, J
    CIRCULATION, 1975, 52 (04) : 213 - 213
  • [22] MEASUREMENTS OF CARDIAC-OUTPUT AND STROKE VOLUME IN MONKEYS BY THORACIC TETRAPOLAR RHEOGRAPHY
    BELKANIA, GS
    DARTSMELIA, VA
    KOSMICHESKAYA BIOLOGIYA I AVIAKOSMICHESKAYA MEDITSINA, 1983, 17 (04): : 75 - 78
  • [23] Cardiac Output Measurement Using the Ultrasonic Cardiac Output Monitor: A Validation Study in Newborn Infants
    Fraga, Maria, V
    Dysart, Kevin C.
    Rintoul, Natalie
    Chaudhary, Aasma S.
    Ratcliffe, Sarah J.
    Fedec, Anysia
    Kren, Stephanie
    Cohen, Meryl S.
    Kirpalani, Haresh
    NEONATOLOGY, 2019, 116 (03) : 260 - 268
  • [24] The Use of Epinephrine in Caudal Anesthesia Increases Stroke Volume and Cardiac Output in Children
    Liu, Chang Amber
    Sui, Jinghu
    Cote, Charles J.
    Anderson, Thomas A.
    REGIONAL ANESTHESIA AND PAIN MEDICINE, 2016, 41 (06) : 780 - 786
  • [25] Use of epinephrine in caudal anaesthesia increases stroke volume and cardiac output in children
    Liu, C. A.
    Sui, J.
    Cote, C. J.
    Anderson, T. A.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 117 : 74 - 74
  • [26] Cardiac output measurement in children:: comparison of Aesculon® cardiac output monitor and thermodilution
    Tomaske, M.
    Knirsch, W.
    Kretschmar, O.
    Woitzek, K.
    Balmer, C.
    Schmitz, A.
    Bauersfeld, U.
    Weiss, M.
    BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) : 517 - 520
  • [27] An assessment of the accuracy of non-invasive measurements of cardiac output in patients requiring intensive care using an ultrasonic cardiac output monitor (the USCOM®)
    Boyle, Martin
    Steel, Liz
    Murgo, Margherita
    Nicholson, Lisa
    O'Brien, Maureen
    Flynn, Gordon
    Bihari, David
    AUSTRALIAN CRITICAL CARE, 2009, 22 (01) : 47 - 47
  • [28] TRANSPULMONARY THERMODILUTION VERSUS TRANSTHORACIC ECHOCARDIOGRAPHY FOR CARDIAC OUTPUT MEASUREMENTS IN SEVERELY BURNED CHILDREN
    Wurzer, Paul
    Branski, Ludwik K.
    Jeschke, Marc G.
    Ali, Arham
    Kinsky, Michael P.
    Bohanon, Fredrick J.
    Hundeshagen, Gabriel
    Norbury, William B.
    Williams, Felicia N.
    Kamolz, Lars-P.
    Finnerty, Celeste C.
    Herndon, David N.
    SHOCK, 2016, 46 (03): : 249 - 253
  • [29] Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement
    Walter Knirsch
    Oliver Kretschmar
    Maren Tomaske
    Kathrina Stutz
    Nicole Nagdyman
    Christian Balmer
    Achim Schmitz
    Dominique Béttex
    Felix Berger
    Urs Bauersfeld
    Markus Weiss
    Intensive Care Medicine, 2008, 34 : 1060 - 1064
  • [30] Cardiac output measurement in children:: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement
    Knirsch, Walter
    Kretschmar, Oliver
    Tomaske, Maren
    Stutz, Kathrina
    Nagdyman, Nicole
    Balmer, Christian
    Schmitz, Achim
    Bettex, Dominique
    Berger, Felix
    Bauersfeld, Urs
    Weiss, Markus
    INTENSIVE CARE MEDICINE, 2008, 34 (06) : 1060 - 1064