Stroke Volume Measurements by Echocardiography and Ultrasonic Cardiac Output Monitor in Children

被引:0
|
作者
Fremuth, Jiri [1 ,2 ]
Huml, Michal [1 ]
Pomahacova, Tereza [1 ]
Kobr, Jiri [1 ]
Kormunda, Stanislav [1 ]
Sykora, Josef [1 ]
机构
[1] Charles Univ Prague, Fac Med Pilsen, Fac Hosp Pilsen, Dept Pediat, Prague, Czech Republic
[2] Charles Univ Prague, Fac Hosp Pilsen, Fac Med Pilsen, Dept Pediat, Alej Svobody 80, Plzen 30460, Czech Republic
关键词
ultrasonic cardiac output monitor; echocardiography; cardiac output; METAANALYSIS; PRECISION; INDEX;
D O I
10.1097/PEC.0000000000003018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
ObjectivesStroke volume (SV) and cardiac output monitoring is a cornerstone of hemodynamic assessment. Noninvasive technologies are increasingly used in children. This study compared SV measurements obtained by transcutaneous Doppler ultrasound techniques (ultrasonic cardiac output monitor [USCOM]), transthoracic echocardiography jugular (TTE-J), and parasternal (TTE-P) views performed by pediatric intensivists (OP-As) with limited training in cardiac sonography (20 previous examinations) and pediatric cardiologists (OP-Bs) with limited training in USCOM (30 previous examinations) in spontaneously ventilating children.MethodsA single-center study was conducted in 37 children. Each operator obtained 3 sets of USCOM SV measurements within a period of 3 to 5 minutes, followed with TTE measurements from both apical and jugular views. The investigators were blinded to each other's results to prevent visual and auditory bias.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ConclusionsOur findings show that the methods are not interchangeable because SV values by USCOM are higher in comparison with the SV values obtained by TTE. Both methods have low level of intraobserver variability. The SV measurements obtained by TTE-P were significantly lower compared with the TTE-J for the operator with limited training in echocardiography. The TTE-P requires longer practice compared with the TTE-J; therefore, we recommend to prefer TTE-J to TTE-P for inexperienced operators.
引用
收藏
页码:680 / 684
页数:5
相关论文
共 50 条
  • [31] THE STROKE VOLUME AND THE CARDIAC OUTPUT BY THE IMPEDANCE CARDIOGRAPHY
    Ipate, M. C.
    Dobre, A. A.
    Morega, A. M.
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN SERIES C-ELECTRICAL ENGINEERING AND COMPUTER SCIENCE, 2012, 74 (03): : 219 - 232
  • [32] RADIOLOGICAL DETERMINATION OF STROKE VOLUME AND CARDIAC OUTPUT
    NEUMANN, J
    MICHALJANIC, A
    VRLA, J
    ZEITSCHRIFT FUR KREISLAUFFORSCHUNG, 1968, 57 (05): : 445 - +
  • [33] REGULATION OF CARDIAC OUTPUT THROUGH STROKE VOLUME
    WARNER, HR
    TORONTO, AF
    CIRCULATION RESEARCH, 1960, 8 (03) : 549 - 552
  • [34] Stroke volume and cardiac output measurement in cardiac patients during a rehabilitation program: comparison between tonometry, impedancemetry and echocardiography
    Alicia Gonzalez-Represas
    Laurent Mourot
    The International Journal of Cardiovascular Imaging, 2020, 36 : 447 - 455
  • [35] Noninvasive and simultaneous measurements of circulating blood volume and cardiac output in infants and children
    Tanigami, H
    He, YL
    Ueyama, H
    Taniguchi, A
    Mashimo, T
    Yoshiya, I
    ANESTHESIA AND ANALGESIA, 1997, 84 : S274 - S274
  • [36] Stroke volume and cardiac output measurement in cardiac patients during a rehabilitation program: comparison between tonometry, impedancemetry and echocardiography
    Gonzalez-Represas, Alicia
    Mourot, Laurent
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2020, 36 (03): : 447 - 455
  • [37] Identification of volume parameters monitored with a noninvasive ultrasonic cardiac output monitor for predicting fluid responsiveness in children after congenital heart disease surgery
    Cheng, Yu-wei
    Xu, Feng
    Li, Jing
    MEDICINE, 2018, 97 (39)
  • [38] AGREEMENT OF STROKE VOLUME, CARDIAC OUTPUT, AND SYSTEMIC VASCULAR RESISTANCE BETWEEN SWAN-GANZ AND ECHOCARDIOGRAPHY
    Priyana, Andria
    Soesanto, Amiliana M.
    Joesoef, Andang H.
    HEART, 2013, 99 : E199 - E199
  • [39] Ultrasound cardiac output monitor measurements in pediatric cardiac transplant patients
    Eguchi, J. H.
    Mathur, M.
    Larsen, R.
    De-Amaya, A.
    Bailey, L. L.
    Chinnock, R.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2007, 26 (02): : S136 - S136
  • [40] REPRODUCIBILITY OF CARDIAC STROKE VOLUME ESTIMATED BY DOPPLER ECHOCARDIOGRAPHY
    IHLEN, H
    ENDRESEN, K
    MYRENG, Y
    MYHRE, E
    AMERICAN JOURNAL OF CARDIOLOGY, 1987, 59 (09): : 975 - 978