Cardiac Output Measurement Using the Ultrasonic Cardiac Output Monitor: A Validation Study in Newborn Infants

被引:11
|
作者
Fraga, Maria, V [1 ]
Dysart, Kevin C. [1 ]
Rintoul, Natalie [1 ]
Chaudhary, Aasma S. [2 ]
Ratcliffe, Sarah J. [3 ]
Fedec, Anysia [4 ]
Kren, Stephanie [4 ]
Cohen, Meryl S. [4 ]
Kirpalani, Haresh [1 ]
机构
[1] Childrens Hosp Philadelphia, Dept Pediat, Div Neonatol, Philadelphia, PA 19104 USA
[2] Univ Penn, Hosp Univ Penn, Div Neonatol, Perelman Sch Med, Philadelphia, PA 19104 USA
[3] Univ Penn, Dept Biostat & Epidemiol, Perelman Sch Med, Philadelphia, PA 19104 USA
[4] Childrens Hosp Philadelphia, Dept Pediat, Div Cardiol, Philadelphia, PA 19104 USA
关键词
Ultrasonic Cardiac Output Monitor; Cardiac output; Newborn infants; VENTRICULAR OUTPUT; CHILDREN; COMPLICATIONS; REPEATABILITY;
D O I
10.1159/000501005
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objectives: We aimed to determine the accuracy and validity of the Ultrasonic Cardiac Output Monitor (USCOM) measurements of cardiac output (CO) compared to echocardiography in newborn infants, and the inter-rater agreement of USCOM measurements. Methods: In a single-center study we prospectively evaluated neonates undergoing an echocardiographic evaluation. USCOM measurements of CO were obtained at the pulmonary and aortic valve by 2 physicians blinded to the echocardiographic results. All echocardiographic measurements were performed blinded to USCOM measurements. We first enrolled an ascertainment cohort which was subsequently validated in an independent new cohort. Agreement between echocardiography and USCOM methods was assessed by Bland-Altman analysis. Intra-class correlation coefficients (ICC) assessed the agreement between the 2 operators. The ascertainment cohort correction factors were applied in a second validation cohort and agreement of the calibrated measures evaluated with repeat Bland-Altman comparisons. Results: A total of 50 infants were enrolled in the initial cohort and 15 in the validation cohort. There was a high degree of correlation between the USCOM operators (ICC = 0.975). USCOM measurements of CO were significantly higher compared to echocardiography (left ventricular output bias 95 +/- 52 mL/kg/min and right ventricular output bias 64 +/- 30 mL/kg/min). There was no difference in the subgroup of infants with and without a ductus arteriosus. After the correction was applied to the validation cohort, there was no longer a significant difference between the measures. Conclusions: CO measured by USCOM consistently overestimated the results obtained from echocardiography. USCOM is not adequate to provide absolute estimates of CO. However, it may allow longitudinal hemodynamic assessment of sick neonates.
引用
收藏
页码:260 / 268
页数:9
相关论文
共 50 条
  • [1] Cardiac output measurement in newborn infants using the ultrasonic cardiac output monitor: an assessment of agreement with conventional echocardiography, repeatability and new user experience
    Patel, Neil
    Dodsworth, Melissa
    Mills, John F.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2011, 96 (03): : F206 - F211
  • [2] Ultrasonic cardiac output monitor for cardiac output measurement: is it equal to conventional echocardiography?
    Schiffer, P. Meyer
    Daster, C.
    Adams, M.
    Mieth, R. Arlettaz
    SWISS MEDICAL WEEKLY, 2013, 143 : 21S - 21S
  • [3] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Wonshill Koh
    Kristin A. Schneider
    Huaiyu Zang
    Sarosh P. Batlivala
    Matthew P. Monteleone
    Alexis L. Benscoter
    Meghan M. Chlebowski
    Ilias D. Iliopoulos
    David S. Cooper
    Pediatric Cardiology, 2022, 43 : 1205 - 1213
  • [4] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Koh, Wonshill
    Schneider, Kristin A.
    Zang, Huaiyu
    Batlivala, Sarosh P.
    Monteleone, Matthew P.
    Benscoter, Alexis L.
    Chlebowski, Meghan M.
    Iliopoulos, Ilias D.
    Cooper, David S.
    PEDIATRIC CARDIOLOGY, 2022, 43 (06) : 1205 - 1213
  • [5] CARDIAC OUTPUT IN NEWBORN INFANTS
    EMMANOUILIDES, GC
    MOSS, AJ
    MONSETCO.M
    MARCANO, BA
    RZEZNIC, B
    BIOLOGY OF THE NEONATE, 1970, 15 (3-4): : 186 - +
  • [6] Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement
    Walter Knirsch
    Oliver Kretschmar
    Maren Tomaske
    Kathrina Stutz
    Nicole Nagdyman
    Christian Balmer
    Achim Schmitz
    Dominique Béttex
    Felix Berger
    Urs Bauersfeld
    Markus Weiss
    Intensive Care Medicine, 2008, 34 : 1060 - 1064
  • [7] Cardiac output measurement in children:: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement
    Knirsch, Walter
    Kretschmar, Oliver
    Tomaske, Maren
    Stutz, Kathrina
    Nagdyman, Nicole
    Balmer, Christian
    Schmitz, Achim
    Bettex, Dominique
    Berger, Felix
    Bauersfeld, Urs
    Weiss, Markus
    INTENSIVE CARE MEDICINE, 2008, 34 (06) : 1060 - 1064
  • [8] Hemodynamic evaluation in preterm infants using ultrasonic cardiac output monitor (USCOM)
    Doni, Daniela
    Faraguna, Martha Caterina
    Zannin, Emanuela
    Rinaldi, Alessandro
    Cafolla, Claudia
    Iozzi, Lucia
    Cavalleri, Valeria
    Rigotti, Camilla
    Sinelli, Mariateresa
    Fedeli, Tiziana
    Ventura, Maria Luisa
    EUROPEAN JOURNAL OF PEDIATRICS, 2024, 183 (05) : 2183 - 2192
  • [9] Hemodynamic evaluation in preterm infants using ultrasonic cardiac output monitor (USCOM)
    Daniela Doni
    Martha Caterina Faraguna
    Emanuela Zannin
    Alessandro Rinaldi
    Claudia Cafolla
    Lucia Iozzi
    Valeria Cavalleri
    Camilla Rigotti
    Mariateresa Sinelli
    Tiziana Fedeli
    Maria Luisa Ventura
    European Journal of Pediatrics, 2024, 183 : 2183 - 2192
  • [10] Validation of a New Cardiac Output Monitor
    Vannucci, Andrea
    Krejci, Vladimir
    Kangrga, Ivan
    LIVER TRANSPLANTATION, 2009, 15 (11) : 1650 - 1650