Effect of False Lumen Occlusion Treatment With AFX VELA™, Candy-Plug Technique for Chronic Aortic Dissection

被引:0
|
作者
Yasuhara, Kiyomitsu [1 ]
Obayashi, Tamiyuki [1 ]
Ohki, Satoshi [1 ]
Okonogi, Shuichi [1 ]
Nagasawa, Ayako [1 ]
Yamaguchi, Ryo [1 ]
Kato, Yusuke [1 ]
Miki, Takao [1 ]
Abe, Tomonobu [2 ]
机构
[1] Isesaki City Hosp, Dept Cardiovasc Surg, 12-1 Tsunatori Hon Machi, Isesaki 3720817, Japan
[2] Gunma Univ, Dept Gen Surg Sci, Div Cardiovasc Surg, Grad Sch Med, Maebashi, Japan
关键词
chronic aortic dissection; thoracic endovascular repair; candy-plug technique; false-lumen occlusion; AFX VELA; ENDOVASCULAR REPAIR; OUTCOMES; DEVICE;
D O I
10.1177/15385744241229594
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction:<bold> </bold>We sought to examine midterm results and remodeling effect of false-lumen occlusion treatment using AFX VELA in case of chronic dissection repair. Material and methods:<bold> </bold>From June 2019 to May 2022, we performed false lumen occlusion treatment using a modified Candy-Plug technique with AFX VELA on 8 chronic aortic dissection patients with a patent false lumen. We collected operative data, short-term clinical outcomes, mid-term clinical outcomes and imaging test results. We conducted follow-up examinations at postoperative, 6-month and 1-, 2- and 3-year intervals, including contrast-enhanced computed tomography to evaluate the diameter, false lumen thrombosis and any events. Results:<bold> </bold>The average time from the symptom onset to the thoracic endovascular repair was 81.5 (35-155) months. The aorta showed aneurysmal dilation with an average maximum short-axis diameter of 58.9 (41-91) mm. Two cases needed emergency surgery due to rupture and impending rupture. There were no postoperative deaths. Complete thrombosis within the false lumen was achieved in 6 cases (75%), but 2 cases had incomplete thrombosis, requiring additional treatment. The mean maximum diameter showed a significant decrease at 6 months, 1 year and 2 years postoperatively compared to preoperative measurements (P < .05). Conclusion:<bold> </bold>We showed the results of false lumen occlusion treatment using the AFX VELA cuff. We observed favorable clinical outcomes and remodeling effects. While the long-term durability and efficacy of this technique in aortic remodeling will need to be monitored with further observation, the use of this cuff is considered a reliable approach to false lumen occlusion treatment.
引用
收藏
页码:505 / 511
页数:7
相关论文
共 50 条
  • [21] Modified 'candy-plug' technique for chronic type B aortic dissection with aneurysmal dilatation: a case report
    Kotani, Sohsyu
    Inoue, Yoshito
    Kasai, Mio
    Suzuki, Satoru
    Hachiya, Takashi
    JOURNAL OF CARDIOTHORACIC SURGERY, 2017, 12
  • [22] Outcomes of the Candy-Plug Technique to Achieve Thrombosis of Thoracic False-Lumen Postdissection Aneurysms
    Tsilimparis, Nikolaos
    Rohlffs, Fiona
    Heidemann, Franziska
    Debus, E. Sebastian
    Koelbel, Tilo
    JOURNAL OF VASCULAR SURGERY, 2016, 63 (06) : 43S - 44S
  • [23] Modified ‘candy-plug’ technique for chronic type B aortic dissection with aneurysmal dilatation: a case report
    Sohsyu Kotani
    Yoshito Inoue
    Mio Kasai
    Satoru Suzuki
    Takashi Hachiya
    Journal of Cardiothoracic Surgery, 12
  • [24] Design, evolution, and experience with the candy plug device for endovascular false lumen occlusion of chronic aortic dissections
    Grandi, Alessandro
    D'Oria, Mario
    Panuccio, Giuseppe
    Rohlffs, Fiona
    Eleshra, Ahmed
    Torrealba, Jose
    Nana, Petroula
    Lepidi, Sandro
    Melloni, Andrea
    Bertoglio, Luca
    Kolbel, Tilo
    EXPERT REVIEW OF MEDICAL DEVICES, 2024, 21 (06) : 463 - 466
  • [25] Extended Use of the Candy-Plug Technique with the Cook Dissection Stent
    Tehrani, Nader
    Soult, Michael
    Bechara, Carlos F.
    ANNALS OF VASCULAR SURGERY, 2021, 72 : 662.e1 - 662.e5
  • [26] Staged Fenestrated/Branched Repair of Postdissecting Thoracoabdominal Aneurysm With Candy-Plug False Lumen Occlusion for Spinal Cord Preconditioning
    Carta, Niccoloi
    Salvati, Simone
    Melissano, Germano
    Chiesa, Roberto
    Bertoglio, Luca
    JOURNAL OF ENDOVASCULAR THERAPY, 2020, 27 (02) : 221 - 227
  • [27] False Lumen Occlusion for Chronic Dissection
    Roselli, Eric E.
    JOURNAL OF ENDOVASCULAR THERAPY, 2013, 20 (04) : 490 - 491
  • [28] Clinical Utility of the Candy-Plug Technique Using an Excluder Aortic Extender
    Ogawa, Yukihisa
    Nishimaki, Hiroshi
    Chiba, Kiyoshi
    Iraha, Tomotaka
    Maruhashi, Takaaki
    Sakurai, Yuka
    Miyairi, Takeshi
    Nakajima, Yasuo
    ANNALS OF VASCULAR DISEASES, 2021, 14 (02) : 139 - 145
  • [29] Open and endovascular reinterventions after Candy Plug false lumen embolization of type B aortic dissection
    Loschi, Diletta
    Santoro, Annarita
    Rinaldi, Enrico
    Anselmi, Claudia
    Grignani, Camilla
    Chiesa, Roberto
    Melissano, Germano
    JOURNAL OF VASCULAR SURGERY, 2023, 78 (06) : 1409 - 1417
  • [30] Homemade Candy Plug Using a Zenith Alpha Thoracic Stent-Graft for False Lumen Distal Occlusion in Acute-on-Chronic Type B Aortic Dissection
    Yap, Kok Hooi
    Chuan Tham
    Tay, Kiang Hiong
    Tec, Chong Tze
    Irani, Farah Gillan
    Kian, Ch'ng Jack
    Chao, Victor T. T.
    JOURNAL OF ENDOVASCULAR THERAPY, 2019, 26 (05) : 732 - 735