Design, evolution, and experience with the candy plug device for endovascular false lumen occlusion of chronic aortic dissections

被引:0
|
作者
Grandi, Alessandro [1 ,2 ]
D'Oria, Mario [3 ,4 ]
Panuccio, Giuseppe [1 ]
Rohlffs, Fiona [1 ]
Eleshra, Ahmed [1 ]
Torrealba, Jose [1 ]
Nana, Petroula [1 ]
Lepidi, Sandro [3 ]
Melloni, Andrea [2 ]
Bertoglio, Luca [2 ]
Kolbel, Tilo [1 ]
机构
[1] Univ Med Ctr Eppendorf, German Aort Ctr Hamburg, Dept Vasc Med, Hamburg, Germany
[2] Univ Brescia, Dept Expt & Clin Sci, Div Vasc Surg, Brescia, Italy
[3] Univ Trieste, Dept Clin Surg & Hlth Sci, Div Vasc & Endovasc Surg, Trieste, Italy
[4] Univ Hosp Trieste ASUGI, Cardiovasc Dept, Div Vasc & Endovascular Surg, Trieste, Italy
关键词
Candy plug; thoracic endovascular aortic repair; aortic dissection; false lumen; entry tear; distal reperfusion; REPAIR;
D O I
10.1080/17434440.2024.2360987
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: The management of the false lumen (FL) when dealing with aortic dissection is a crucial aspect since inducing its thrombosis is necessary in order to achieve aortic remodeling. One of the pitfalls of endovascular treatment of aortic dissection (AD) is retrograde distal FL perfusion and pressurization, which prevents FL thrombosis and thus aortic remodeling, while being associated with aneurysmal degeneration of the FL and poor long-term outcomes. Areas Covered: Currently, there is no CE/FDA approved device for FL closure, however different techniques and devices have been proposed to overcome this challenge, the most known of which is the Candy Plug (CP). This review aims to describe the CP device, its implantation technique, and the available data in the literature (PubMed, Cochrane, and EMBASE databases; last queried, December 31, 2023). Expert Opinions: While the treatment of AD remains technically challenging, the use of the CP technique to close any distal FL reperfusion proved to be feasible and safe with excellent rates of both technical and clinical success. Furthermore, recent studies have shown a quick learning curve with this technique.
引用
收藏
页码:463 / 466
页数:4
相关论文
共 50 条
  • [1] Modified candy-plug device for aneurysmal false lumen occlusion in chronic type B aortic dissection
    Wu, I-Hui
    Chan, Chih-Yang
    Luo, Chien-Ming
    Wang, Shoei-Shen
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2018, 155 (05): : 1970 - +
  • [2] False Lumen Patency After Occlusion With Physician-Modified Candy-Plug in Aortic Dissections: A 2-Center Experience
    Karaolanis, Georgios I.
    Kotelis, Drosos
    Jungi, Silvan
    D'Amico, Rosalinda
    Mujagic, Edin
    Makaloski, Vladimir
    JOURNAL OF ENDOVASCULAR THERAPY, 2024,
  • [3] Custom Made Candy Plug for Distal False Lumen Occlusion in Aortic Dissection: International Experience
    Eleshra, Ahmed
    Haulon, Stephan
    Bertoglio, Luca
    Lindsay, Thomas
    Rohlffs, Fiona
    Dias, Nuno
    Tsilimparis, Nikolaos
    Panuccio, Giuseppe
    Koelbel, Tilo
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2023, 66 (01) : 50 - 56
  • [4] False Lumen Occlusion in Chronic Aortic Dissection: The New Generation Candy-Plug II
    Rohlffs, Fiona
    Tsilimparis, Nikolaos
    Mogensen, John
    Makaloski, Vladimir
    Debus, Sebastian
    Koelbel, Tilo
    ANNALS OF VASCULAR SURGERY, 2019, 57 : 261 - 265
  • [5] The Candy-Plug Technique: Technical Aspects and Early Results of a New Endovascular Method for False Lumen Occlusion in Chronic Aortic Dissection
    Rohlffs, Fiona
    Tsilimparis, Nikolaos
    Fiorucci, Beatrice
    Heidemann, Franziska
    Debus, Eike Sebastian
    Koelbel, Tilo
    JOURNAL OF ENDOVASCULAR THERAPY, 2017, 24 (04) : 549 - 555
  • [6] Candy Plug Generation II for False Lumen Occlusion in Chronic Aortic Dissection: Feasibility and Early Results
    Eleshra, Ahmed
    Kolbel, Tilo
    Tsilimparis, Nikolaos
    Panuccio, Giuseppe
    Scheerbaum, Martin
    Debus, E. Sebastian
    Rohlffs, Fiona
    JOURNAL OF VASCULAR SURGERY, 2019, 69 (06) : E80 - E81
  • [7] Self-occluding Candy-Plug: Implantation Technique to Obtain False Lumen Thrombosis in Chronic Aortic Dissections
    Bertoglio, Luca
    Bilman, Victor
    Rohlffs, Fiona
    Panuccio, Giuseppe
    Chiesa, Roberto
    Koelbel, Tilo
    JOURNAL OF ENDOVASCULAR THERAPY, 2023, 30 (06) : 811 - 816
  • [8] Modified Candy-plug Device for Dilated False Lumen Occlusion: Is a Candle Better Than a Candy?
    Isomura, Shogo
    Azuma, Takashi
    Yokoi, Yoshihiko
    Domoto, Satoru
    Niinami, Hiroshi
    ANNALS OF THORACIC SURGERY, 2022, 114 (03): : E219 - E221
  • [9] Effect of False Lumen Occlusion Treatment With AFX VELA™, Candy-Plug Technique for Chronic Aortic Dissection
    Yasuhara, Kiyomitsu
    Obayashi, Tamiyuki
    Ohki, Satoshi
    Okonogi, Shuichi
    Nagasawa, Ayako
    Yamaguchi, Ryo
    Kato, Yusuke
    Miki, Takao
    Abe, Tomonobu
    VASCULAR AND ENDOVASCULAR SURGERY, 2024, 58 (05) : 505 - 511
  • [10] Candy-Plug Generation II for False Lumen Occlusion in Chronic Aortic Dissection: Feasibility and Early Results
    Eleshra, Ahmed
    Koelbel, Tilo
    Tsilimparis, Nikolaos
    Panuccio, Giuseppe
    Scheerbaum, Martin
    Debus, E. Sebastian
    Mogensen, John
    Rohlffs, Fiona
    JOURNAL OF ENDOVASCULAR THERAPY, 2019, 26 (06) : 782 - 786