Accuracy of total knee arthroplasty using hospital-based 3D printed patient-specific instrumentation: A radiology research

被引:0
|
作者
Nang, Vo Sy Quyen [1 ,2 ,3 ]
Hieu, Pham Trung [3 ]
Nam, Vu Tu [3 ]
Dung, Tran Trung [2 ,3 ]
Toan, Duong Dinh [1 ]
机构
[1] Hanoi Med Univ, Hanoi, Vietnam
[2] Vinmec Healthcare Syst, Ctr Orthopaed & Sport Med, Hanoi, Vietnam
[3] VinUniversity, 3D Technol Med Ctr, Hanoi, Vietnam
关键词
3D printing; Total knee arthroplasty; TKA; In-hospital 3D printing; Patient-specific instrumentation; PSI; ROTATIONAL ALIGNMENT; TIBIAL COMPONENT; PAIN;
D O I
10.1016/j.jor.2024.10.032
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Patient-specific instruments (PSIs) were reported to improve implant position and rotation accuracy in total knee arthroplasty (TKA), among other benefits. Most PSIs used in TKA were commercial products from implant manufacturers, which could be time-consuming and could potentially affect accuracy as they relied on engineer's decision. This study aimed to present the radiological outcomes of PSI produced by the 3D printing facility at our hospital, with direct involvement of the surgeon in its design. Methods: This prospective study enrolled 45 patients who underwent mechanically aligned TKA. Preoperative whole-limb 3D-CT scans were obtained to create the sizing plan. The surgeon identified the position of the cut plane and guide pin locations on software, which were then forwarded to engineers for further processing. Subsequently, the PSI was fabricated using in-hospital 3D resin printers. During surgery, after removing all remaining cartilage, the PSI was utilized on the bone surfaces to assist surgeons in drilling pinhole positions for the manufacturer's cutting instruments. Post-surgery, CT scans were acquired and overlaid onto the 3D model of the implant to compare to the plan. Results: On average, it took three days to prepare for the surgery using PSI. 100 % of patients had implant sizes that matched the preconception template. The mean postoperative hip-knee angle was 178.1 degrees f 1.34 degrees. The mean femoral component rotation angle relative to the surgical transepicondylar axis (sTEA) and the mean rotation mismatch between the femur and tibial component were 0.99 degrees f 0.84 degrees and 0.91 degrees f 1.61 degrees, respectively. Conclusions: Preliminary radiological results demonstrated that in-hospital PSI was a useful technique for improving the accuracy of implant positioning in TKA, especially in controlling the rotational alignment.
引用
收藏
页码:70 / 74
页数:5
相关论文
共 50 条
  • [31] Patient-Specific Instrumentation Does Not Improve Alignment in Total Knee Arthroplasty
    Russell, Robert
    Brown, Timothy
    Huo, Michael
    Jones, Richard
    JOURNAL OF KNEE SURGERY, 2014, 27 (06) : 501 - 504
  • [32] Outcomes following total knee arthroplasty with CT-based patient-specific instrumentation
    Meng Zhu
    Jerry Yongqiang Chen
    Hwei Chi Chong
    Andy Khye Soon Yew
    Leon Siang Shen Foo
    Shi-Lu Chia
    Ngai Nung Lo
    Seng Jin Yeo
    Knee Surgery, Sports Traumatology, Arthroscopy, 2017, 25 : 2567 - 2572
  • [33] Effectiveness and Accuracy of a Patient-Specific Instrumentation System for Total Hip Arthroplasty
    Zhang, Teng
    Jia, Zhao
    Han, Wei
    Wang, Junqiang
    Li, Jinqi
    Gong, Maoqi
    Jiang, Xieyuan
    ORTHOPAEDIC SURGERY, 2023, 15 (03) : 878 - 887
  • [34] Outcomes following total knee arthroplasty with CT-based patient-specific instrumentation
    Zhu, Meng
    Chen, Jerry Yongqiang
    Chong, Hwei Chi
    Yew, Andy Khye Soon
    Foo, Leon Siang Shen
    Chia, Shi-Lu
    Lo, Ngai Nung
    Yeo, Seng Jin
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2017, 25 (08) : 2567 - 2572
  • [35] Accuracy of patient-specific instrumentation in anatomic and reverse total shoulder arthroplasty
    Dallalana, Richard James
    McMahon, Ryan A.
    East, Ben
    Geraghty, Liam
    INTERNATIONAL JOURNAL OF SHOULDER SURGERY, 2016, 10 (02): : 59 - 66
  • [36] Accuracy of patient-specific instrumentation in total ankle arthroplasty: A comparative study
    Saito, Guilherme H.
    Sanders, Austin E.
    O'Malley, Martin J.
    Deland, Jonathan T.
    Ellis, Scott J.
    Demetracopoulos, Constantine A.
    FOOT AND ANKLE SURGERY, 2019, 25 (03) : 383 - 389
  • [37] Contact kinematics of patient-specific instrumentation versus conventional instrumentation for total knee arthroplasty
    Broberg, Jordan S.
    Naudie, Douglas D. R.
    Howard, James L.
    Vasarhelyi, Edward M.
    McCalden, Richard W.
    Teeter, Matthew G.
    KNEE, 2020, 27 (05): : 1501 - 1509
  • [38] Accuracy of a Novel 3D-Printed Patient-Specific Intramedullary Guide to Control Femoral Component Rotation in Total Knee Arthroplasty
    Sun, Mao-lin
    Zhang, Ying
    Peng, Yang
    Fu, De-jie
    Fan, Hua-quan
    He, Rui
    ORTHOPAEDIC SURGERY, 2020, 12 (02) : 429 - 441
  • [39] 3D Printed Patient-Specific Acetabular Jig for Cup Placement in Total Hip Arthroplasty
    Abhishek Mishra
    Tarun Verma
    Gaurang Rajkumar
    Amit Agarwal
    Lalit Sharma
    Indian Journal of Orthopaedics, 2020, 54 : 174 - 180
  • [40] 3D Printed Patient-Specific Acetabular Jig for Cup Placement in Total Hip Arthroplasty
    Mishra, Abhishek
    Verma, Tarun
    Rajkumar
    Agarwal, Gaurang
    Sharma, Amit
    Maini, Lalit
    INDIAN JOURNAL OF ORTHOPAEDICS, 2020, 54 (02) : 174 - 180