Validity of Wrist-Worn Activity Tracker Heart Rate Detection in Fontan Patients during Exercise

被引:0
|
作者
Pierick, Alyson R. [1 ]
Burke, Kelly J. [1 ]
Prusi, Megan [1 ]
Largent, Bethany [1 ]
Yu, Sunkyung [1 ]
Lowery, Ray E. [1 ]
Duimstra, Ashley [1 ]
Hansen, Jesse E. [1 ]
机构
[1] Univ Michigan, Div Pediat Cardiol, 1540 East Hosp Dr, Ann Arbor, MI 48109 USA
关键词
PHYSICAL ACTIVITY; SINGLE VENTRICLE; AMBULATORY TRACKING; EXERCISE PHYSIOLOGY; QUALITY-OF-LIFE; PHYSICAL-ACTIVITY; CIRCULATION; ACCURACY; MONITORS; CHILDREN; IMPROVE; ADULTS;
D O I
10.1249/MSS.0000000000003567
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Purpose Physical activity and a healthy lifestyle play an essential role in optimizing long-term health in patients with Fontan physiology. Wrist-worn activity trackers may be useful in medically directed exercise programs for patients with Fontan physiology. The objective of this study was to measure the validity of Garmin and Fitbit activity tracker heart rate detection in patients with Fontan circulation when compared with electrocardiogram during cardiopulmonary exercise testing (CPET). Methods Forty-seven Fontan patients undergoing CPET for clinical indications were included and wore activity trackers during CPET. Heart rate via the activity tracker was collected at baseline, maximal exercise, and recovery. Patient heart rates, peak VO2, and peak respiratory exchange ratio were collected using standard CPET protocols and equipment. Heart rate at each time point was compared between the activity trackers and CPET electrocardiogram. Results Median age of participants was 17.1 yr, 15.1 yr since Fontan completion. Mean percentage of predicted peak VO2 was 56.8%, with a z-score of -3.2, and 61.7% of participants completing a maximal CPET. Baseline mean oxygen saturation was 92.9% and 90.0% at maximal exercise. Activity trackers demonstrated mean absolute percentage error <10% at most time points, comparable with other studies. Demographics, Fontan-associated comorbidities, and echocardiogram findings did not impact the accuracy. Conclusions Consumer-oriented wrist-worn activity trackers show promising accuracy for heart rate monitoring in medically directed exercise programs for adolescents and young adults with Fontan physiology. Further validation across different exercise modalities is needed.
引用
收藏
页码:280 / 284
页数:5
相关论文
共 50 条
  • [31] Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
    Benjamin W. Nelson
    Carissa A. Low
    Nicholas Jacobson
    Patricia Areán
    John Torous
    Nicholas B. Allen
    npj Digital Medicine, 3
  • [32] Predicting Resistance Training Exercise Repetitions Using A Wrist-worn Activity Monitor
    Conger, Scott A.
    Guo, Jun
    Mercier, Kenzie
    Needham, Cameron D.
    Zamzow, Clare
    Mecham, Christopher
    Chen, Hao
    Bassett, David R., Jr.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2018, 50 (05): : 298 - 298
  • [33] Accuracy of Wrist-Worn Activity Monitors during Treadmill and Elliptical Ergometry
    Smith, John D.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05): : 365 - 365
  • [34] Appropriate heart rate during exercise in Fontan patients
    Hedlund, Eva R.
    Soderstrom, Liselott
    Lundell, Bo
    CARDIOLOGY IN THE YOUNG, 2020, 30 (05) : 674 - 680
  • [35] The validity and accuracy of wrist-worn activity monitors in lower-limb prosthesis users
    Smith, John D.
    Guerra, Gary
    Burkholder, Brian G.
    DISABILITY AND REHABILITATION, 2020, 42 (22) : 3182 - 3188
  • [36] A wrist-worn oximetry in the detection of obstructive sleep apnea in stroke patients
    Wang, Xiaoyu
    Zhang, Chunrong
    Li, Weiwei
    Lei, Yanfeng
    An, Yuan
    NEUROLOGY ASIA, 2024, 29 (04) : 1113 - 1121
  • [37] Photoplethysmography-based estimation of heart rate recovery using a wrist-worn device
    Sokas, Daivaras
    Petrenas, Andrius
    Daukantas, Saulius
    Rapalis, Andrius
    Marozas, Vaidotas
    2018 16TH BIENNIAL BALTIC ELECTRONICS CONFERENCE (BEC), 2018,
  • [38] Forecasting of heart rate variability using wrist-worn heart rate monitor based on Hidden Markov Model
    Yun, Sanghun
    Son, Chang-Sik
    Lee, Sang-Ho
    Kang, Won-Seok
    2018 INTERNATIONAL CONFERENCE ON ELECTRONICS, INFORMATION, AND COMMUNICATION (ICEIC), 2018, : 408 - 409
  • [39] Accuracy of Wrist-Worn Photoplethysmography Devices at Measuring Heart Rate in the Laboratory and During Free-Living Activities
    Giggins, Oonagh M.
    Doyle, Julie
    Sojan, Nisanth
    Moran, Orla
    Crabtree, Daniel R.
    Fraser, Matthew
    Muggeridge, David J.
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 6970 - 6973
  • [40] Radar-Based Passive Step Counter and Its Comparison with a Wrist-Worn Physical Activity Tracker
    Muaaz, Muhammad
    Waqar, Sahil
    Patzold, Matthias
    INTELLIGENT TECHNOLOGIES AND APPLICATIONS, 2022, 1616 : 259 - 272