Does Root Canal Shaping Effect the Accuracy of Electronic Apex Locators in Curved and Straight Root Canals?

被引:0
|
作者
Tekinarslan, Dide [1 ]
Erkal, Damla [1 ]
Ercan, Esen [1 ]
Koc, Simay [1 ]
Er, Kursat [1 ]
机构
[1] Akdeniz Univ, Fac Dent, Dept Endodont, Antalya, Turkiye
来源
关键词
Working length; root canal preparation; curvature; straight canal; WORKING LENGTH; ABILITY; INSTRUMENTATION; FREQUENCY;
D O I
10.33808/clinexphealthsci.1354100
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective: To assess the influence of root canal preparation (RCP) on the precision of electronic apex locators (EALs) in both straight and curved root canals. Methods: The experimental setup included forty-two extracted human mandibular molars featuring both straight distal canals (0-10 degrees) and curved mesial root canals (30-50 degrees). Initial working length (AWL1) measurements were conducted under a stereomicroscope at X15 magnification prior to root canal preparation (RCP). Subsequently, electronic working length (EWL1) was determined using two different electronic apex locators (EALs) - Propex-Pixi and EndoRadar. RCP was carried out with ProTaper Ultimate F1, F2 files in mesial canals, and F1, F2, F3 files in distal straight root canals. The RCP procedure involved the implementation of a standard irrigation protocol along with the use of a sonic device for activation. Following RCP, measurements were repeated under a stereomicroscope and documented as AWL2. EWL measurements were also repeated and recorded as EWL2 after the preparation. The variation was calculated by subtracting AWL from EWL for each tooth (EWL1-AWL1 and EWL2-AWL2). Statistical analysis employed Paired Sample t-test and Wilcoxon tests. Results: There was no statistically significant difference observed in the accuracy of Propex-Pixi and EndoRadar Pro before and after root canal preparation for both straight and curved canals. Conclusion: Root canal preparation did not alter the accuracy of the electronic apex locators (EALs) used in this study, whether applied to curved or straight root canals.
引用
收藏
页码:727 / 732
页数:6
相关论文
共 50 条
  • [41] An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals
    Ebrahim, A. K.
    Wadachi, R.
    Suda, H.
    AUSTRALIAN DENTAL JOURNAL, 2007, 52 (03) : 193 - 197
  • [42] Accuracy of an Electronic Apex Locator (EAL) in Measuring the Working Length of Root Canals in Comparison With Radiographs: An In Vivo Study
    Bhagat, Esha
    Agnihotri, Yoshaskam
    Das, Abhisek
    Das, Sambarta
    Panda, Swagat
    Hota, Sudeepta
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (05)
  • [43] Influence of operator expertise on glide path and root canal preparation of curved root canals with rotary and reciprocating motions
    Belen Dablanca-Blanco, Ana
    Arias, Ana
    Jose Ginzo-Villamayor, Maria
    Consuelo Perez, Maria
    Castelo-Baz, Pablo
    Martin-Biedma, Benjamin
    AUSTRALIAN ENDODONTIC JOURNAL, 2022, 48 (01) : 37 - 43
  • [44] AN INVITRO METHOD FOR COMPARING THE EFFECTS OF DIFFERENT ROOT-CANAL PREPARATION TECHNIQUES ON THE SHAPE OF CURVED ROOT CANALS
    TANG, MPF
    STOCK, CJR
    INTERNATIONAL ENDODONTIC JOURNAL, 1989, 22 (02) : 49 - 54
  • [45] The effect of progressive root canal enlargement on root surface strain during obturation of oval root canals
    Saber, Shehab El-Din Mohamed
    Nassif, Mohammad Salah
    ENDO-ENDODONTIC PRACTICE TODAY, 2011, 5 (04): : 275 - 280
  • [46] Morphological study of the Er,Cr:YSGG laser for root canal preparation in mandibular incisors with curved root canals
    Matsuoka, E
    Jayawardena, JA
    Matsumoto, K
    PHOTOMEDICINE AND LASER SURGERY, 2005, 23 (05) : 480 - 484
  • [47] Effect of Prefllaring on Accuracy of Endodontic Length Determination by Four Electronic Apex Locators
    Izadi, Arash
    Bayandari, Maliheh
    Kazeminejad, Ezatolah
    JOURNAL OF DENTISTRY INDONESIA, 2022, 29 (03) : 165 - 171
  • [48] Accuracy of an electronic apex locator in primary teeth with root resorption
    Angwaravong, O.
    Panitvisai, P.
    INTERNATIONAL ENDODONTIC JOURNAL, 2009, 42 (02) : 115 - 121
  • [49] Effect of chloroform, orange solvent and eucalyptol on the accuracy of four electronic apex locators
    Al-Hadlaq, Solaiman M.
    AUSTRALIAN ENDODONTIC JOURNAL, 2013, 39 (03) : 112 - 115
  • [50] EFFECTS OF THE FLEXOGATES AND CANAL MASTER-U ON ROOT-CANAL CONFIGURATION IN SIMULATED CURVED CANALS
    CAMPS, J
    MACOUIN, G
    PERTOT, WJ
    INTERNATIONAL ENDODONTIC JOURNAL, 1994, 27 (01) : 21 - 25