Accuracy and safety of percutaneous pedicle screw placement using the K-wireless minimally invasive spine percutaneous pedicle screw system in Japan: A randomized active controlled study

被引:1
|
作者
Ohmori, Kazuo [1 ]
Terayama, Sei [2 ]
Ono, Koichiro
Sakamoto, Miyuki [3 ]
Horikoshi, Yukie
机构
[1] Nippon Koukan Hosp, Ctr Spinal Surg, 1-2-1 Koukandori,Kawasaki Ku, Kawasaki, Kanagawa, Japan
[2] Shunyokai Cent Hosp, Dept Orthoped Surg, 485 Niitomi, Kimotuki, Kagoshima, Japan
[3] Johnson & Johnson KK, Med Co, 3-5-2 Nishikanda,Chiyoda Ku, Tokyo, Japan
来源
关键词
Randomized controlled trial; K-wireless MIS PPS system; Accuracy of screw insertion; Percutaneous Pedicle Screw; Minimally Invasive Spine; PPS; FIXATION; FUSION;
D O I
10.1016/j.xnsj.2022.100121
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Minimally invasive lumbar fusion has recently become a widely used technique worldwide. This randomized active controlled study was conducted to demonstrate the non-inferiority of the K-wireless Minimally Invasive Spine (MIS) Percutaneous Pedicle Screw (PPS) system compared with use of the six pedicle screw systems currently used in our practices with respect to the accuracy of pedicle screw placement.Also to compare the screw- insertion time and number of fluoroscopic observations during screw insertion between the groups. Methods: A total of 80 patients with degenerative spinal diseases or vertebral fractures were assigned, including 41 patients in the K-wireless MIS PPS system group (K-wireless group) and 39 in the control group (K-wire group). The accuracy of the screw insertion, screw-insertion time, number of fluoroscopic observations during screw insertion, and the incidence of adverse events were compared between the K-wireless group and the K-wire group. The accuracy rate was calculated as the number of screws with no breach divided by the total number of screws. Results: The accuracy rates of screw insertion were 85.7% and 75.0% in the K-wireless and K-wire groups, respectively, with an intergroup difference of 10.7% (95% confidence interval: 2.3-19.1%). The K-wireless group demonstrated non-inferiority compared with the K-wire group. The mean screw-insertion time was significantly shorter in the K-wireless group (2.62 and 2.97 min in the K-wireless and K-wire groups, respectively; P = 0.005). There were also significantly fewer fluoroscopies in the K-wireless group (10.7 and 17.4 in the K-wireless and K-wire groups, respectively; P < 0.001). There were no device-related or study treatment-related adverse events in either group. Conclusions: The accuracy of pedicle screw insertion using the K-wireless MIS PPS system was not inferior to that of existing products. In terms of safety, no product-related or treatment-related adverse events were identified in this study and no new safety concerns were noted.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Feasibility of laser-guided percutaneous pedicle screw placement in the lumbar spine using a hybrid-OR
    P. H. Richter
    F. Gebhard
    M. Salameh
    K. Schuetze
    M. Kraus
    International Journal of Computer Assisted Radiology and Surgery, 2017, 12 : 873 - 879
  • [32] Feasibility of laser-guided percutaneous pedicle screw placement in the lumbar spine using a hybrid-OR
    Richter, P. H.
    Gebhard, F.
    Salameh, M.
    Schuetze, K.
    Kraus, M.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2017, 12 (05) : 873 - 879
  • [33] Accuracy of pedicle screw insertion in the thoracic and lumbar spine: a comparative study between percutaneous screw insertion and conventional open technique
    Hiroko Ikeuchi
    Ko Ikuta
    Archives of Orthopaedic and Trauma Surgery, 2016, 136 : 1195 - 1202
  • [34] Accuracy of pedicle screw insertion in the thoracic and lumbar spine: a comparative study between percutaneous screw insertion and conventional open technique
    Ikeuchi, Hiroko
    Ikuta, Ko
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2016, 136 (09) : 1195 - 1202
  • [35] Percutaneous Pedicle Screw Placement in the Lumbar Spine A Comparison Study Between the Novel Guidance System and the Conventional Fluoroscopy Method
    Gu, Guangfei
    Zhang, Hailong
    He, Shisheng
    Cai, Xiaobing
    Gu, Xin
    Jia, Jianbo
    Fu, Qingsong
    Zhou, Xu
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2015, 28 (09): : E522 - E527
  • [36] Tracked Ultrasound Snapshots in Percutaneous Pedicle Screw Placement Navigation: A Feasibility Study
    Ungi, Tamas
    Moult, Eric
    Schwab, Joseph H.
    Fichtinger, Gabor
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (12) : 4047 - 4055
  • [37] Percutaneous Pedicle Screw Placement Aided by a New Drill Guide Template Combined with Fluoroscopy: An Accuracy Study
    Wu, Chao
    Deng, Jiayan
    Li, Tao
    Tan, Lun
    Yuan, Dechao
    ORTHOPAEDIC SURGERY, 2020, 12 (02) : 471 - 479
  • [38] Accuracy of pedicle and mass screw placement in the spine without using fluoroscopy: a prospective clinical study
    Kotil, Kadir
    Bilge, Turgay
    SPINE JOURNAL, 2008, 8 (04): : 591 - 596
  • [39] Long-Term Prospective Analysis of Minimally Invasive Transforaminal Lumbar Interbody Fusion and Percutaneous Pedicle Screw Placement
    Cruet, Mick Perez
    Hiremath, Girish
    de la Torre, Ramiro Perez
    Cooley, David
    Perez, Sophia Lang
    JOURNAL OF NEUROSURGERY, 2011, 115 (02) : A446 - A446
  • [40] Assessing Procedural Accuracy in Lateral Spine Surgery: A Retrospective Analysis of Percutaneous Pedicle Screw Placement with Intraoperative CT Navigation
    Hiyama, Akihiko
    Sakai, Daisuke
    Katoh, Hiroyuki
    Nomura, Satoshi
    Watanabe, Masahiko
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (21)