A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy

被引:0
|
作者
Alzahrani, Meshal [1 ,2 ,3 ]
Broadbent, David A. [2 ,3 ]
Teh, Irvin [3 ]
Al-Qaisieh, Bashar [2 ]
Johnstone, Emily [4 ]
Speight, Richard [2 ]
机构
[1] King Abdulaziz Univ, Fac Appl Med Sci, Dept Radiol Sci, Jeddah, Saudi Arabia
[2] Leeds Teaching Hosp NHS Trust, Dept Med Phys & Engn, Leeds, England
[3] Univ Leeds, Leeds Inst Cardiovasc & Metab Med, Leeds, England
[4] Christie NHS Fdn Trust, Christie Med Phys & Engn, Manchester, England
关键词
Magnetic resonance imaging; Radiotherapy; MRI in RT; Quality assurance; ACR phantom; Multimodal anthropomorphic phantom; IMPLEMENTATION; SIMULATION;
D O I
10.1016/j.phro.2025.100707
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The use of magnetic resonance imaging (MRI) for radiotherapy (RT) simulation has grown, prompting quality assurance (QA) guidelines by the Institute of Physics and Engineering in Medicine (IPEM) and the American Association of Physicists in Medicine (AAPM). This study compares a novel multimodality anthropomorphic phantom to an American College of Radiology (ACR) phantom for a subset of these MRI-specific QA tests in RT. Materials and methods: A novel 3D-printed multimodality head-and-neck anthropomorphic phantom was compared to an ACR large MRI phantom. IPEM and AAPM-recommended QA tests were conducted, including informatics/connectivity/data transfer, MRI-CT registration, end-to-end QA, and signal-to-noise ratio (SNR)/ percentage integral uniformity (PIU) assessments using RT accessories. Results: Both phantoms were suitable for informatics/connectivity/data transfer. In MRI-CT registration, no errors were found; the ACR phantom offered more quantitative landmarks, while the anthropomorphic phantom provided limited structures. Both phantoms achieved target registration errors (TREs) below 0.97 mm and dice similarity coefficient (DSC) values above 0.9, meeting guidelines. For end-to-end QA, the anthropomorphic phantom facilitated dose measurements of 1.994 Gy versus a calculated 2.01 Gy (-0.8 %). SNR and PIU assessments showed higher values in radiology setups compared to RT setups for both phantoms. Conclusions: Multimodality anthropomorphic phantoms compatible with dosimetric equipment allow realistic end-to-end QA, unlike the ACR phantom. While the ACR phantom is suitable for informatics and MRI-CT registration, anthropomorphic phantoms better represent clinical scenarios. For comprehensive QA, both ACR and anthropomorphic phantoms are required. Additionally, large field-of-view (FOV) phantoms are crucial for evaluating large FOV MRI distortions.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] SEMIAUTOMATED QUALITY ASSURANCE FOR QUANTITATIVE MAGNETIC-RESONANCE-IMAGING
    BARKER, GJ
    TOFTS, PS
    MAGNETIC RESONANCE IMAGING, 1992, 10 (04) : 585 - 595
  • [22] A Novel Integrated Quality Assurance Phantom for Radiographic and Nonradiographic Radiotherapy Localization and Positioning Systems
    Fowler, T.
    Yu, Amy S.
    MEDICAL PHYSICS, 2017, 44 (06) : 3060 - 3060
  • [23] Design of a fused phantom for quantitative evaluation of brain metabolites and enhanced quality assurance testing for magnetic resonance imaging and spectroscopy
    Song, Kyu-Ho
    Kim, Sang-Young
    Lee, Do-Wan
    Jung, Jin-Young
    Lee, Jung-Hoon
    Baek, Hyeon-Man
    Choe, Bo-Young
    JOURNAL OF NEUROSCIENCE METHODS, 2015, 255 : 75 - 84
  • [24] Remote Cherenkov imaging-based quality assurance of a magnetic resonance image-guided radiotherapy system
    Andreozzi, Jacqueline M.
    Mooney, Karen E.
    Bruza, Petr
    Curcuru, Austen
    Gladstone, David J.
    Pogue, Brian W.
    Green, Olga
    MEDICAL PHYSICS, 2018, 45 (06) : 2647 - 2659
  • [25] A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
    Yu, Amy S.
    Fowler, Tyler L.
    Dubrowski, Piotr
    MEDICAL PHYSICS, 2018, 45 (07) : 2857 - 2863
  • [26] Quality assurance of human functional magnetic resonance imaging: a literature review
    Lu, Weizhao
    Doug, Kejiang
    Cui, Dong
    Jiao, Qing
    Qiu, Jianfeng
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (06) : 1147 - 1162
  • [27] The Osteoarthritis Initiative (OAI) magnetic resonance imaging quality assurance update
    Schneider, E.
    NessAiver, M.
    OSTEOARTHRITIS AND CARTILAGE, 2013, 21 (01) : 110 - 116
  • [28] A Novel Polymer Gel-Loading Phantom for Magnetic Resonance Imaging
    Cheng, Kai-Yuan
    Chen, Sih-Ying
    Shieh, Jiunn-I
    Wu, Yi-Ying
    Hsieh, Ling-Ling
    ARABIAN JOURNAL FOR SCIENCE AND ENGINEERING, 2016, 41 (01) : 163 - 167
  • [29] A Novel Polymer Gel-Loading Phantom for Magnetic Resonance Imaging
    Kai-Yuan Cheng
    Sih-Ying Chen
    Jiunn-I Shieh
    Yi-Ying Wu
    Ling-Ling Hsieh
    Arabian Journal for Science and Engineering, 2016, 41 : 163 - 167
  • [30] QUANTITATIVE DCE-CT IMAGING QUALITY ASSURANCE WITH A NOVEL DYNAMIC FLOW PHANTOM
    Driscoll, B.
    Keller, H.
    Coolens, C.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S178 - S179