Dynamic B0 field shimming for improving pseudo-continuous arterial spin labeling at 7 T

被引:0
|
作者
Ji, Yang [1 ,2 ]
Woods, Joseph G. [1 ]
Li, Hongwei [1 ,3 ]
Okell, Thomas W. [1 ]
机构
[1] Univ Oxford, Wellcome Ctr Integrat Neuroimaging, Nuffield Dept Clin Neurosci, FMRIB Div, Oxford, England
[2] Univ Sci & Technol China, Sch Informat Sci & Technol, Dept Elect Engn & Informat Sci, Hefei, Peoples R China
[3] Fudan Univ, Inst Sci & Technol Brain Inspired Intelligence, Shanghai, Peoples R China
基金
英国惠康基金; 中国国家自然科学基金;
关键词
7; T; B0; inhomogeneity; dynamic B0 field shimming; pseudo-continuous arterial spin labeling (PCASL); OFF-RESONANCE; PERFUSION; QUANTIFICATION; INVERSION; ROBUST;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeB0 field inhomogeneity within the brain-feeding arteries is a major issue for pseudo-continuous arterial spin labeling (PCASL) at 7 T because it reduces the labeling efficiency and leads to a loss of perfusion signal. This study aimed to develop a vessel-specific dynamic B0 field shimming method for 7 T PCASL to improve the labeling efficiency by correcting off-resonance within the arteries in the labeling region.MethodsWe implemented a PCASL sequence with dynamic B0 shimming at 7 T that compensates for B0 field offsets in the brain-feeding arteries by updating linear shimming terms and adding a phase increment to the PCASL RF pulses. Rapidly acquired vessel-specific B0 field maps were used to calculate dynamic B0 shimming parameters. We evaluated both 2D and 3D variants of our method, comparing their performance against the established global frequency offset and optimal encoding scheme-based corrections. Cerebral blood flow (CBF) maps were quantified before and after corrections, and CBF values from different methods were compared across the whole brain, white matter, and gray matter regions.ResultsAll off-resonance correction methods significantly recovered perfusion signals across the brain. The proposed vessel-specific dynamic B0 shimming method improved the labeling efficiency while maintaining optimal static shimming in the imaging region. Perfusion-weighted images demonstrated the superiority of the 3D dynamic B0 shimming method compared to global or 2D-based correction approaches. CBF analysis revealed that 3D dynamic B0 shimming significantly increased CBF values relative to the other methods.ConclusionOur proposed dynamic B0 shimming method offers a significant advancement in PCASL robustness and effectiveness, enabling full utilization of 7 T ASL high sensitivity and spatial resolution.
引用
收藏
页数:16
相关论文
共 50 条
  • [31] Reliability and Reproducibility of Hadamard Encoded Pseudo-Continuous Arterial Spin Labeling in Healthy Elderly
    Neumann, Katja
    Schidlowski, Martin
    Guenther, Matthias
    Stoecker, Tony
    Duezel, Emrah
    FRONTIERS IN NEUROSCIENCE, 2021, 15
  • [32] Reduction of BOLD interference in pseudo-continuous arterial spin labeling: towards quantitative fMRI
    Warnock, Geoffrey
    Oezbay, Pinar S.
    Kuhn, Felix P.
    Nanz, Daniel
    Buck, Alfred
    Boss, Andreas
    Rossi, Cristina
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2018, 38 (05): : 847 - 856
  • [33] Inter-Vendor Reproducibility of Pseudo-Continuous Arterial Spin Labeling at 3 Tesla
    Mutsaerts, Henri J. M. M.
    Steketee, Rebecca M. E.
    Heijtel, Dennis F. R.
    Kuijer, Joost P. A.
    van Osch, Matthias J. P.
    Majoie, Charles B. L. M.
    Smits, Marion
    Nederveen, Aart J.
    PLOS ONE, 2014, 9 (08):
  • [34] Cerebral Perfusion in Adolescent Major Depression: A Pseudo-Continuous Arterial Spin Labeling Study
    Ho, Tiffany C.
    Wu, Jing
    Shin, David D.
    Yang, Guang
    Banerjee, Dipavo
    Connolly, Colm G.
    Hendren, Robert
    Simmons, Alan G.
    Yang, Tony
    BIOLOGICAL PSYCHIATRY, 2013, 73 (09) : 165S - 165S
  • [35] Cerebral Blood Flow Quantification in Swine Using Pseudo-Continuous Arterial Spin Labeling
    Johnston, Megan E.
    Zheng, Zhenlin
    Maldjian, Joseph A.
    Whitlow, Christopher T.
    Morykwas, Michael J.
    Jung, Youngkyoo
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (05) : 1111 - 1118
  • [36] Simultaneous measurement of brain perfusion and labeling efficiency in a single pseudo-continuous arterial spin labeling scan
    Chen, Zhensen
    Zhao, Xihai
    Zhang, Xingxing
    Guo, Rui
    Teeuwisse, Wouter M.
    Zhang, Bida
    Koken, Peter
    Smink, Jouke
    Yuan, Chun
    van Osch, Matthias J. P.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (04) : 1922 - 1930
  • [37] Perfusion measurement in brain gliomas using velocity-selective arterial spin labeling: comparison with pseudo-continuous arterial spin labeling and dynamic susceptibility contrast MRI
    Qu, Yaoming
    Kong, Dexia
    Wen, Haitao
    Ou, Xiaochan
    Rui, Qihong
    Wang, Xianlong
    Lin, Doris D.
    Qin, Qin
    Wen, Zhibo
    EUROPEAN RADIOLOGY, 2022, 32 (05) : 2976 - 2987
  • [38] Perfusion measurement in brain gliomas using velocity-selective arterial spin labeling: comparison with pseudo-continuous arterial spin labeling and dynamic susceptibility contrast MRI
    Yaoming Qu
    Dexia Kong
    Haitao Wen
    Xiaochan Ou
    Qihong Rui
    Xianlong Wang
    Doris D. Lin
    Qin Qin
    Zhibo Wen
    European Radiology, 2022, 32 : 2976 - 2987
  • [39] Pseudo-continuous arterial spin labeling at very high magnetic field (11.75 T) for high-resolution mouse brain perfusion imaging
    Duhamel, Guillaume
    Callot, Virginie
    Tachrount, Mohamed
    Alsop, David C.
    Cozzone, Patrick J.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) : 1225 - 1236
  • [40] Neural plasticity secondary to carpal tunnel syndrome: a pseudo-continuous arterial spin labeling study
    Deng, Xue
    Chau, Phoebe Lai-Heung
    Chiu, Suk-Yee
    Leung, Kwok-Pui
    Hu, Yong
    Ip, Wing-Yuk
    NEURAL REGENERATION RESEARCH, 2021, 16 (01) : 158 - 165