Evaluation of dynamic accuracy and latency of a surface-guided radiotherapy system

被引:0
|
作者
Yamauchi, Ryohei [1 ]
Tomita, Fumihiro [1 ]
机构
[1] St Lukes Int Hosp, Dept Radiat Oncol, 9-1 Akashi Cho,Chuo Ku, Tokyo 1048560, Japan
关键词
Surface-guided radiotherapy; Respiratory motion; Dynamic accuracy; Latency; RADIATION-THERAPY; QUALITY-ASSURANCE;
D O I
10.1007/s12194-024-00866-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to evaluate the dynamic accuracy and latency of the surface-guided radiotherapy (SGRT) system using TrueBeam and AlignRT in compliance with SGRT guidelines. Beam characteristics-flatness, symmetry, beam quality, and output-were compared between gated and nongated beams using a two-dimensional ionization chamber array and a Farmer-type chamber. Dynamic accuracy was assessed using a moving platform and breast phantom, with measurements taken for various shift values (5, 10, 30 mm), region-of-interest (ROI) shapes, reference-surface image types (DICOM and capture), surface resolutions, and room illuminations. Latency due to differences in frame rates was evaluated using radiochromic film, calculated from position displacements of profiles at two speeds. Differences in beam characteristics between gated and nongated beams were within 0.1%. Dynamic accuracy showed minimal dependence on settings, with deviations of < 1 mm for a 10-mm shift. A maximum displacement of 1.9 mm was observed with a 30-mm shift at the body ROI. Beam-on latency at 12, 16, 25, and 35 frames per second was 253.2 +/- 21.9, 225.7 +/- 33.7, 177.1 +/- 43.0, and 112.4 +/- 29.2 ms, respectively, with similar trends for beam-off latency. This study is the first to evaluate the dynamic accuracy of the TrueBeam and AlignRT system under SGRT-specific settings. While accuracy was generally maintained (< 1 mm), ROI shape significantly impacted results. Latency results indicate that certain frame rates may not meet guideline limits, underscoring the need for careful SGRT system use in clinical applications.
引用
收藏
页码:136 / 146
页数:11
相关论文
共 50 条
  • [21] Surface-guided radiotherapy for breast cancer in deep inspiration breath hold
    Beer, K. T.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S539 - S541
  • [22] Comparison of patient setup accuracy for optical surface-guided and X-ray-guided imaging with respect to the impact on intracranial stereotactic radiotherapy
    Schoepe, Michael
    Sahlmann, Jacob
    Jaschik, Stefan
    Findeisen, Anne
    Klautke, Gunther
    STRAHLENTHERAPIE UND ONKOLOGIE, 2024, 200 (01) : 60 - 70
  • [23] Comparison of patient setup accuracy for optical surface-guided and X-ray-guided imaging with respect to the impact on intracranial stereotactic radiotherapy
    Michael Schöpe
    Jacob Sahlmann
    Stefan Jaschik
    Anne Findeisen
    Gunther Klautke
    Strahlentherapie und Onkologie, 2024, 200 : 60 - 70
  • [24] Failure modes and effects analysis for surface-guided DIBH breast radiotherapy
    Bright, Megan
    Foster, Ryan D.
    Hampton, Carnell J.
    Ruiz, Justin
    Moeller, Benjamin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (04):
  • [25] Reply to comparison of patient setup accuracy for optical surface-guided and X-ray-guided imaging with respect to the impact on intracranial stereotactic radiotherapy
    Schoepe, M.
    Sahlmann, J.
    Jaschik, S.
    Findeisen, A.
    Klautke, G.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2024, 200 (07) : 644 - 645
  • [26] Permanent marker free surface-guided breast radiotherapy: Implementing a new technique
    Goldsmith, Christy
    Dobson, Deirdre
    Littler, Adam
    Sawyer, Elinor
    Moriggi, Georgina
    Chauhan, Roshni
    CANCER RESEARCH, 2021, 81 (04)
  • [27] Design and virtual evaluation of a customized surface-guided knee implant
    Pejhan, Shabnam
    Bohm, Eric
    Brandt, Jan-Mels
    Wyss, Urs
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (10) : 949 - 961
  • [28] Surface-Guided Radiotherapy Systems As Source of Big Data: Using Sgrt Data to Study the Radiotherapy Process
    Jacqmin, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E731 - E732
  • [29] Patient positioning accuracy with surface guided radiotherapy
    Mychko, D.
    Dziameshka, P.
    Zverava, A.
    Siamkousky, S.
    Shautsou, N.
    Chyrets, M.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S2022 - S2022
  • [30] Breathing Amplitude Variations in Left Breast Radiotherapy using Surface-Guided DIBH Technique
    Aman, Abid Ali Mirza
    Al-Ajmi, Amna Abdullah
    Pervez, Nadeem
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S5454 - S5454