Effect of crystalline lens decentration and tilt on visual performance in eyes implanted with bifocal or extended depth of focus intraocular lenses

被引:0
|
作者
Zhang, Yu [1 ,2 ]
Zhang, Jiaqing [1 ]
Liang, Chen [1 ]
Xu, Yifan [1 ,2 ]
Wu, Jiaying [2 ]
Wu, Junming [2 ]
Zhang, Yifan [1 ,2 ]
Jin, Aixia [1 ,2 ]
Tan, Xuhua [1 ]
Luo, Lixia [1 ]
机构
[1] Sun Yat Sen Univ, Zhongshan Ophthalm Ctr, State Key Lab Ophthalmol, Guangdong Prov Key Lab Ophthalmol & Visual Sci, Guangzhou, Peoples R China
[2] Sun Yat Sen Univ Med Sci, Guangzhou, Peoples R China
基金
中国国家自然科学基金;
关键词
Cataract; Intraocular lens; IOL; Presbyopia-correcting IOL; Decentration; Tilt; RETINAL IMAGE QUALITY; OPTICAL-QUALITY; CONTRAST SENSITIVITY; ABERRATION; PRECISION; VISION;
D O I
10.1186/s12886-025-03884-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeTo explore the potential of crystalline lens decentration and tilt as indicators for screening cataract patients for presbyopia-correcting intraocular lens (IOL) implantation. MethodsProspective observational study. Patients undergoing phacoemulsification with bifocal (Tecnis ZMB00) or extended depth-of-focus (EDOF) (Tecnis ZXR00) IOL implantation were consecutively enrolled. The decentration and tilt of the crystalline lens and the IOL were quantified through the utilization of swept-source optical coherence tomography (SS-OCT, Casia2). Postoperative visual acuity (VA), contrast sensitivity (CS), objective optical quality, and patient-reported outcomes were assessed at a 3-month follow-up. A LOWESS (LOcally WEighted Scatterplot Smoothing) curve was employed to analyze the changes in VA, CS, and objective optical quality relative to the decentration or tilt of crystalline lenses. A further comparison of visual outcomes was conducted based on the inflection points suggested by the curves. ResultsEighty-seven patients with ZMB00 IOL and 76 patients with ZXR00 IOL were included. Multiple 6-mm internal aberrations showed a nonlinear increase with greater crystalline lens decentration. The inflection points for the steep increase were observed to be 0.28 mm for the bifocal group and 0.35 mm for the EDOF group. Beyond these points, internal aberrations such as coma increased significantly (all P < 0.01). Patient satisfaction decreased (P < 0.01). Moreover, the bifocal group exhibited a decline in spectacle-independence from 98.67% to 83.33% (P = 0.049), along with a reduction in CS (P < 0.05). ConclusionsThe increased decentration of crystalline lenses compromises specific visual quality aspects in eyes implanted with ZMB00 and ZXR00 IOLs, with cut-off values of 0.28 mm and 0.35 mm, respectively.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Postoperative Visual Function of Extended Depth-of-Focus Intraocular Lenses Versus Monofocal Lenses
    Komatsu, Koji
    Masuda, Yoichiro
    Tachibana, Sei
    Sano, Kei
    Iida, Masanobu
    Ichihara, Kosuke
    Oki, Tetutaro
    Fukai, Kota
    Tatemichi, Masayuki
    Nakano, Tadashi
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (07) : E499 - +
  • [32] Effect of decentration and tilt on four novel extended range of vision intraocular lenses regarding far distance
    Schmid, Ruediger
    Luedtke, Holger
    Borkenstein, Andreas F.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2023, 33 (02) : 933 - 942
  • [33] Visual Performance of Non-Diffractive Extended Depth of Focus and Neutral Aspheric Monofocal Intraocular Lenses
    O'Rourke, Tara
    Pantanelli, Seth
    Bolognia, Olga
    Longenecker, Amy
    Lehman, Erik
    Scruggs, Kathleen
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [34] Visual function in eyes with diffractive extended depth-of-focus and monofocal intraocular lenses: 2-year comparison
    Hisai, Takahiro
    Takada, Keita
    Tokuda, Shota
    Nejima, Ryohei
    Mori, Yosai
    Minami, Keiichiro
    Miyata, Kazunori
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 261 (09) : 2567 - 2573
  • [35] Visual function in eyes with diffractive extended depth-of-focus and monofocal intraocular lenses: 2-year comparison
    Takahiro Hisai
    Keita Takada
    Shota Tokuda
    Ryohei Nejima
    Yosai Mori
    Keiichiro Minami
    Kazunori Miyata
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2023, 261 : 2567 - 2573
  • [36] Effect of blocking violet light in extended depth of focus intraocular lenses
    Faria-Ribeiro, Miguel
    Jenkins, Mark
    Rosen, Robert
    van der Mooren, Marrie
    Canovas, Carmen
    Piers, Patricia
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [37] The Effect of a Spectral Filter on Visual Quality in Patients with an Extended-Depth-Of-Focus Intraocular Lens
    Labuz, Grzegorz
    Auffarth, Gerd U.
    Ozen, Aydin
    Van den Berg, Thomas J. T. P.
    Yildirim, Timur M.
    Son, Hyeck-Soo
    Khoramnia, Ramin
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2019, 208 : 56 - 63
  • [38] Visual Performance, Subjective Satisfaction and Quality of Life Effect of a New Refractive Intraocular Lens with Central Extended Depth of Focus
    Gillmann, Kevin
    Mermoud, Andre
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2019, 236 (04) : 384 - 390
  • [39] Effect of Violet Light-Filtering and Manufacturing Improvements in an Extended Depth-of-Focus Intraocular Lens on Visual Performance
    van der Mooren, Marrie
    Alarcon, Aixa
    Sanchez, Mark D. Jenkins
    Chang, Daniel H.
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 701 - 709
  • [40] Difference in Quality of Vision Outcome among Extended Depth of Focus, Bifocal, and Monofocal Intraocular Lens Implantation
    Chao, Chen-Cheng
    Lin, Hung-Yuan
    Lee, Chia-Yi
    Mai, Elsa Lin-Chin
    Lian, Ie-Bin
    Chang, Chao-Kai
    HEALTHCARE, 2022, 10 (06)