Electrical Bioimpedance-Based Monitoring of Intracochlear Tissue Changes After Cochlear Implantation

被引:0
|
作者
Sijgers, Leanne [1 ]
Geys, Marlies [1 ]
Geissler, Gunnar [2 ]
Boyle, Patrick [2 ]
Huber, Alexander [1 ]
Pfiffner, Flurin [1 ]
机构
[1] Univ Zurich, Univ Hosp Zurich, Dept Otorhinolaryngol Head & Neck Surg, CH-8091 Zurich, Switzerland
[2] Adv Bion GmbH, European Res Ctr, D-30625 Hannover, Germany
关键词
digital biomarkers; remote hearing healthcare; cochlear implant; neuroprosthetics; hearing technology; auditory rehabilitation; HEARING PRESERVATION; IMPEDANCE CHANGES; AUDITORY-NERVE; IN-VITRO; STIMULATION; ELECTRODES; MODEL; ELECTROCOCHLEOGRAPHY; OUTCOMES; HEALTH;
D O I
10.3390/s24237570
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background: This study examined electrical bioimpedance as a biomarker for intracochlear tissue changes after cochlear implant surgery, comparing monopolar, three-point, and four-point impedance measurements over time and evaluating different measurement systems and approaches. Methods: Impedance measurements were obtained from 21 participants during surgery and at four postoperative stages. Monopolar impedances were recorded using the Bionic Ear Data Collection System (BEDCS) and the Active Insertion Monitoring (AIM) system. Three- and four-point impedances were recorded directly using BEDCS, and indirect three-point impedances were additionally derived from Electrical Field Imaging matrices recorded using BEDCS or AIM. Results: There was an 11% relative error between monopolar measurements from BEDCS and AIM and a 25% discrepancy between direct and indirect three-point measurements. Despite this, direct and indirect measurements from both systems were useful for tracking postoperative impedance shifts. Three- and four-point measurements showed a strong relationship both during and after surgery. Our results suggest that three- and four-point measurements are more specific than monopolar impedances in capturing localized tissue changes. Conclusions: Three- and four-point impedance measurements are potential markers of intracochlear tissue changes over time. While direct three-point impedance measurements offer higher accuracy, indirect measurements provide a feasible alternative for monitoring intracochlear changes in clinical settings lacking the option of direct measurements.
引用
收藏
页数:18
相关论文
共 50 条
  • [21] Changes in residual hearing after cochlear implantation
    Leung, EKS
    Wong, TKC
    Tong, MCF
    van Hasselt, CA
    UPDATES IN COCHLEAR IMPLANTATION, 2000, 57 : 397 - 400
  • [22] Changes in vowel quality after cochlear implantation
    Schenk, BS
    Baumgartner, WD
    Hamzavi, JS
    ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 2003, 65 (03): : 184 - 188
  • [23] CHANGES IN THE CAT COCHLEAR NUCLEUS FOLLOWING NEONATAL DEAFENING AND CHRONIC INTRACOCHLEAR ELECTRICAL-STIMULATION
    LUSTIG, LR
    LEAKE, PA
    SNYDER, RL
    REBSCHER, SJ
    HEARING RESEARCH, 1994, 74 (1-2) : 29 - 37
  • [24] The Effects of Gusher-Related Intracochlear Pressure Changes on Hearing Preservation in Cochlear Implantation: A Comparative Series
    Crohan, William
    Krishnaswamy, Jay
    Rajan, Gunesh
    AUDIOLOGY AND NEURO-OTOLOGY, 2018, 23 (03) : 181 - 186
  • [25] Ultra-High-Resolution CT to Detect Intracochlear New Bone Formation after Cochlear Implantation
    Heutink, Floris
    Klabbers, Tim M.
    Huinck, Wendy J.
    Lucev, Federica
    van der Woude, Willem Jan
    Mylanus, Emmanuel A. M.
    Verbist, Berit M.
    RADIOLOGY, 2022, 302 (03) : 605 - 612
  • [26] Electrode impedance changes after implantation of a dexamethasone-eluting intracochlear array
    Needham, Karina
    Stathopoulos, Dimitra
    Newbold, Carrie
    Leavens, Jason
    Risi, Frank
    Manouchehri, Shahram
    Durmo, Irfan
    Cowan, Robert
    COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (02) : 98 - 109
  • [27] Real-Time Bioimpedance-Based Biopsy Needle Can Identify Tissue Type with High Spatial Accuracy
    Halonen, Sanna
    Kari, Juho
    Ahonen, Petri
    Kronstrom, Kai
    Hyttinen, Jari
    ANNALS OF BIOMEDICAL ENGINEERING, 2019, 47 (03) : 836 - 851
  • [28] Air caloric computer system application in monitoring vestibular function changes after cochlear implantation
    Szirmai, A
    Ribári, O
    Répássy, G
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2001, 125 (06) : 631 - 634
  • [29] Intraoperative Cochlear Nerve Monitoring in Cochlear Implantation after Vestibular Schwannoma Resection
    Fiasca, Valerio Maria Di Pasquale
    Tealdo, Giulia
    AUDIOLOGY RESEARCH, 2023, 13 (03) : 398 - 407
  • [30] Modelling the Recovery of Residual Acoustic Hearing after Cochlear Implantation by Using Feasible Intracochlear Acoustic Devices
    Slater, David
    Elliott, Stephen
    Verschuur, Carl
    NONLINEARITY AND HEARING: ADVANCES IN THEORY AND EXPERIMENT, 2024, 3062