Electrical Bioimpedance-Based Monitoring of Intracochlear Tissue Changes After Cochlear Implantation

被引:0
|
作者
Sijgers, Leanne [1 ]
Geys, Marlies [1 ]
Geissler, Gunnar [2 ]
Boyle, Patrick [2 ]
Huber, Alexander [1 ]
Pfiffner, Flurin [1 ]
机构
[1] Univ Zurich, Univ Hosp Zurich, Dept Otorhinolaryngol Head & Neck Surg, CH-8091 Zurich, Switzerland
[2] Adv Bion GmbH, European Res Ctr, D-30625 Hannover, Germany
关键词
digital biomarkers; remote hearing healthcare; cochlear implant; neuroprosthetics; hearing technology; auditory rehabilitation; HEARING PRESERVATION; IMPEDANCE CHANGES; AUDITORY-NERVE; IN-VITRO; STIMULATION; ELECTRODES; MODEL; ELECTROCOCHLEOGRAPHY; OUTCOMES; HEALTH;
D O I
10.3390/s24237570
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background: This study examined electrical bioimpedance as a biomarker for intracochlear tissue changes after cochlear implant surgery, comparing monopolar, three-point, and four-point impedance measurements over time and evaluating different measurement systems and approaches. Methods: Impedance measurements were obtained from 21 participants during surgery and at four postoperative stages. Monopolar impedances were recorded using the Bionic Ear Data Collection System (BEDCS) and the Active Insertion Monitoring (AIM) system. Three- and four-point impedances were recorded directly using BEDCS, and indirect three-point impedances were additionally derived from Electrical Field Imaging matrices recorded using BEDCS or AIM. Results: There was an 11% relative error between monopolar measurements from BEDCS and AIM and a 25% discrepancy between direct and indirect three-point measurements. Despite this, direct and indirect measurements from both systems were useful for tracking postoperative impedance shifts. Three- and four-point measurements showed a strong relationship both during and after surgery. Our results suggest that three- and four-point measurements are more specific than monopolar impedances in capturing localized tissue changes. Conclusions: Three- and four-point impedance measurements are potential markers of intracochlear tissue changes over time. While direct three-point impedance measurements offer higher accuracy, indirect measurements provide a feasible alternative for monitoring intracochlear changes in clinical settings lacking the option of direct measurements.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] Bioimpedance-Based Respiration Monitoring With a Defibrillator
    Martinsen, Orjan G.
    Nordbotten, Bernt
    Grimnes, Sverre
    Fossan, Helge
    Eilevstjonn, Joar
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (06) : 1858 - 1862
  • [2] Pneumolabyrinth, intracochlear and vestibular fluid loss after cochlear implantation
    Moteki, Hideaki
    Fujinaga, Yasunari
    Goto, Tetsuya
    Usami, Shin-ichi
    AURIS NASUS LARYNX, 2018, 45 (05) : 1116 - 1120
  • [3] Intracochlear Auditory Nerve Monitoring and Cochlear Implantation After Total Resection of a Vestibular Schwannoma: Case Report
    Mertens, Griet
    Lammers, Marc Jan-Willem
    Vanderveken, Olivier
    Van Rompaey, Vincent
    B-ENT, 2024, 20 (01) : 61 - 68
  • [4] Flexible bioimpedance-based dynamic monitoring of stress levels in live oysters
    Zhang, Luwei
    Li, You
    Du, Jiacheng
    Mu, Boyu
    Hu, Jinyou
    Zhang, Xiaoshuan
    AQUACULTURE, 2023, 577
  • [5] Application of extra- and intracochlear electrocochleography during and after cochlear implantation
    Dalbert, Adrian
    Weder, Stefan
    HNO, 2025, 73 (01) : 14 - 21
  • [6] Quality Control after Intracochlear Intralabyrinthine Schwannoma Resection and Cochlear Implantation
    Sudhoff, Holger
    Scholtz, Lars Uwe
    Gehl, Hans Bjoern
    Todt, Ingo
    BRAIN SCIENCES, 2021, 11 (09)
  • [7] Application of extra- and intracochlear electrocochleography during and after cochlear implantation
    Dalbert, Adrian
    Weder, Stefan
    HNO, 2025, 73 (01) : 14 - 21
  • [8] A Bioimpedance-based Transducer for Insulin Bioavailability Assessment after Subcutaneous Administration
    Arpaia, Pasquale
    Cuneo, Davide
    Mancino, Francesca
    Moccaldi, Nicola
    2022 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC 2022), 2022,
  • [9] Monitoring Cochlear Health With Intracochlear Electrocochleography During Cochlear Implantation: Findings From an International Clinical Investigation
    O'Leary, S.
    Mylanus, E.
    Venail, F.
    Lenarz, T.
    Birman, C.
    Di Lella, F.
    Roland Jr, J. T.
    Gantz, B.
    Beynon, A.
    Sicard, M.
    Buechner, A.
    Lai, W. K.
    Boccio, C.
    Choudhury, B.
    Tejani, V. D.
    Plant, K.
    English, R.
    Arts, R.
    Bester, C.
    EAR AND HEARING, 2023, 44 (02): : 358 - 370
  • [10] Correlation between word recognition score and intracochlear new bone and fibrous tissue after cochlear implantation in the human
    Kamakura, Takefumi
    Nadol, Joseph B., Jr.
    HEARING RESEARCH, 2016, 339 : 132 - 141