Signal Acquisition and Algorithm Design for Bioimpedance-Based Heart Rate Estimation from the Wrist

被引:0
|
作者
Lapsa, Didzis [1 ,2 ]
Metshein, Margus [3 ]
Krivosei, Andrei [3 ]
Janeliukstis, Rims [1 ]
Martens, Olev [3 ]
Elsts, Atis [1 ]
机构
[1] Inst Elect & Comp Sci, Dzerbenes St 14, LV-1006 Riga, Latvia
[2] Univ Latvia, Fac Comp, Raina Bulvaris 19, LV-1586 Riga, Latvia
[3] Tallinn Univ Technol, Thomas Johann Seebeck Dept Elect, Ehitajate Tee 5, EE-19086 Tallinn, Estonia
来源
APPLIED SCIENCES-BASEL | 2024年 / 14卷 / 21期
关键词
bipolar electrode system; electrical bioimpedance; heart rate estimation; peak detection; signal-to-noise ratio; tetrapolar electrode system;
D O I
10.3390/app14219632
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Background: Heart rate (HR) is a critical biomarker that provides insights into overall health, stress levels, and the autonomic nervous system. Pulse wave signals contain valuable information about the cardiovascular system and heart status. However, signal acquisition in wearables poses challenges, particularly when using electrical sensors, due to factors like the distance from the heart, body movement, and suboptimal electrode placement. Methods: Electrical bioimpedance (EBI) measurements using bipolar and tetrapolar electrode systems were employed for pulse wave signal acquisition from the wrist in both perpendicular and distal configurations. Signal preprocessing techniques, including baseline removal via Hankel matrix methods, normalization, cross-correlation, and peak detection, were applied to improve signal quality. This study describes the combination of sensor-level signal acquisition and processing for accurate wearable HR estimation. Results: The bipolar system was shown to produce larger Delta Z(t), while the tetrapolar system demonstrated higher sensitivity. Distal placement of the electrodes yielded greater Delta Z(t) (up to 0.231 Omega) when targeting both wrist arteries. Bandpass filtering resulted in a better signal-to-noise ratio (SNR), achieving 3.6 dB for the best bipolar setup and 4.8 dB for the tetrapolar setup, compared to 2.6 and 3.3 dB SNR, respectively, with the Savitzky-Golay filter. The custom HR estimation algorithm presented in this paper demonstrated improved accuracy over a reference method, achieving an average error of 1.8 beats per minute for the best bipolar setup, with a mean absolute percentage error (MAPE) of 8%. Conclusions: The analysis supports the feasibility of using bipolar electrode setups on the wrist and highlights the importance of electrode positioning relative to the arteries. The proposed signal processing method, featuring a preprocessing pipeline and HR estimation algorithm, provides a proof-of-concept demonstration for HR estimation from EBI signals acquired at the wrist.
引用
收藏
页数:22
相关论文
共 50 条
  • [21] Heart rate detection from plantar bioimpedance measurements
    Gonzalez-Landaeta, Rafael
    Casas, Oscar
    Pallas-Areny, Ramon
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (03) : 1163 - 1167
  • [22] Heart rate detection from plantar bioimpedance measurements
    Gonzalez Landaeta, R.
    Casas, O.
    Pallas-Areny, R.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 3817 - +
  • [23] Nightbeat: Heart Rate Estimation From a Wrist-Worn Accelerometer During Sleep
    Moebus, Max
    Hauptmann, Lars
    Kopp, Nicolas
    Demirel, Berken Utku
    Braun, Bjorn
    Holz, Christian
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2025, 29 (02) : 870 - 877
  • [24] A microcontroller based system for real-time heart rate estimation from ECG signal
    Chatterjee, H. K.
    Gupta, R.
    Mitra, M.
    2012 ANNUAL IEEE INDIA CONFERENCE (INDICON), 2012, : 1020 - 1025
  • [25] Chip rate estimation of the DS/SS signal based on SSME algorithm
    Peng Y.-H.
    Tang B.
    Lü M.
    Dianzi Yu Xinxi Xuebao/Journal of Electronics and Information Technology, 2010, 32 (07): : 1649 - 1653
  • [26] Robust Heart Rate Monitoring by a Single Wrist-Worn Accelerometer Based on Signal Decomposition
    Zhao, Chao
    Zeng, Wenru
    Hu, Dandan
    Liu, Hong
    IEEE SENSORS JOURNAL, 2021, 21 (14) : 15962 - 15971
  • [27] Design, Development and Clinical Validation of a Wrist-based Optical Heart Rate Monitor
    Preejith, S. P.
    Alex, Annamol
    Joseph, Jayaraj
    Sivaprakasam, Mohanasankar
    2016 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2016, : 493 - 498
  • [28] Learning based Quality Indicator Aiding Heart Rate Estimation in Wrist-Worn PPG
    Lutin, E.
    Biswas, D.
    Simoes-Capela, N.
    Van Hoof, C.
    Van Helleputte, N.
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 7063 - 7067
  • [29] Photoplethysmography-based estimation of heart rate recovery using a wrist-worn device
    Sokas, Daivaras
    Petrenas, Andrius
    Daukantas, Saulius
    Rapalis, Andrius
    Marozas, Vaidotas
    2018 16TH BIENNIAL BALTIC ELECTRONICS CONFERENCE (BEC), 2018,
  • [30] Design of Fast Acquisition Algorithm for TDRSS Burst Signal
    Ren, Qi-yi
    Yao, Zheng
    Lu, Ming-quan
    PROCEEDINGS OF 2015 IEEE 5TH INTERNATIONAL CONFERENCE ON ELECTRONICS INFORMATION AND EMERGENCY COMMUNICATION, 2015, : 206 - 209