Heart rate detection from plantar bioimpedance measurements

被引:39
|
作者
Gonzalez-Landaeta, Rafael [1 ]
Casas, Oscar [1 ]
Pallas-Areny, Ramon [1 ]
机构
[1] Univ Politecn Cataluna, Instrumentat Sensors & Interface Grp, Barcelona 08860, Spain
关键词
coherent demodulation; heart rate detection; plantar bioimpedance; synchronous sampling;
D O I
10.1109/TBME.2007.906516
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, a novel technique for heart rate measurement on a standing subject is proposed that relies on electrical impedance variations detected by a plantar interface with booth feet, such as those in some bathroom weighting scales for body composition analysis. Heart-related impedance variations in the legs come from arterial blood circulation and are below 500 m Omega. To detect them, we have implemented a system with a gain in excess of 600, and whose fully differential AC input amplifier has a gain of 4.5 and a common-mode rejection ratio (CMRR) higher than 90 dB at 10 kHz. Differential coherent demodulation based on synchronous sampling yields a signal-to-noise ratio (SNR) of about 54 dB. The system sensitivity is 610 mV/Omega. The technique has been demonstrated on 18 volunteers, whose bioimpedance signal and ECG were simultaneously recorded. A Bland-Altman plot shows a mean bias of -0.2 ms between the RR time intervals obtained from these two signals, which is negligible. The technique is simple and user friendly and does not require any additional sensors or electrodes attached to the body, hence no conductive gel or skin preparation.
引用
收藏
页码:1163 / 1167
页数:5
相关论文
共 50 条
  • [1] Heart rate detection from plantar bioimpedance measurements
    Gonzalez Landaeta, R.
    Casas, O.
    Pallas-Areny, R.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 3817 - +
  • [2] Heart Rate Detection from Single-Foot Plantar Bioimpedance Measurements in a Weighing Scale
    Diaz, Delia H.
    Casas, Oscar
    Pallas-Areny, Ramon
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 6489 - 6492
  • [3] Electrode Comparison for Heart Rate Detection via Bioimpedance Measurements
    Lapsa, Didzis
    Janeliukstis, Rimas
    Elsts, Atis
    2022 WORKSHOP ON MICROWAVE THEORY AND TECHNIQUES IN WIRELESS COMMUNICATIONS (MTTW), 2022, : 41 - 46
  • [4] Transthoracic Bioimpedance Measurements for Heart Failure Detection
    Reljin, Natasa
    2018 14TH SYMPOSIUM ON NEURAL NETWORKS AND APPLICATIONS (NEUREL), 2018,
  • [5] The alcohol detection using heart rate variability and bioimpedance
    Chen, Chien-Hung
    Chen, Cheng-Yu
    Huang, Min-Wei
    Cheng, Kuo-Sheng
    3RD KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2006, 2007, 15 : 598 - +
  • [6] The feasibility of transoesophageal bioimpedance measurements for the detection of heart graft rejection
    Giovinazzo, G.
    Ribas, N.
    Cinca, J.
    Rosell-Ferrer, J.
    PHYSIOLOGICAL MEASUREMENT, 2011, 32 (07)
  • [7] A Real-Time Heart Rate Detection Algorithm Based on Peripheral Electrical Bioimpedance
    Mansouri, Sofiene
    Chabchoub, Souhir
    Alharbi, Yousef
    Alshrouf, Anwar
    Nebhen, Jamel
    IEEJ TRANSACTIONS ON ELECTRICAL AND ELECTRONIC ENGINEERING, 2022, 17 (07) : 1054 - 1060
  • [8] Dynamic heart rate measurements from video sequences
    Yu, Yong-Poh
    Raveendran, P.
    Lim, Chern-Loon
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (07): : 2466 - 2480
  • [9] Signal Acquisition and Algorithm Design for Bioimpedance-Based Heart Rate Estimation from the Wrist
    Lapsa, Didzis
    Metshein, Margus
    Krivosei, Andrei
    Janeliukstis, Rims
    Martens, Olev
    Elsts, Atis
    APPLIED SCIENCES-BASEL, 2024, 14 (21):
  • [10] Combining Bioimpedance and EMG Measurements for Reliable Muscle Contraction Detection
    Kusche, Roman
    Ryschka, Martin