Comparison of intelligent 4D CT sequence scanning and conventional spiral 4D CT: A first comprehensive phantom study

被引:13
|
作者
Werner R. [1 ]
Szkitsak J. [2 ,3 ]
Sentker T. [1 ]
Madesta F. [1 ]
Schwarz A. [3 ,4 ]
Fernolendt S. [3 ,4 ]
Vornehm M. [3 ,4 ]
Gauer T. [1 ]
Bert C. [2 ,3 ]
Hofmann C. [4 ]
机构
[1] University Medical Center Hamburg-Eppendorf, Martinistr. 52, Hamburg
[2] Department of Radiation Oncology, Universitätsklinikum Erlangen, Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen
[3] Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen
[4] Siemens Healthcare GmbH, Siemensstr. 3, Forchheim
来源
Physics in Medicine and Biology | 2021年 / 66卷 / 01期
关键词
4D CT; I4DCT; Motion artifacts; Radiotherapy;
D O I
10.1088/1361-6560/abc93a
中图分类号
学科分类号
摘要
4D CT imaging is a cornerstone of 4D radiotherapy treatment. Clinical 4D CT data are, however, often affected by severe artifacts. The artifacts are mainly caused by breathing irregularity and retrospective correlation of breathing phase information and acquired projection data, which leads to insufficient projection data coverage to allow for proper reconstruction of 4D CT phase images. The recently introduced 4D CT approach i4DCT (intelligent 4D CT sequence scanning) aims to overcome this problem by breathing signal-driven tube control. The present motion phantom study describes the first in-depth evaluation of i4DCT in a real-world scenario. Twenty-eight 4D CT breathing curves of lung and liver tumor patients with pronounced breathing irregularity were selected to program the motion phantom. For every motion pattern, 4D CT imaging was performed with i4DCT and a conventional spiral 4D CT mode. For qualitative evaluation, the reconstructed 4D CT images were presented to clinical experts, who scored image quality. Further quantitative evaluation was based on established image intensity-based artifact metrics to measure (dis)similarity of neighboring image slices. In addition, beam-on and scan times of the scan modes were analyzed. The expert rating revealed a significantly higher image quality for the i4DCT data. The quantitative evaluation further supported the qualitative: While 20% of the slices of the conventional spiral 4D CT images were found to be artifact-affected, the corresponding fraction was only 4% for i4DCT. The beam-on time (surrogate of imaging dose) did not significantly differ between i4DCT and spiral 4D CT. Overall i4DCT scan times (time between first beam-on and last beam-on event, including scan breaks to compensate for breathing irregularity) were, on average, 53% longer compared to spiral CT. Thus, the results underline that i4DCT significantly improves 4D CT image quality compared to standard spiral CT scanning in the case of breathing irregularity during scanning. © 2021 Institute of Physics and Engineering in Medicine.
引用
收藏
相关论文
共 50 条
  • [41] Verification of 4D Dose Delivery Using 4D Digital Human Phantom
    Zhang, F.
    Qin, Y.
    Segars, W.
    Yin, F.
    Cai, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3859 - 3859
  • [42] Development of an anthropomorphic 4D Phantom for multimodal imaging, 4D radiation and SGRT
    Moghaddam, Anahita Bakhtiari
    Runz, Armin
    Haering, Peter
    Seeber, Steffen
    Moeller, Michael
    Echner, Gernot
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S2083 - S2084
  • [43] A comparison of amplitude- and phase-based 4D CT
    Thorndyke, B
    Schreibmann, E
    Li, T
    Boyer, A
    Xing, L
    MEDICAL PHYSICS, 2005, 32 (06) : 1919 - 1919
  • [44] Clinical introduction of respiration correlated (4D) CT/PET scanning.
    Bosmans, G.
    Nagel, C.
    Potargent, I.
    Dekker, A.
    Oellers, M.
    de Ruysscher, D.
    Minken, A.
    Lambin, P.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S212 - S213
  • [45] Accuracy of 4D Cone Beam CT and 4D Fan Beam CT for Thoracic Radiation Therapy Simulation and Verification Imaging
    Nyflot, M.
    Bowen, S. R.
    Lee, T.
    Meyer, J.
    Sandison, G. A.
    Kinahan, P. E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S640 - S640
  • [46] On-Beam 4D CT Reconstruction Using Motion Amplitude Change Provided in Projection Images and Planning 4D CT
    Jung, J.
    Park, S.
    Kim, J.
    Yeo, I.
    Yi, B.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [47] Quantitative evaluation of a new 4D MRI sequence: a motion phantom study
    Parent, L.
    Nemtanu, T.
    Celicanin, Z.
    Bieri, O.
    Cattin, P.
    Ken, S.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1210 - S1210
  • [48] Extension of the 4D NCAT phantom to dynamic x-ray CT simulation
    Segars, WP
    Tsui, BMW
    Frey, EC
    Fishman, EK
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 3195 - 3199
  • [49] A comprehensive quality assurance procedure for 4D CT commissioning and periodic QA
    Polizzi, Mitchell
    Kim, Siyong
    Rosu-Bubulac, Mihaela
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (11):
  • [50] Evaluation of Liver Deformation During Breathing Using Deformable Image Registration: A Comparison Between 4D CT and 4D MRI
    Liang, X.
    Czito, B.
    Palta, M.
    Bashir, M.
    Yin, F.
    Cai, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 171 - 171