Comparison of intelligent 4D CT sequence scanning and conventional spiral 4D CT: A first comprehensive phantom study

被引:13
|
作者
Werner R. [1 ]
Szkitsak J. [2 ,3 ]
Sentker T. [1 ]
Madesta F. [1 ]
Schwarz A. [3 ,4 ]
Fernolendt S. [3 ,4 ]
Vornehm M. [3 ,4 ]
Gauer T. [1 ]
Bert C. [2 ,3 ]
Hofmann C. [4 ]
机构
[1] University Medical Center Hamburg-Eppendorf, Martinistr. 52, Hamburg
[2] Department of Radiation Oncology, Universitätsklinikum Erlangen, Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen
[3] Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen
[4] Siemens Healthcare GmbH, Siemensstr. 3, Forchheim
来源
Physics in Medicine and Biology | 2021年 / 66卷 / 01期
关键词
4D CT; I4DCT; Motion artifacts; Radiotherapy;
D O I
10.1088/1361-6560/abc93a
中图分类号
学科分类号
摘要
4D CT imaging is a cornerstone of 4D radiotherapy treatment. Clinical 4D CT data are, however, often affected by severe artifacts. The artifacts are mainly caused by breathing irregularity and retrospective correlation of breathing phase information and acquired projection data, which leads to insufficient projection data coverage to allow for proper reconstruction of 4D CT phase images. The recently introduced 4D CT approach i4DCT (intelligent 4D CT sequence scanning) aims to overcome this problem by breathing signal-driven tube control. The present motion phantom study describes the first in-depth evaluation of i4DCT in a real-world scenario. Twenty-eight 4D CT breathing curves of lung and liver tumor patients with pronounced breathing irregularity were selected to program the motion phantom. For every motion pattern, 4D CT imaging was performed with i4DCT and a conventional spiral 4D CT mode. For qualitative evaluation, the reconstructed 4D CT images were presented to clinical experts, who scored image quality. Further quantitative evaluation was based on established image intensity-based artifact metrics to measure (dis)similarity of neighboring image slices. In addition, beam-on and scan times of the scan modes were analyzed. The expert rating revealed a significantly higher image quality for the i4DCT data. The quantitative evaluation further supported the qualitative: While 20% of the slices of the conventional spiral 4D CT images were found to be artifact-affected, the corresponding fraction was only 4% for i4DCT. The beam-on time (surrogate of imaging dose) did not significantly differ between i4DCT and spiral 4D CT. Overall i4DCT scan times (time between first beam-on and last beam-on event, including scan breaks to compensate for breathing irregularity) were, on average, 53% longer compared to spiral CT. Thus, the results underline that i4DCT significantly improves 4D CT image quality compared to standard spiral CT scanning in the case of breathing irregularity during scanning. © 2021 Institute of Physics and Engineering in Medicine.
引用
收藏
相关论文
共 50 条
  • [21] On-Treatment 4D CT Reconstruction From Planning 4D CT Using Linear Amplitude Scaling
    Park, S.
    Jung, J.
    Kim, J.
    Yeo, I.
    Yi, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3434 - 3434
  • [22] THE USE OF 4D CT SCANNING IN RADICAL RADIOTHERAPY FOR LUNG CANCER
    Huang, V.
    Liu, M.
    Khaira, M.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S59 - S59
  • [23] Technical Considerations for Automated Low-Pitch Spiral 4D CT Scanning Protocol Selection
    Werner, Rene
    Sothmann, Thilo
    Madesta, Frederic
    Gauser, Tobias
    Hofmann, Christian
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [24] A 4D CT digital phantom of an individual human brain for perfusion analysis
    Manniesing, Rashindra
    Brune, Christoph
    van Ginneken, Bram
    Prokop, Mathias
    PeerJ, 2016, 4
  • [25] Developing a 4D Movement Phantom to Examine the Effect of Movements in CT and Irradiation
    Berndt, J.
    Borm, K.
    Duma, M. N.
    Wilkens, J. J.
    Geinitz, H.
    Oechsner, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2013, 189 : 142 - 143
  • [26] Feasibility of 4D VMAT-CT
    Zhao, Xiaodong
    Zhang, Rui
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (06)
  • [27] Optimization in 4D CT/PET acquisitions
    Mari, A.
    Valenti, M.
    Maggi, S.
    Cardinali, M.
    Ascoli, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S507 - S507
  • [28] 4D Iterative Reconstruction in Cardiac CT
    Bruder, H.
    Raupach, R.
    Allmendinger, T.
    Sunnegardh, J.
    Stierstorfer, K.
    Flohr, T.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [29] Tools for Development of 4D Proton CT
    Dou, T.
    Ramos-Mendez, J.
    Piersimoni, P.
    Giacometti, V.
    Penfold, S.
    Censor, Y.
    Faddegon, B.
    Low, D.
    Schulte, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3298 - 3299
  • [30] Adaptive 4D cone beam CT
    Lu, J.
    Zhu, R.
    Munro, P.
    Jeung, A.
    Mostafave, H.
    Chi, P.
    Balter, P.
    Guerrero, T.
    Mohan, R.
    Pan, T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S609 - S609