Task-based assessment for radiation dose optimization in CT abdominal examinations: A phantom study

被引:0
|
作者
Halim, Esnu [1 ]
Kechik, Mohd Mustafa Awang [1 ]
Ibahim, Mohamad Johari [2 ]
Harun, Hanif Haspi [1 ]
Shaffiq, Said Mohd [3 ]
Yusof, Aimi Adibah [1 ]
Karim, Muhammad Khalis Abdul [1 ]
机构
[1] Univ Putra Malaysia, Fac Sci, Dept Phys, Serdang 43400, Selangor, Malaysia
[2] Univ Teknol MARA, Fac Med, Sungai Buloh, Selangor, Malaysia
[3] Natl Canc Inst, Putrajaya 62250, Malaysia
关键词
Abdominal CT; Optimization; Radiation dose; Signal-to-noise ratio; Figure-of-merit; IMAGE QUALITY; COMBINATION;
D O I
10.1016/j.radphyschem.2024.111966
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Recent advancements in Computed Tomography (CT) imaging technology have significantly enhanced healthcare delivery. However, public concern regarding the safety and health risks associated with increased exposure to ionizing radiation has escalated. In response, this study performs a quantitative analysis comparing the image quality of the standard CT abdomen protocol (P1 protocols) with alternative optimization techniques. Adjustments to standard CT scan parameters, including tube potential (P2 protocols), effective mAs value (P3), and pitch factor (P4, P5, P6 protocols), were made to evaluate the effectiveness of these optimization strategies. A 64slice CT scanner (Canon, Japan) and a polymethylmethacrylate (PMMA) textural water phantom was utilized in this study. CT dose descriptors, volume weighted CT Dose Index (CTDI vol ) and dose-length product (DLP) alongside evaluation parameters such as Hounsfield Unit (HU) number, noise levels, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured and analysed. Furthermore, this research introduced a Figure of Merit (FOM) based on the ratio of image quality parameters (SNR and CNR) to the CTDI vol , establishing a quantitative measure for optimization. A notable reduction in radiation dose -34.85% below the standard protocol (P1) -was observed in protocol P6, attributed to an increased pitch factor. Conversely, increase the effective mAs value has improved SNR and CNR values. The findings reveal that protocols P1 and P2 deliver the highest FOM for soft tissue and fat imaging, respectively. Protocol P1, with its higher attenuation factor, is optimal for abdominal examinations targeting soft tissues. Meanwhile, protocol P2 's increase effective mAs value enhances beam intensity, improving CT numbers for fat tissues with their high-density composition. The study also found a strong correlation between FOM based on SNR and CNR values. In conclusion, adjustments to primary CT parameters significantly influence image quality. The FOM serves as an effective tool for measuring the overall performance of image quality, facilitating post-optimization assessment and contributing to the development of safer, more efficient CT imaging protocols.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study
    Damien Racine
    Nick Ryckx
    Alexandre Ba
    Fabio Becce
    Anais Viry
    Francis R. Verdun
    Sabine Schmidt
    European Radiology, 2018, 28 : 5203 - 5210
  • [22] Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study
    Racine, Damien
    Ryckx, Nick
    Ba, Alexandre
    Becce, Fabio
    Viry, Anais
    Verdun, Francis R.
    Schmidt, Sabine
    EUROPEAN RADIOLOGY, 2018, 28 (12) : 5203 - 5210
  • [24] Estimating Image Quality and Radiation Dose Reduction with In-Plane Bismuth Shielding for CT Examinations: Phantom Study
    Liao, Y. L.
    Tsai, H. Y.
    Tyan, Y. S.
    Cheng, K. Y.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [25] Impact of four kVp combinations available in a dual-source CT on the spectral performance of abdominal imaging: A task-based image quality assessment on phantom data
    Greffier, Joel
    Frandon, Julien
    Sadate, Alexandre
    Akessoul, Philippe
    Belaouni, Asmaa
    Beregi, Jean-Paul
    Dabli, Djamel
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 243 - 254
  • [26] BEST IN PHYSICS (IMAGING): Task-Based Parameter Optimization for Low Signal Correction in Low Dose CT
    Gomez-Cardona, D.
    Hayes, J.
    Zhang, R.
    Li, K.
    Chen, G.
    MEDICAL PHYSICS, 2017, 44 (06) : 3285 - 3285
  • [27] The impact of ASiR-V on abdominal CT radiation dose and image quality-A phantom study
    Morup, Svea Deppe
    Stowe, John
    Kusk, Martin Weber
    Precht, Helle
    Foley, Shane
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2022, 53 (03) : 453 - 459
  • [28] Task-based optimization and performance assessment in optical coherence imaging
    Rolland, J
    O'Daniel, J
    Akcay, C
    DeLemos, T
    Lee, KS
    Cheong, KI
    Clarkson, E
    Chakrabarti, R
    Ferris, R
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2005, 22 (06) : 1132 - 1142
  • [29] Influence of deep learning reconstruction on task-based model observer performance in CT: an anthropomorphic head phantom study
    Hernandez-Giron, Irene
    Kaasalainen, Touko
    Makela, Teemu
    Peltonen, Juha
    Kortesniemi, Mika
    MEDICAL IMAGING 2022: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2022, 12035
  • [30] 3D task-based performance assessment metrics for optimization of performance and dose in breast tomosynthesis
    Richard, S.
    Samei, E.
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961